Failure
Code
Text message
F008
Error multi sensor commu-
nication
F013
Sensor signals a fault
F016
Adjustment data exchanged
F017
Adjustment span too small
F025
Invalid linearization table
F030
Process value out of limits
F034
EPROM hardware error
F035
EPROM data error
F036
Faulty program memory
F037
RAM hardware error
F038
Slave signals failure
F040
Hardware error
F041
Photomultiplier error
F045
Error on the current output
F052
Faulty configuration
F114
Error real time clock
POINTRAC 31 • Four-wire 8/16 mA/HART
This status message is inactive by default.
Cause
Additional sensors not switched on
EMC influences
No other sensor available
Error on the current input
No valid measured value
Connected instruments without function
Values of the min. and max. adjustment
exchanged
The values of the min. and max. adjust-
ment are too close together
Wrong value in the linearization table
Process values are not within the adjust-
ed measuring range
Electronics defective
Error in the internal instrument commu-
nication
Error during software update
Error in RAM
Connection cable to the Slave instru-
ment interrupted
Instrument not defined as Slave instru-
ment
Sensor defective
Error in the measured value recording
Error on the current output
Invalid parameter adjustment
Discharge accumulator
9 Diagnostics and servicing
Rectification
Check wiring between the sensors
Connect the sensors correctly and make
them ready for operation
Check current input
Check connected instruments (Slaves)
Correct adjustment data
Correct adjustment data
Correct linearization table
Repeat adjustment
Exchanging the electronics
Carry out a reset
Exchanging the electronics
Repeat software update
Exchanging the electronics
Exchanging the electronics
Define instrument as Slave
Check the connection cable to the Slave
instrument
Exchanging the electronics
Exchanging the electronics
Check wiring of the current output
Exchanging the electronics
Carry out a reset
Readjust real time clock
45
Need help?
Do you have a question about the POINTRAC 31 and is the answer not in the manual?
Questions and answers