Power Consumption; Power Specification Pinaacle 900T And Pinaacle 900Z; Power Specification Pinaacle 900H - PerkinElmer PINAACLE 900 Series Customer Hardware And Service Manual

Table of Contents

Advertisement

Preparing Your Laboratory
48
It is essential to avoid ground loops.

Power Consumption:

10100 VA for the PinAAcle 900T and the PinAAcle 900Z
10100 VA for the PinAAcle 900H
800 VA for the PinAAcle 900F
140 VA for the Cooling System

Power Specification PinAAcle 900T and PinAAcle 900Z

230 VAC
The PinAAcle 900T and 900Z will ship standard an IEC 60309 32A power
connector.
Operating range of 207 - 253 VAC/60 Hz, 30 A
(2 pole- 3 wire-grounding), and a dedicated circuit.
fuse 30 A slow blow, must be able to hold a 3x surge current for 3 seconds.
The mains line house wiring must have a minimum wire cross section of 6 mm
(8 gauge) or 10 mm

Power Specification PinAAcle 900H

230 VAC
The PinAAcle 900H will ship standard an IEC 60309 16/20 A power connector.
Operating range of 207 - 253 VAC/60 Hz, 20 A
('2 pole- 3 wire-grounding), and a dedicated circuit.
fuse 20 A slow blow, must be able to hold a 3x surge current for 3 seconds.
The mains line house wiring must have a minimum wire cross section of 4 mm²
(10 gauge) or 6 mm² (8 gauge) for distances > 10 m.
10%, 50 or 60 Hz ± 0.3 Hz, single phase alternating current.
²
(6 gauge) for distances > 10 m.
10%, 50 or 60 Hz ± 0.3 Hz, single phase alternating current.
²

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PINAACLE 900 Series and is the answer not in the manual?

Questions and answers

charles chatham
April 16, 2025

i'm getting a "changing BOC" error. What is it?

Table of Contents

Save PDF