Note
The system security pop-up screen appears only if the analyzer options are activated. Enter the log-in
name and password then press enter.
See Also
For information about creating passwords, refer to the Software Chapter.
6. The window closes and the main screen appears. See the illustration below for the
location of the menu bar.
7. Perform the following daily maintenance procedures. Refer to the Maintenance Chapter
for detailed procedures.
Check the sample and reagent syringes for leaks or condensation.
Fill the concentrated wash solution bottles.
displays on the screen if there is insufficient concentrated wash solution.
Check the sample probes, reagent probes and mix bars.
and clean. The mix bars should be clean with no chips in the Teflon. Select the following
menus: [Maintenance], [Analyzer Maintenance], H. Prime Washing-line. Press the
STAT/Rotation Diag. button. Verify that a thin, straight stream of diluent dispenses from
all probes. If the probes or mix bars appear dirty, clean them with an alcohol prep.
Verify that the printer "Ready" light is on and that there is sufficient printer paper.
Change DI water in the pre-dilution bottle.
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
OLU1010E
WARM UP
(R) Routine
(P) Parameters
(A) Auxiliary
Login
Login Name
Password
OLP1042
WARM UP
(R) Routine
(P) Parameters
(A) Auxiliary
(M)Maintenance
?
1997/09/03 10:10
(M)Maintenance
(U) User
(L) Logout
OK
Cancel
ALARMCLEAR
?
1997/09/03 10:10
(U) User
(L) Logout
Menu bar
ALARMCLEAR
Analysis pauses and a warning message
The probes should be straight
Chapter C - Basic Operations