Advertisement

Instructions for use
SI-923 / SI-915

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OsseoSet 200 and is the answer not in the manual?

Questions and answers

Summary of Contents for W&H OsseoSet 200

  • Page 1 Instructions for use SI-923 / SI-915...
  • Page 2: W&H Symbols

    W&H Symbols WARNING! ATTENTION! General explanations, (Risk of injury) Important notes! without risk to (to prevent damage occurring) persons or objects Thermo washer disinfectable Sterilizable W&H Service up to the stated temperature Only for USA Caution: Federal law restricts this device to sale by or on the order of a dentist, physician or any other practitioner licensed by the law of the state in which he or she practices to use or order the use of the device.
  • Page 3: Table Of Contents

    Contents W&H Symbols ................................2 Introduction ................................4 – 5 Unpacking ..................................6 Equipment supplied ..............................7 Safety notes ................................8 – 11 Description of front panel ............................12 Description of rear panel ............................13 Description of motor with cable ..........................14 Starting operation foot control ..........................15 – 20 Starting operation –...
  • Page 4 1. Introduction For your safety and the safety of your patients These Instructions for Use explain how to use your OsseoSet 200. However, we must also warn against possible hazardous situations. Your safety, the safety of your team, and of course, the safety of your patients are of paramount importance to us.
  • Page 5: Introduction

    Introduction Production according to EU Directive EU Directive 93/42/EEC has been used as a basis in the design and manufacture of this medical product and this applies to the dental surgical units – OsseoSet 200 SI-915 and – OsseoSet 200 SI-923 in the condition as supplied by us.
  • Page 6: Unpacking

    2. Unpacking ➊ ➍ Lift out insert with stand. Remove carton with mains cable. ➋ ➎ Remove foot control and Remove carton containing insert. motor, motor support, drill gauge, locking pins and instrument. ➌ The packaging is environmentally friendly and can be Lift out insert with control disposed of by industrial recycling companies.
  • Page 7: Equipment Supplied

    3. Equipment supplied ❍ Control unit REF NB05018200 (230 V) ❍ Foot control S-NU 1, REF NB04491600 ❍ Motor with 1.8 m cable including 5 clips, REF NB04720016 ❍ Mains cable REF NB01343700 (Europe) / REF NB03212700 (UK, IRL) / REF NB02909300 (AUS, NZ) ❍...
  • Page 8 4. Safety notes Please ensure that you carry out the following instructions – Only fit the handpieces when the motor is at a complete standstill. – Never touch rotary instruments which are still rotating. – Never touch the chuck mechanism of straight and contra-angle handpieces while they are still running. –...
  • Page 9: Safety Notes

    Safety notes Danger zones M and G In accordance with IEC 60601-1 / UL 60601-1, the control unit is not suitable for use in potentially explosive atmospheres or with potentially explosive mixtures of anaesthetic substances containing oxygen or nitrous oxide. Control unit The control unit is classed as "conventional equipment"...
  • Page 10 Safety notes Mains cable Only connect to an earthed socket outlet. Power failure In the event of a power failure, if the OsseoSet 200 is switched off, or when alternating between programs, the last values set are saved and re-activated on power-up. System failure A total system failure of the OsseoSet 200 does not constitute a critical fault.
  • Page 11 Safety notes Coolant The OsseoSet 200 is designed for use with physiological saline solution. Use only suitable irrigation fluids and comply with the medical data and instructions from the manufacturer. Coolant bottle, coolant bag and sterile fittings can be obtained from a pharmacist. Rotational energy Fast deceleration of the bur can, at times, cause the selected torque to be overloaded as a result of the rotational energy stored in the drive system.
  • Page 12: Description Of Front Panel

    5. Description of front panel Program buttons Pump arm Stand holder Display OPEN Pump arm Motor connection socket...
  • Page 13: Description Of Rear Panel

    6. Description of rear panel Stand holder Connecting socket Power switch for foot control ON / OFF Power socket Fuse holder with 2 fuses REF 04014700 (250 V – 1.25A slow-blow)
  • Page 14: Description Of Motor With Cable

