With foundation fieldbus communication, pressure transmitters (92 pages)
Summary of Contents for ABB 4600 Series
Page 1
Operating Instructions 4600 Series Models 4681 and 4686 Oxygen Transmitters, for use with ZGP2 Probes 20.0 % Oxygen 700°C ABB Instrumentation...
Page 2
St Neots, U.K. – Cert. No. Q5907 environmental applications. Stonehouse, U.K. – Cert. No. FM 21106 As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide. EN 29001 (ISO 9001) We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and support.
1 INTRODUCTION 2 PREPARATION 2.1 Checking the Code Number The Models 4681 and 4686 Oxygen Transmitters are designed for continuous monitoring of Oxygen content in 2.1.1 Wall-/Pipe-mounted Instruments – Fig. 2.1 applications using 'in situ' ZGP2 probes. Operation and programming of each Transmitter is via four tactile membrane switches and a digital display located on the front of the instrument.
3 MECHANICAL INSTALLATION 3.1 Siting Requirements Caution. • Mount in a location free from excessive vibration. Maximum Distance • Mount away from harmful vapours and/or dripping 100m fluids. Information. It preferable mount instrument at eye level, allowing an unrestricted view of A –...
4 ELECTRICAL CONNECTIONS Warning. Before making any connections, ensure that the power supply, any high voltage-operated control circuits and high common mode voltages are switched off. 4.1 Access to Terminals 4.1.1 Wall-/Pipe-mounted Instruments – Fig. 4.1 Remove protection cover slide down Slacken captive...
These components can be connected either across the load or directly across the relay contacts. On 4600 Series instruments the RFI components must be fitted to the relay terminal block along with the supply and load wires – see Fig 4.3.
…4 ELECTRICAL CONNECTIONS 4.5 Selecting the Mains Voltage 4.5.1 Wall-/Pipe-mounted Instruments – Fig. 4.6 Remove cap and screw Information. Use a small, flat-bladed screwdriver to remove Remove front the screw cap from the case. panel Slacken captive screws and remove protection cover Select the mains voltage required...
5 CONTROLS AND DISPLAYS 5.1 Displays – Fig. 5.1 5.2 Switch Functions – Fig. 5.2 The display comprises a 5-digit, 7-segment digital upper display line and a 16-character dot-matrix lower display line. In operation, the upper display line shows actual values of % Advance to oxygen, temperature, cell millivolts or alarm set points.
6 OPERATION 6.1 Instrument Start-up Ensure all electrical connections have been made correctly and apply power to the instrument. 6.2 Operating Page The operating page is a general use page in which continuously updated measured values and preset parameters can be viewed but not altered.
…7 PROGRAMMING Note. Before commencing a gas calibration procedure the analyzer and probe must be switched on and allowed to run for at least one hour to allow the system to stabilize thermally. 7.1 Single-point Calibration The calibration sequence involves standardizing the analyzer and the oxygen probe, using air as the test gas. Until a calibration sequence has been completed successfully, the existing slope remains unaffected.
7 PROGRAMMING… 7.2 Two-point Calibration Oxygen Calibration Sequence ----- OXYGEN CAL. SEQ, Cell Zero mV 0. 0 The upper display shows the millivolt offset of the oxygen probe from the last successful calibration. Cell Constant mV Span % of Theory 100.
Page 18
…7 PROGRAMMING …7.2 Two-point Calibration Continued from previous page Calibration Pass/Fail ----- On completion a calibration status message is displayed. Calibration Pass: Calibration sequence successful Calibration Pass Failed Constant: Cell offset >±10mV Failed Constant (upper display shows cell constant) Cell output <90% or >110% of slope Failed Span %: Failed Span (upper display shows measured slope)
7 PROGRAMMING… 7.3 Preset Calibration Oxygen Calibration Sequence ----- OXYGEN CAL. SEQ. Cell Zero mV 0. 0 The upper display shows the millivolt offset of the oxygen probe from the last successful calibration. Cell Zero mV Span % of Theory 100.
…7 PROGRAMMING 7.4 Access to Secure Parameters Page A 5-digit security code is used to prevent tampering with the secure parameters. Security Code ----- [00000 to 199999] If an incorrect value is entered, access to programming pages is inhibited. Security Code Press to advance to the Language Selection Page...
7 PROGRAMMING… 7.6 Set Up Thermocouple Page Set Up Thermocouple ----- Page header SET UP T/C Thermocouple Type ----- Set the thermocouple type required: Type K, R, S, B or NONE. If thermocouple type is set to NONE, a preset temperature must be used – see below. T/C Type NONE Preset Temperature...
…7 PROGRAMMING 7.7 Set Up Outputs Page Set Up Outputs ----- Page header SET UP OUTPUTS Alarm A1 Action ----- Set the required alarm action from the following table: A1 Action The set point band is defined as the actual value of the set point plus or minus the hysteresis value.
Page 23
7 PROGRAMMING …7.7 Set Up Outputs Page Continued from previous page Retransmission Zero 00. 0 0 Set the required retransmission zero value, in % oxygen units. [0% to 20.00% (linear output)] Retrans Zero [0.01% to 0.25% (logarithmic output)]. Retransmission Span 25.
APPENDIX A – ELECTRICAL CALIBRATION Note. The instrument is calibrated by the company prior to despatch and an electrical calibration should be carried out if the instrument's accuracy is suspect suitably calibrated test equipment is available. only A1 Equipment Required a) Millivolt source (cell input simulator), –20.0 to 180.0mV.
APPENDIX A – ELECTRICAL CALIBRATION A3 Electrical Calibration Page x x x x x In this section the actual values denoted by ' ' are unimportant and are used to determine display reading stability when carrying out the electrical calibration procedure. Electrical Calibration Page ----- (To access this page refer to Fig.
Page 27
PRODUCTS & CUSTOMER SUPPORT A Comprehensive Instrumentation Range Customer Support Analytical Instrumentation ABB Instrumentation provides a comprehensive after sales service via a Worldwide Service Organization. Contact one of • Transmitters the following offices for details on your nearest Service and...
Need help?
Do you have a question about the 4600 Series and is the answer not in the manual?
Questions and answers