16. Error messages

Error no. Description
00
Electronics overheating–Safety shutdown
01
Electronics overloaded
Foot control error –
07
initializing
09
Foot control error
19
Running time limiter
99
System failure
Motor temperature too high
Motor plug-in connection
If one of the error messages described above cannot be rectified by switching off the Implantmed and then switching
it on again, the equipment must be checked by an authorized W&H service partner (see Page 52). If a total failure of
the equipment occurs caused by external circumstances, the equipment must be switched off and then on again.
Solution
Switch off device, allow device to cool for at least
10 minutes, re-start
Switch off device, allow device to cool for at least
10 minutes, re-start
Switch off device, re-start, do not actuate
foot control when switching on
Switch off device, check plug-in
connection of foot control, re-start
Switch off device and re-start
Switch off device, allow motor to cool for at least
10 minutes, re-start
Switch off device, check plug-in motor connection,
allow motor to cool for at least 10 minutes, re-start
33

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the implantmed SI-923 and is the answer not in the manual?

Questions and answers

Ahmed Esawi
February 10, 2025

Error message 00 I was trying off the switch but the same error message 00

This manual is also suitable for:

Implantmed si-915

Table of Contents

Save PDF