DMM Settings
Range
OCO Resistance
1 MΩ
Off
1 MΩ
100 kΩ Off
100 kΩ
Note
0 Ω measurement as an offset null.
10. Connect the PXIe-4080/4082 to the calibrator using low thermal EMF cables, as shown in
the following figure.
Figure 3. 2-Wire Resistance Low Range Connection Configuration
11. Wait two minutes for the thermal EMF to stabilize.
12. Verify the accuracy for each configuration in the following table:
a)
Acquire the measurement at each specified configuration.
b)
Compare to the specified tolerances.
Calibrator Settings
Ext
Comp
Sense
Off
None
Off
None
For the 1 MΩ and 100 kΩ measurements, subtract the corresponding
NI DMM
HI
LO
HI
LO
Not Connected
PXIe-4080/4082 Calibration Procedure | © National Instruments | 13
Tolerances
As-Found
ppm of
ppm of
Reading
Range
95
10
95
6
Calibrator
HI
LO
HI SENSE
LO SENSE
V-GUARD
As-Left
ppm of
ppm of
Reading
Range
32
2
15
2
OUTPUT
SENSE
A
V-GUARD
AUX
CURRENT
GROUND