13. Error messages
Error no. Description
00
Electronics overheated – safety shutdown
01
Electronics overloaded
02
Voltage too high
07
Initialization error
09
Foot control error
19
Running time limiter
99
System failure
Motor plug-in connection
> Switch the control unit off and on again. If any of the described problems cannot be corrected by the remedy provided
above, the unit will need to be inspected by an authorized W&H service partner.
> Switch the control unit off and on again in case of a total system failure.
Solution
Switch off device, allow device to cool for at least
10 minutes, re-start
Switch off device, allow device to cool for at least
10 minutes, re-start
Switch off device, check voltage, re-start
Switch off device, re-start, do not actuate foot control
and display when switching on
Switch off device, check plug contacts
of foot control, re-start
Switch off device and re-start
Switch off device, allow device to cool
for at least 10 minutes, re-start
Switch off device, check plug contacts,
re-start
37