Emd (Electromagnetic Disturbances) - Canon AX-C4343W Technical Document

Digital x-ray imager 2.4 ghz/5 ghz
Table of Contents

Advertisement

12
Guidance and manufacturer's declaration for EMD

EMD (Electromagnetic Disturbances)

The AX-C4343W is designed and tested to comply with IEC 60601-1-2 (EN 60601-1-2)
which is applicable regulations regarding EMD for devices and need to be installed and
put into service according to the following information.
1. Electrical equipment needs special precautions regarding EMD and needs to be
installed and put into service according to the information provided in the document.
(Table 1 to Table 4)
2. WARNING:
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
3. To maintain the optimum EMD performance, use only the designated cables.
4. WARNING:
Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.
5. WARNING:
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the AX-C4343W, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.