Error Codes
Table 3-2 Error Codes (Continued)
Error Code
80000 - 8FFFF
90000 - 90002
90003
90004 and 90005
90006
90007 and 90008
90009
9000A
A0000-A7FFF
A8000 - AFFFF
B0000 - BFFFF
F0000 - FFFFF
F0002
F0003
F0004
F0007
3-12
Meaning
IRDA failure.
Self Test failure - RAM/ROM or Gate
Array.
Self Test failure - Data Card circuits.
Self Test failure - Codec/time base.
Self Test failure - SpO 2.
Self Test failure - Pacer or
Defib.This error can be caused by
pressing the Pacer key during the
self-test.
Self Test failure - Front End.
Lithium backup battery failure.
Data Card failure.
Data Archival error.
Audio failure.
RAM/ROM failure.
Promo Mode failure.
Keyscanner failure.
System Communication error.
System Communication error
Possible Solutions
1. Replace Interface PCA.
2. Replace Control PCA.
1. Replace lithium battery.
2. Replace Control PCA.
Replace Control PCA.
1. Replace lithium battery.
2. Replace Control PCA.
1. Replace SpO2 PCA.
2. Replace Control PCA.
3. Replace Interface PCA.
1. Run the self-test again, making
sure not to touch the unit during
the test.
2. Replace Power PCA.
3. Replace Control PCA.
1. Replace Power PCA.
2. Replace Control PCA.
3. Replace Interface PCA.
1. Replace lithium battery.
2. Replace Control PCA.
1. Replace Data Card.
2. Replace Control PCA.
1. Replace Data Card.
2. Replace Control PCA.
Replace Control PCA.
Replace Control PCA.
Replace Control PCA.
1. Replace Keyscan PCA.
2. Replace Control PCA.
Replace Control PCA.
Cycle power.