... 00 02
NOT_STABLE_PRESS_TO_FULFILL_MANBA
G_OFFSET_CAL_1
... 00 07
NOT_STABLE_PRESS_TO_FULFILL_MANBA
G_GAIN_CAL1
... 00 09
NOT_STABLE_PRESS_TO_FULFILL_MANBA
G_GAIN_CAL2
... 00 0D
NOT_STABLE_PRESS_TO_FULFILL_MANBA
G_OFFSET_CAL_2
The PEEP pressure is not within -1 cmH2O to 0.5 cmH2O as required to fulfill the first
offset calibration of the MAN pressure transducer.
Stable Pressure is assumed when pressure is within -1 cmH2O to 0.5 cmH2O during 4
seconds. The timeout to reach stable pressure is 30 seconds.
Displays on Control panel:
The Fresh Gas Flow display shows the pressure measured by the INSP pressure
transducer until a stable pressure is reached.
The Preset O2 Conc. display shows the pressure measured by the EXP pressure
transducer until a stable pressure is reached.
The Minute Volume display shows the pressure measured by the PEEP pressure
transducer until a stable pressure is reached.
What to do: Check that the Manual breathing bag is deflated and not squeezed during
calibration.
The PEEPpressure is not within 16 cmH2O – 37 cmH2O as required to fulfill the first
gain calibration of the MAN pressure transducer.
Stable Pressure is assumed when pressure is within 16 cmH2O to 37 cmH2O during 4
seconds. The timeout to reach stable pressure is 30 seconds.
Displays on Control panel: Refer to (... 00 02 ) above.
What to do: The manual breathing bag may be too stiff or too elastic. Replace the
Manual breathing bag.
The PEEP pressure is not within 19 cmH2O – 40 cmH2O as required to fulfill the
second gain calibration of the MAN pressure transducer.
Stable Pressure is assumed when pressure is within 19 cmH2O to 40 cmH2O during 4
seconds. The timeout to reach stable pressure is 30 seconds
Displays on Control panel: Refer to (... 00 02 ) above.
What to do: The manual breathing bag may be too stiff or too elastic. Replace the
Manual breathing bag.
The PEEP pressure is not within -1 cmH2O – 0.3 cmH2O as required to fulfill the
second offset calibration of the MAN pressure transducer.
Stable Pressure is assumed when pressure is within -1 cmH2O to 0.3 cmH2O during 4
seconds. The timeout to reach stable pressure is 30 seconds.
Displays on Control panel: Refer to (... 00 02 ) above.
What to do: Check that the Manual breathing bag is deflated and not squeezed during
calibration.