STATUS STATUS BAR STATUS BAR LAYOUT STATUS BAR MESSAGE AREA STATUS BAR ICONS STATUS BAR ICON LEVEL STATUS BAR MENU OPTIONS STANDARD OPTIONS USING THE MENU MENU ITEMS MEASURING FLUE GASSES 17-18 PRINTING KANE INFRARED PRINTER 2 KANE 358 MANUAL...
Page 3
PRESSURE & TEMPERATURE TESTING TEMPERATURE & PRESSURE DISPLAY VIEWING, SENDING & PRINTING PRESSURE MEASUREMENT GOOD PRACTICE LARGE BORE TUBING ISSUES TESTS 24-27 COMMISSIONING TEST 24-25 LET-BY & TIGHTNESS TESTING 26-27 PRINTOUTS SPECIFICATIONS 29-30 EU DECLARATION OF CONFORMITY COLD WEATHER PRECAUTIONS KANE 358 MANUAL...
The display’s bottom line also highlights analyser status at all times. Your KANE358 can print out test results using an optional infrared printer or to KANE’s wireless printer App. Your KANE358 stores up to 30 logs of any combination of Combustion, AUX, Temperature &...
Navigate down – Short press to scroll down Data hold – Short press to hold current data on screen (see status bar section) Pump toggle – long press to toggle the pump on and Function buttons 6 KANE 358 MANUAL...
Watertrap drain with red plug Pressure connection P1 Red connector – plugs into Flue Flue Probe Temperature Plug – Gas Inlet plugs into T1 The narrower (+ve) pin must be on the right hand side Flue Probe Hose KANE 358 MANUAL...
If your batteries discharge so that the analyser enters a low power shutdown, 1 hour’s charge provides approx 2 hours continuous use. BATTERY DISPOSAL Always dispose of depleted batteries using approved disposal methods that protect the environment. KANE 358 MANUAL...
● Altitude to 2000m ● Ambient temperature 0°C-40°C ● Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50%RH at 40°C ● Mains supply fluctuations not to exceed 10% of the nominal voltage 10 KANE 358 MANUAL...
Your analyser’s large display shows 5 lines of tests & a status bar. The backlight activates on each button press then turns off after 10 seconds. Navigate through your options and menu choices via the 3 dedicated & buttons Button presses are either short or long presses. KANE 358 MANUAL...
Status bar STATUS BAR The Status bar shows instrument status and offers options based on your settings. Navigate through the status bar options via the & buttons when the status bar is on the display. Status bar 12 KANE 358 MANUAL...
CALIBRATION DUE between calibration due WARNING MESSAGE symbol & current time Display alternates AIR PURGE between calibration due “SNOOZE” TIMER symbol & current time Display alternates LOW BATTERY between low battery WARNING MESSAGE symbol & current time KANE 358 MANUAL...
Icons give quick & simple status information: STATUS BAR ICON LEVEL ICON LIST PUMP Pump status On or Off Analyser has logged/ LOGGED DATA stored data Data being sent to App or SENDING DATA Printer Battery Status indication BATTERY 14 KANE 358 MANUAL...
Displays selected fuel type - To change, long press button to display selection indicators. Press buttons to select fuel then press button Turn rotary dial to Aux position to make changes to the AUX screen then long press button until cursor flashes KANE 358 MANUAL...
As you navigate up or down through the menu items will move up or down the screen returning to the beginning. NOTE: To exit Menu simply turn your analyser’s rotary dial to any position but any changes not entered will not be stored. 16 KANE 358 MANUAL...
With balanced flues, make sure the probe is positioned far enough into the flue so no air can “back flush” into the probe. WARNING Ensure the flue probe handle does not get hot! KANE 358 MANUAL...
PRINTING To print, simply press and release the button. Select your print destination to either KANE IRP1 optional printer or App. Printing can be stopped by pressing the button again. KANE INFRARED PRINTER Switch on your printer and make ready to accept data with its infrared receiver in line with your analyser’s emitter on top of the...
