Siemens KION Service Manual page 134

Anesthesia workstation
Table of Contents

Advertisement

x
250
... 00 FA ...
CON:
DIAG_BREATH_PHASE_FAILURE_SCON_ST
ATE
Sub data:
... AB CD
ACTIVE_SCON_STATE
x
251
... 00 FB ...
CON:
DIAG_BREATH_PHASE_FAIL_BREATHSYNC
_CMV
Sub data:
... 00 00
Inspiration
... 00 01
Pause
... 00 02
Expiration
x
252
... 00 FC ...
CON:
DIAG_BREATH_PHASE_FAILURE_PEND_VE
NTMODE
Sub data:
... 00 01
Manual ventilation
... 00 02
Volume Control
... 00 03
Pressure Support
... 00 04
Pressure Control
... 00 05
N/A
... 00 06
N/A
Fault: Internal program error.
Description: The software has detected a breathing phase synchronization failure and
has been synchronized again. A new expiration has been generated to synchronize
the software processes to solve this situation.
What to do:
1.
Single event: Report to EM-HSC.
2.
Repeted events: Replace PC 1730 Control. If the error remains, remove the
workstation from operation. Report to EM-HSC.
The failure has occurred when the "System Control"- process was in the state:
ACTIVE_SCON_STATE (hexadecimal form).
Fault: Internal program error.
Description: The software has detected a breathing phase synchronization failure and
has been synchronized again. A new expiration has been generated to synchronize
the software processes to solve this situation.
What to do:
1.
Single event: Report to EM-HSC.
2.
Repeted events: Replace PC 1730 Control. If the error remains, remove the
workstation from operation. Report to EM-HSC.
The failure has occurred in the breathing phase: Inspiration
The failure has occurred in the breathing phase: Pause
The failure has occurred in the breathing phase: Expiration
Fault: Internal program error
Description: The software has detected a breathing phase synchronization failure and
has been synchronized again. A new expiration has been generated to synchronize
the software processes to solve this situation.
What to do:
1.
Single event: Report to EM-HSC.
2.
Repeted events: Replace PC 1730 Control. If the error remains, remove the
workstation from operation. Report to EM-HSC.
The error occurred when pending ventilation mode: Manual ventilation.
The error occurred when pending ventilation mode: Volume control.
The error occurred when pending ventilation mode: Pressure support.
The error has occurred when pending ventilation mode: Pressure control.
Internal program state.
Internal program state.