M3500B/M5500B Service Manual
Replacement
1. Using the new gasket that came with the connector, install the gasket in
place on the connector. Slide the cable and ferrite back through the hole
in the case.
2. Align connector and gasket with notches in case and insert in hole. Make
sure the gasket is in place properly, between the connector and the outside
of the case.
3. Press in firmly on the connector to compress the gasket. Slide the metal
retainer clip into place behind the 2 plastic lugs on the connector, being
sure it locks into place. Use the new retainer clip that came with the con-
nector.
4. Loop the cable around the case post and secure with the cable tie pro-
vided.
5. Connect the ECG connector to the Interface PCA, being sure the ECG
cable is above the SpO
6. Replace the plastic insulating shield over the SpO
Align the hole in its wide end over the case post. Slide the left edge of its
narrow end under the lips on the metal retaining clips on the ECG and
SpO
connectors. Secure the shield to the clips with the original tape, or
2
use other plastic electrical tape.
7. Replace and reconnect the Control Stack. See "Control Stack" on page 4-
33.
flex circuit.
2
Top Case Assemblies
and Interface PCAs.
2
4-41