Momentary Messages
Table 3-4 Momentary Messages
Message
Attach Leads
Attach Pads
Defib Disarmed
No Shock Delivered
Check Printer
Data Card Full
Data Card Interrupted
Data Card Not In Ser-
vice
Incompatible Data
Card
3-16
Momentary Messages
Momentary messages are temporary and only appear on the display for a few
seconds. They are intended for the end user and appear in the Instructions
for Use. They are duplicated here for the reference of the service person.
Before replacing any parts, check to see if the cables and flex circuits are
properly connected. See "Servicing Notes" on page 4-1.
Possible Cause
An attempt was made to begin pacing in
Demand Mode without ECG leads
attached to the patient.
The multifunction defib electrode pads are
not making proper contact with the patient.
The pads connection is compromised.
The mode is changed from Manual to
AED while the defibrillator is charged.
SHOCK
is not pressed within 30 seconds
of the defibrillator being charged.
DISARM
is pressed.
Patient impedance is too high or too low.
Printer paper is absent or jammed; the
printer door is not closed properly.
The incident is more than 2 hours in dura-
tion, causing the Data Card to fill.
An empty Data Card was not inserted for
the incident; the Data Card filled sooner.
No data is being recorded on the Data
Card because the Data Card was
removed during an incident.
The Data Card is inserted while the
M3500B/M5500B is on.
A Data Card other than the M3510A is
inserted.
Corrective Action
Attach leads to patient
1. Check the pads are applied to the
patient, as directed on the package.
2. Replace pads if the prompt continues.
Check the pads are applied to the patient
properly.
If a shock is indicated, deliver the shock
before changing modes.
To deliver a shock, press
30 seconds of the defibrillator being
charged.
None.
1. Make sure pads are applied properly.
2. Replace the pads.
3. Replace the pads cable.
1. Reload printer paper.
2. Make sure the door is closed properly.
None. A new Data Card can not be
inserted during an incident.
Use one empty Data Card per incident/
patient to decrease the chance of the card
filling.
• During incident, none. The Data Card
cannot be inserted during an incident.
• After incident, remove card, turn
power off, insert card, turn power on.
None. A Data Card must be inserted prior
to turning the M3500B/M5500B on for the
current patient.
Use only M3510A/M5510A Data Cards.
SHOCK
within