    7. Description of motor with cable The motor may not be dismantled! The motor bearings are lubricated for life. No oil lubrication or other maintenance is necessary. To prevent the instrument on the motor attachment from turning during the transmission, the locking pin supplied can be pushed into one of the two holes (see illustration).
  • Page 15 8. Starting operation foot control – Fitting / Removing the handle Fitting the handle For easy fitting of the handle rotate the foot control by 180°. ➊ Remove the protective cap from the handle hole. ➋ Locate the guide pin of the handle in the recess in the hole within the foot control. ➌...
  • Page 16 Starting operation foot control – Disassembly of foot control Disassembly of foot control – removing the pedal for disinfection and cleaning (see Page 30) – Remove the handle from the foot control. – Press on one of the two guide pins with the tool provided and lift the pedal slightly. –...
  • Page 17 Starting operation foot control – Reassembly of foot control – Make sure that both springs are inserted in the appropriate recess. – Position the pedal at the end of the foot control into the mounting plate. – Press the two guide pins in front of the foot control at the same time and allow the pedal to engage. –...
  • Page 18 Starting operation foot control – Description of pedal ORANGE Change program BLACK (Pedal) Motor start LIGHT-GREY DARK-GREY Pump Change motor direction Function BLACK VARIABLE or ON / OFF (Factory setting = variable) LIGHT-GREY Pump ON / OFF ORANGE Program 1 to 5 DARK-GREY Motor forward / reverse drive...
  • Page 19 Starting operation foot control – Description of pedal Changing program The Programs 1 – 5 can be selected in ascending sequence by operating the ORANGE button. When changing from Program 5 to Program 1, a longer confirmation signal is heard. With each program change, the speed is automatically set to forward operation.
  • Page 20: Starting Operation Foot Control

    Starting operation foot control – To change from VARIABLE to ON / OFF Keep Program button P3 depressed throughout this procedure! ➊ Press P3 for approx. 4 seconds ➋ Keep P3 depressed and simultaneously press PLUS and MINUS ➌ Keep P3 depressed and carry out adjustment. 01 = VARIABLE (factory setting) –...
  • Page 21: Starting Operation - General

    9. Starting operation – General ➊ ➍ Connect the mains cable Fit the motor support. and foot control. ➋ ➎ Connect the motor cable. Fit the irrigation tubing Note the positioning! set. ➌ ➏ Insert the stand. Close pump arm. Note the positioning!
  • Page 22 10. Reset factory settings Factory setting always starts with Program 1 (P1). ➊ Switch off the control unit ➋ Keep P1 depressed and simultaneously switch on the control unit ➌ Keep P1 depressed as long as on the display the adjustment “DE FAU” appears...
  • Page 23: Factory Settings

    Factory settings (P1 – P5) Transmission ratio 20:1 20:1 Factory setting rpm 40 000 2 000 Adjustable range rpm 300 – 40 000 15 – 2 000 15 – 2 000 Motor direction forward forward forward Pump Factory setting Ncm max.
  • Page 24 11. Control unit operation – Changing the program (P1 – P5) Activate the required Program (P1 – P5) by pressing the corresponding Program button. During selection an audible signal can be heard and the Program button is illuminated. On the display the selected Program appears with the adjusted range in rpm, e.g.
  • Page 25: Control Unit Operation

    Control unit operation – Changing the speed (P1 – P3) Keeping PLUS / MINUS depressed activates the repeat function and the values are continuously increased / decreased. ➊ Press Program button (P1–P3) ➋ Increase speed ➋ Decrease speed...
  • Page 26 Control unit operation – Changing the torque (P4 – P5) Adjustable range from 5 up to 50 Ncm. Keeping PLUS / MINUS depressed activates the repeat function and the values are continuously increased / decreased. When changing from 5 to 50 Ncm and 50 to 5 Ncm, a longer confirmation signal is heard. When the adjustable torque in the the forward/reverse drive mode is reached, the motor switches off automatically.
  • Page 27 Control unit operation – Changing the coolant flow (P1 – P5) Factory setting 100 %. Adjustable range 65 %, 80 % and 100 %. Depressing PLUS / MINUS the values are continuously increased / decreased. Keep Program button P2 depressed throughout this procedure! ➊...
  • Page 28 12. Error messages Error No. Description Remedy Electronics overheating – Switch off the equipment, allow to cool safety shutdown for at least 10 minutes, then re-start Electronics overloaded Switch off the equipment and then re-start Foot control error – Switch off the equipment, re-start, do not initializing actuate foot control when switching on...
  • Page 29: Error Messages

    Error messages If one of the error messages described above cannot be rectified by switching off the OsseoSet 200 and then switching it on again, the equipment must be checked by a service and repair organization authorized by W&H (see Page 45 – 47). If a total failure of the equipment occurs caused by external circumstances, the equipment must be switched off and then on again.
  • Page 30 13. Disinfection, Cleaning, Sterilization – Wear protective gloves. – Disinfect and clean the motor immediately after every treatment! – Sterilize the motor following disinfection and cleaning. – Motor with cable and motor support must be sterilized prior to every use. Control unit, foot control, handle Manual disinfection and cleaning The front panel of the control unit is sealed and may be wiped clean.
  • Page 31: Disinfection, Cleaning, Sterilization

    Disinfection, Cleaning, Sterilization Motor with cable Do not twist and kink the motor cable! Do not coil it too tightly! Before disinfection and cleaning push the motor protective cover onto the motor attachment! Manual disinfection – If heavily soiled: Clean first with disinfectant cloths. –...
  • Page 32 Disinfection, Cleaning, Sterilization Sterilization and storage – Note the instructions of the unit manufacturer. – Only sterilize cleaned motors and accessories. – Wrap motors and accessories in sterile goods packing according to EN 868-5. – Store sterile goods dust-free and dry. W&H recommends sterilization according to EN 13060, class B Permitted sterilization methods Vacuum sterilization –...
  • Page 33: Accessories

