Philips Respironics E30 User Manual page 42

Ventilation solution
Table of Contents

Advertisement

Philips Respironics E30 User Manual
Alert
SD Card Error:
Remove and
Reinsert
SD Card Full.
Humidification
Error. Contact
support if the
problem persists.
The attached
power supply
does not support
humidification.
Activate MODE?
Check Power
Busy
36
Alarms, alerts, and troubleshooting
Icon
Priority
Notification
Notification
Status
Alert
Notification
Notification
Status
Description
SD card error detected.
Device cannot read the SD
card.
A problem may exist with
the SD card or it was
ejected during a writing
activity, or it was inserted
incorrectly.
SD card is full.
Humidifier error (only when
humidifier is present)
Humidifier heater plate
error or humidifier not
properly connected to
therapy device.
Indicates that the attached
power supply is not
capable of supporting
humidification.
Indicates mode change
while blower is running or
therapy is being delivered.
Indicates an incompatible
power supply is attached.
Incompatible power supply,
or power cord is not fully
inserted into device's power
inlet.
Displayed when the device
is temporarily inaccessible
due to data communication.
User action
Remove SD card and reinsert.
If alert continues to occur,
replace with another card or
contact Philips Respironics or
the product support team.
Remove SD card and contact
Philips Respironics or the
product support team for a
new SD card.
Turn off device and
disconnect from power.
Detach the humidifier,
visually check that electrical
contacts are clear, then
reconnect humidifier
and power cord. If alert
continues, contact Philips
Respironics or the product
support team.
Contact Philips Respironics
or the product support team
to obtain a compatible
Philips Respironics power
supply. Or, operate therapy
device without humidifier.
Choose Yes to accept new
therapy mode, No to stay in
current mode.
Confirm power cord is fully
inserted into device's power
inlet. Confirm a compatible
Philips Respironics power
supply is attached. Switch to
compatible power supply if
needed.
No action needed.