Top Case Assemblies
4-46
Remove the 4 Torx #10 screws now visible around the edge of the
b.
keypad. See Figure 4-27.
Remove the Manual Door shaft and rails. Note the orientation of the
c.
shaft between the rails. If at all possible keep the shaft captured
between the rails.
Using a sharp tool such as a utility knife, carefully pick up one corner
d.
of the membrane switches and peel the entire switch assembly off the
case. Be careful not to gouge the case. The membrane switches will
likely peel apart into several layers - be sure to remove all layers.
Slide the 2 flex circuit tails out of the slot in the case.
e.
7. Turn the top case over and loosen the Interface PCA (do not remove it) as
follows:
Remove (1) T10 screw from the Interface PCA (near the ECG and
a.
SpO
connections). See Figure 4-26 on page 4-44.
2
Release the retaining clip at the end of the PCA nearest the Ambient
b.
Light Sensor. Use a small flat bladed screwdriver to gently pry
between the clip and the PCA, levering the PCA up and free of the
retaining clip. Simply leave the PCA loose in position.
Removal and Replacement