    Use only original W&H accessories/spare parts 14. Accessories or accessories approved by W&H For spare parts and other accessories not listed here, please contact your local W&H representative or service station. NB04720016 NB04006800 NB02610500 Motor with 1.8 m cable Locking pin Cannula for internal bur cooling NB10207554 NB10207513...
  • Page 34 15. Servicing Regular checking of OsseoSet 200 and accessories Regular servicing of function and safety of the OsseoSet 200 including the accessories is necessary and should be carried out at least once every three years, unless shorter intervals are prescribed by law. The inspection must be undertaken by a qualified organization and needs to include the following procedures: –...
  • Page 35: Servicing

    Servicing Motor and cable The standard ISO 11498 stipulates a durability of at least 250 sterilization cycles. In the case of the motor and cable from W&H, regular servicing is specified only after 500 sterilization cycles, or one year. Repairs If a defect occurs, always return all the equipment, due to the fact that with motor malfunctions, an inspection of the electronic controls is also necessary! Returns...
  • Page 36 16. Technical data OsseoSet SI-923 SI-915 Supply voltage: 220 – 240 V 100 – 130 V Rated current: 0.1 – 0.8 A 0.2 – 1.6 A Permitted voltage fluctuation: ± 10 % ± 10 % Frequency: 50 – 60 Hz 50 –...
  • Page 37: Technical Data

    Technical data Physical characteristics Temperature in storage: -40 °C (-40 °F) to +70 °C (+158 °F) Air humidity in storage: 8 % to 80 % (relative), non-condensing at +40 °C (+104 °F) Temperature in operation: +10 °C (+50 °F) to +40 °C (+104 °F) Air humidity in operation: 15 % to 80 % (relative), non-condensing at +40 °C (+104 °F) Recycling...
  • Page 38: Letter Of Indemnity

    Letter of indemnity This W&H product has been manufactured with great care by highly qualified specialists. A wide variety of tests and controls guarantee faultless operation. Please note that claims under warranty can only be validated when all the directions in the instructions for use have been followed. As manufacturer, W&H is liable for material or manufacturing defects within a warranty period of one year from the date of purchase.
  • Page 40 ✄ ✄ ✄ ✄...
  • Page 41: Authorized W&H Service Organizations

    Authorized W&H Service organizations W&H Dentalmechanik GmbH, Ignaz-Glaser-Straße 60, A-5111 Bürmoos, Tel. +43 (0)6274 / 6236-239, Fax: +43 (0)6274 / 6236-890, E-mail: service.dept@wh.com W&H Deutschland GmbH & Co.KG, Raiffeisenstraße 4, D-83410 Laufen/Obb., Tel. +49 (0)8682 / 89 67-0, Fax +49 (0)8682 / 89 67-11, E-mail: office.de@wh.com W&H CH-AG, Aathalstrasse 84, CH-8610 Uster ZH, Tel.
  • Page 42 Authorized W&H Service organizations W&H NORDIC AB, Tillverkarvägen 6, S-187 66 Täby, Tel. +46 (0)8 / 445 88 30, Fax +46 (0)8 / 445 88 33, E-mail: office@whnordic.se Dentamechanik s.r.o., Miloš Kuttelvašer, Nádražní 6, CZ – 370 02 České Budějovice Tel.
  • Page 43 Authorized W&H Service organizations W&H Impex, 7475 Tecumseh RD.E., Windsor, Ontario, N8T 1G2, Phone: +1 / 519 / 944 67 39, Fax: +1 / 519 / 974 61 21, e-mail: admin@wnhimpex.com ADEC AUSTRALIA CO. INC., 41-43 Bowden Street, P.O. Box 71, Alexandria NSW 2015, Phone: +61 / 2 / 969 946 00, Fax: +61 / 2 / 969 947 00, E-mail: adec@adec.com.au SHINWON DENTAL CO., LTD., 158, Namdaemun-ro 5-ka Joong-ku, Seoul,...
  • Page 44 Distributed by Distributed in North and South America by Nobel Biocare AB Nobel Biocare USA, LLC, CA, USA Box 5190 SE-402 26 Göteborg SWEDEN Manufacturer ISO 13485, Annex II 93/42/EWG, EC 0297 W&H Dentalwerk Bürmoos GmbH Austria, A-5111 Bürmoos Postfach 1, Ignaz-Glaser-Straße 53 +43 / 6274 / 6236-0 +43 / 6274 / 6236-55 office@wh.com...

This manual is also suitable for:

Si-923Si-915

Table of Contents

Save PDF