The cursor flashes and the line number appears in the status bar. Use to select your option to appear on the line then press to enter this option. O2/EFF SCREEN Oxygen reading Combustion efficiency reading T1 Flue Temperature reading T2 Inlet Temperature reading Nett Temperature reading Status bar KANE 358 MANUAL...
An icon displays when your analyser has stored data. To view current memory turn rotary dial to MENU then select LOGS to display this: No. of Combustion logs stored No. of Auxiliary logs stored No. of Pressure & Temperature logs stored Total memory used Menu options 20 KANE 358 MANUAL...
List of available logs - & Navigate & select using Buttons REPORT VIEW MENU OPTIONS View stored Combustion Reports – if View stored Auxiliary Reports – if any View stored Pressure & Temperature Reports – if any Returns to previous menu KANE 358 MANUAL...
Prints currently selected report Navigate to next available report if there’s more than one report Navigate to previously selected report - only once navigation begins Return to main menu Note: commissioning and tightness are accessed via the TEST position. 22 KANE 358 MANUAL...
VIEWING, SENDING & PRINTING Press button to send a full Pressure & Temperature report either to the optional KANE IRP-2 printer or the App. Press and hold button for 2 seconds to log a pressure and temperature report - See PRINTING to print stored reports.
Rotate dial to TESTS position. Press the followed by the & follow your analyser’s instructions. TEST 1 – CHECK THE APPLIANCE AT MAX GAS RATE Switch on appliance to max rate & zero your analyser in outside fresh air. 24 KANE 358 MANUAL...
Page 25
To finish press the enter symbol. To continue next press down triangle followed by enter symbol TEST 4 – MEASURE FLOW & RETURN TEMPERATURES FROM THE APPLIANCE All measured readings are logged & can be printed to our optional KANE IRP-2 printer or KANE’s App via wireless module. KANE 358 MANUAL...
If Let-by test passes, adjust gas pressure for the tightness test & press to start the stabilisation test – display shows: Real time pressure reading Stabilisation default time is 60 seconds When complete press to start the tightness test: 26 KANE 358 MANUAL...
Page 27
VIEWING/PRINTING Press button to print a full Let-By & Tightness report to our optional KANE IRP-2 printer or KANE’s App via wireless module. Let-by & Tightness reports are automatically stored. Page 20 explains how to view & print stored reports.
Natural Gas, Light Oil, Bio Oil, Coke, LPG, Wood, Town Gas, Butane European & Propane Battery Life >8 hours (continuous with pump on) The KANE358 is independently tested and certified to EN 50379, Certification Parts 1-3 in accordance to 1st German Federal Emission Control Ordinance (BimschV) KANE 358 MANUAL...
Page 30
Operating Conditions Temperatures 0 - 45°C Humidity 15 to 90% RH, (non-condensing) Power Supply Rechargeable batteries, USB Charging Physical Characteristics Weight Approx. 0.625g Dimensions L: 216mm x H: 105mm x W:45mm 30 KANE 358 MANUAL...
This declaration of conformity is issued under the sole responsibility of the manufacturer:- Kane International Ltd. Kane House, 11 Bessemer Road, Welwyn Garden City, Hertfordshire. AL7 1GF, UK. Tel: +44 1707 375550 Web: www.kane.co.uk The KANE358 is in conformity with the relevant Union harmonization legislation below:...
‘ON’ sampling fresh air for a few hours whilst connected to mains power. If, after this, you still experience problems please contact a KANE Service Centre or authorised Service Partner. 32 KANE 358 MANUAL...
Page 33
THIS PRODUCT CONFORMS WITH THE FOLLOWING PLEASE RECYCLE MADE IN THE UK KANE 358 MANUAL sales@norrscope.com...
Need help?
Do you have a question about the 358 and is the answer not in the manual?