Direct Function Keys; Displaying Non-Invasive Blood Pressure - GE Healthcare B30 User's Reference Manual

Patient monitor
Hide thumbs Also See for Healthcare B30:
Table of Contents

Advertisement

B30 Patient Monitor

Direct function keys

There are two module keys for NIBP on the module and on the monitor side panel:
On/Off
Cancel

Displaying non-invasive blood pressure

NIBP can be displayed in the digit field:
1
Figure 16-2 NIBP digit field display
(1)
Systolic and diastolic pressure value of non-invasive blood pressure
(2)
Label
(3)
Mean pressure value of NIBP
(4)
Time since the last autocycle measurement
(5)
NIBP autocycle time indicator
NOTE: When 60 minutes has passed from the latest NIBP measurement, the NIBP
numeric value digits turn gray. When 245 minutes has passed from the latest NIBP
measurement, the gray numeric value digits are replaced by a dashed line.
16-2
Starts and stops autocycling measurements.
Auto
Starts a single measurement, and cancels any measurement.
Start
2
4
3
5