Replacing The Trolley Base - GE Modular MAC Trolley Service Manual

Ecg trolley
Table of Contents

Advertisement

Maintenance
This illustration applies to all trolley styles.
Retain the screws for later reattachment of the column to the base.
3.
Position the new column over the base and align the holes with the holes in
the base.
NOTE:
4.
Using the 1/2" socket wrench supplied with the trolley, insert the four 5/16-18 x
2.25" flanged HHCS and tighten the hardware.
5.
If you removed the trolley top, reattach it.
See
6.
After attaching the top to the trolley column, reattach the MAC device.
• For the Modular MAC Trolley, see
• For the MAC 1600 Trolley, see
7.
Replace the equipment that was removed and reconnect power cables to the
equipment on the trolley.

Replacing the Trolley Base

This procedure applies to all trolley styles.
Before removing the base, remove the MAC unit and accessories from the trolley. See
"Removing the MAC Unit from the Modular MAC Trolley" on page 22
on replacing the MAC unit to the Modular MAC Trolley. See
from the MAC 1600 Trolley" on page 23
to the 1600 Trolley.
26
For the MAC 1600 Trolley, position the column so the angled cut on the top
of the column slopes toward the casters with the brakes.
"Replacing the Modular or MAC 1600 Top " on page 24
MAC Trolley" on page 22
Trolley" on page 23
for instructions on reattaching the MAC device.
MAC™ ECG Trolley
"Removing the MAC Unit from the Modular
for instructions on reattaching the MAC device.
"Removing the MAC Unit from the MAC 1600
for instructions on replacing the MAC unit
for details.
for instructions
"Removing the MAC Unit
2044854-012E