364xS User's Manual
Proof pressure
The transmitter can be exposed without leaking (meet ANSI/ISA–S 82.03 hydrostatic test requirements and
SAMA PMC 27.1) to line pressure of up to 38.5 MPa, 385 bar, 5585 psi.
Electromagnetic compatibility (EMC)
Emission:
Immunity:
Radiated electromagnetic immunity level:
Conducted electromagnetic immunity
level:
Surge immunity level
Fast transient (Burst) immunity level:
Humidity
Relative humidity:
Condensing, icing:
Vibration resistance (according to IEC 60068–2–26)
Accelerations:
Shock resistance (according to IEC 60068–2–27)
Acceleration:
Duration:
Wet and dust-laden atmospheres
The transmitter is dust and sand tight and protected against immersion effects as defined by IEC 60529
(1989) to IP 67 (IP 68 on request) or by NEMA to 4X.
Fill fluid warning
Be sure that the fill fluid can mix safely with the process fluid in case of rupture of the sensor membrane.
Corrosion
A fluid/material compatibility table is available at www.abb.com searching for "CORROSION.pdf" or from
local ABB representatives
Data of the table are based on information from manufacturers.
All data is based on a temperature of 20°C (70°F) unless noted otherwise.
Since corrosion involves many more variables than this table considers, such as trace contaminants,
aeration or temperature-concentration profile, stress corrosion cracking and pitting, the table should be used
only as a reference in narrowing the choice of materials that merit further investigation. Suitability of a
particular material is best determined by field test. For this purpose, please contact the local ABB
representatives.
For safety purpose the design corrosion allowance of differential pressure instrument flanges is
approximately 1.5 mm (0.04 in). Therefore from the safe containment of liquids compatible with a specific
material according to the above mentioned table, the expected instrument lifetime is more than 10 years (the
previous statement on the corrosion influence variables apply).
Comply with EN 61000-6-3
Comply with EN 61000-6-2
30V/m (according to IEC 1000–4–3, EN61000–4–3)
30V (according to IEC 1000–4-6, EN 61000–4–6)
4kV (according to IEC 1000-4–5 EN 61000–4–5)
4kV (according to IEC 1000–4–4 EN 61000–4–4)
up to 100% annual average
admissible
up to 2g at frequency up to 1000Hz
50g
11ms
WARNING !