7.4

EMC information

Your unit has been designed to meet EMC standards throughout its Service Life without additional
maintenance. There is always an opportunity to relocate your device within an environment that contains
other devices with their own unknown EMC behavior. If you believe your unit is affected by locating it closer
to another device, simply separate the devices to remove the condition.
Pressure and Flow Accuracy
The device is designed to perform within the pressure and flowrate accuracies specified in the user manual.
If you suspect that the pressure and/or flow rate accuracy is affected by EMC interference, remove power
and relocate the device to another area. If performance continues to be affected, discontinue use and
contact Philips Respironics.
SpO
and Pulse Rate Accuracy
2
The device is designed to capture the SpO
described in the sensor manufacturer's instructions for use. When 4 hours of successful oximetry data have
been achieved, the device indicates this to the user by displaying "Oximetry: Good Study". If you suspect
that your unit is affected by EMC interference, remove power and relocate the device to another area. If
performance continues to be affected, discontinue use and contact Philips Respironics.
Guidance and Manufacturer's Declaration - Electromagnetic Emissions – This device is intended for use in
the electromagnetic environment specified below. The user of this device should make sure it is used in such
an environment.
E
T
missions
EsT
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
Emission of Radio Frequency
Energy
RTCA/DO-160G Section 21
52
Specifications
and Pulse Rate oximetry data within the accuracy specification
2
C
E
omplianCE
lECTromagnETiC
The device uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class B
The device is suitable for use in all establishments, including
Class A
domestic establishments and those directly connected to the
public low-voltage power supply network.
Complies
This device is suitable for use onboard commercial airplanes
Category M
inside passenger cabin.
E
- g
nvironmEnT
uidanCE