Address 2 - Comp_Temp; Address 3 - Adc_Pres - GE DPS 5000 Series User Manual

I2c-bus pressure transducer
Hide thumbs Also See for DPS 5000 Series:
Table of Contents

Advertisement

Bit 15
Bit 14
Bit 7
Bit 6
Bit 31–0
COMP_PRES [31..0]: Read only – the most recently updated compensated
pressure value in the pressure units defined by the PRES_UNIT register
(address 84). Reading this register whilst the sensor is in the auto-update
mode will clear the CONV (bit 0) flag of the STATUS register (address 0).
0xn..n
Number type float
3.3.4
Address 2 – COMP_TEMP
Bit 31
Bit 30
Bit 23
Bit 22
Bit 15
Bit 14
Bit 7
Bit 6
Bit 31–0
COMP_TEMP [31..0]: Read only – the most recently updated compensated
temperature value in °C. Reading this register whilst the sensor is in the auto-
update mode will clear the CONV (bit 0) flag of the STATUS register (address 0).
0xn..n
Number type float
3.3.5
Address 3 – ADC_PRES
Bit 31
Bit 30
Bit 23
Bit 22
Bit 15
Bit 14
Bit 7
Bit 6
Bit 31–0
ADC_PRES [31..0]: Read only – the most recently updated pressure ADC value.
This value is used to calculate the value held in the MVOLT_PRES register
(address 6). Reading this register whilst the sensor is in the auto-update mode
will clear the CONV (bit 0) flag of the STATUS register (address 0).
0xn..n
Number type unsigned word
8
Bit 13
Bit 12
Bit 11
COMP_PRES [15..8]
R-X
Bit 5
Bit 4
Bit 3
COMP_PRES [7..0]
R-X
Bit 29
Bit 28
Bit 27
COMP_TEMP [31..24]
R-X
Bit 21
Bit 20
Bit 19
COMP_TEMP [23..16]
R-X
Bit 13
Bit 12
Bit 11
COMP_TEMP [15..8]
R-X
Bit 5
Bit 4
Bit 3
COMP_TEMP [7..0]
R-X
Bit 29
Bit 28
Bit 27
ADC_PRES [31..24]
R-X
Bit 21
Bit 20
Bit 19
ADC_PRES [23..16]
R-X
Bit 13
Bit 12
Bit 11
ADC_PRES [15..8]
R-X
Bit 5
Bit 4
Bit 3
ADC_PRES [7..0]
R-X
May 2015
Bit 10
Bit 9
Bit 8
Bit 2
Bit 1
Bit 0
Bit 26
Bit 25
Bit 24
Bit 18
Bit 17
Bit 16
Bit 10
Bit 9
Bit 8
Bit 2
Bit 1
Bit 0
Bit 26
Bit 25
Bit 24
Bit 18
Bit 17
Bit 16
Bit 10
Bit 9
Bit 8
Bit 2
Bit 1
Bit 0
K0582 Revision B

Advertisement

Table of Contents
loading

Table of Contents