5.2

Adjustment

The DPS 5000 sensors are supplied fully calibrated by the manufacturer. When it is
necessary to re-calibrate the sensor, the procedure given in section 4.4.8 should be
followed. The minimum recommended re-calibration interval is once per year and
should be undertaken using a pressure source with a measurement accuracy of at
least ±0.01% of reading.
Note:
GE can provide a calibration service that is traceable to international
standards.
5.3

Repair

The DPS 5000 sensors contain no user serviceable items. For any repairs, return the
sensor to the manufacturer or an approved service agent.
5.4

Disposal

Disposal of the DPS 5000 sensor must be in accordance with the local statutory
regulations. Do not dispose of with regular household refuse.