B30 Patient Monitor
Patient connection
NOTE: Keep the ECG cable, lead set and connectors dry. Avoid excessive use of liquids
when cleaning the cables and connectors.
1
Figure 11-5 Example of the ECG setup with E-PSM(P)W module
(1)
E-PSMW or E-PSMPW module
(2)
ECG trunk cable, or 3-lead ECG cable with integrated leadwires
(3)
3 or 5 leadwire set
ECG electrodes (pre-gelled electrodes are recommended). Check the expiration data.
NOTE: For a comprehensive list of accessories, see the "Supplies and Accessories" catalog.

Lead measurement

The following table lists the electrodes needed to measure different ECG leads:
Lead
I
II
III
aVR
aVL
aVF
V5
11-4
L=YELLOW (IEC)
R=RED (IEC)
LA=BLACK (AAMI)
RA=WHITE (AAMI)
LE AD I
R/R A
L/LA
F/LL
F=GREEN (IEC)
LL=RED (AAMI)
2
Electrodes needed
R/RA, L/LA + F/LL or N/RL
R/RA, F/LL + L/LA or N/RL
LL/LA, F/LL + R/RA or N/RL
N/RL, R/RA, L/LA, F/LL
N/RL, R/RA, L/LA, F/LL
N/RL, R/RA, L/LA, F/LL
N/RL, R/RA, L/LA, F/LL, C5
R=RED (IEC)
L=YELLOW (IEC)
RA=WHITE (AAMI)
LA=BLACK (AAMI)
N=BLACK (IEC)
RL=GREEN (AAMI)
3
C=WHITE (IEC)
V=BROWN (AAMI)
F=GREEN (IEC)
LL=RED (AAMI)