To set up a custom protocol on the Beckman Coulter UniCel DxH 800, follow these steps:
1. Navigate to Menu > Worklist > Custom Tab > Advanced Search. 2. Select an option from the Report Transmitted drop-down list: - Equal To - Not Equal To 3. Choose one of the following statuses from the second drop-down list: - Preliminary Not Sent - Amended Not Sent - Final Not Sent - Preliminary Sending - Amended Sending - Final Sending - Preliminary Queued - Amended Queued - Final Queued - Preliminary Sent - Amended Sent - Final Sent 4. Click OK to save the configuration.
This process allows customization of worklist filters based on report transmission status.
This answer is automatically generated
Related Manuals for Beckman Coulter UniCel DxH 800
Summary of Contents for Beckman Coulter UniCel DxH 800
Page 1
® UniCel DxH 800 ® Coulter Cellular Analysis System Job Aids This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU). Information in the Instructions for Use Manual supersedes information in any other manual. PN: B02684AA (August 2011) Professional Development...
This equipment is used in a manner other than specified. Operate the instrument as instructed in the Product Manuals. • You introduce software that is not authorized by Beckman Coulter into your computer. Only operate your system’s computer with software authorized by Beckman Coulter. •...
Page 3
TRADEMARKS THE BECKMAN COULTER logo, COULTER, DxH, and UniCel are trademarks of Beckman Coulter; the BECKMAN COULTER logo COULTER, and UniCel are registered in the USPTO and SIPO. All other trademarks, service marks, products or services are trademarks or registered trademarks of their respective holders.
Page 4
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. PN: B02684AA (August 2011) Professional Development...
Page 5
UniCel® DxH 800 Coulter® Cellular Analysis System Log off / Log on Step Action To Log Off from the System Manager Select the Log off icon at the top right of the screen. On the Log off screen, select the Log off button. You are now logged off from the System Manager.
Page 6
UniCel® DxH 800 Coulter® Cellular Analysis System Access System Help Step Action Select the System HELP icon. You may search for topics using: “Content”, “Index” and “Search” tabs located on the left side of Help. To place topic in Favorites as a reference, select the Favorites tab. Locate a specific topic of interest.
UniCel® DxH 800 Coulter® Cellular Analysis System System Overview PN: B02684AA (August 2011) Professional Development...
Page 10
UniCel® DxH 800 Coulter® Cellular Analysis System Optional Floor Stand PN: B02684AA (August 2011) Professional Development...
Page 11
UniCel® DxH 800 Coulter® Cellular Analysis System Daily Checks Step Action Select the System Status icon . Look at the System Status screen to give you an overview of basic information about the instrument. Note the date and time of the last Daily Checks, Shutdown, Last Control (run) and Calibration.
Page 12
Step Action Look at the status in the upper left of the screen. Status may be Online, if running samples, Offline, for special cycles and procedures, Standby, for certain diagnostic tests. It may also state Calibration, Carryover or Repeatability when those procedures are run. Select the Daily Checks alert status icon A DxH 800 dialog box displays the following: Are you sure you want to start a Daily Checks procedure? Press “OK”...
Page 13
UniCel® DxH 800 Coulter® Cellular Analysis System Shutdown The DxH 800 allows you to program the length of time the instrument is in Shutdown. The default time is 30 minutes. This is the minimum time required for Shutdown. You may increase the time to up to 24 hours. At the end of your programmed time, the system removes DxH Cleaner and replaces it with DxH Diluent.
Page 14
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. PN: B02684AA (August 2011) Professional Development...
Page 15
UniCel® DxH 800 Coulter® Cellular Analysis System Quality Control Step Action The frequency of running controls is dependent upon regulatory and laboratory policy for quality assurance programs. Remove control tube(s) from the refrigerator and warm at AMBIENT TEMPERATURE for 10-15 minutes. After warming, mix by hand as follows: a.
Page 16
Reviewing Control Results Step Action Select the Quality Control status icon If any of the control runs had out-of-range values, a dialogue box opens with a list of those controls. If everything was within range, the Quality Control (Data View) screen opens with the last control that was viewed.
Page 17
UniCel® DxH 800 Coulter® Cellular Analysis System Supplies Step Action The hand-held Bar-Code scanner allows you to automatically upload DxH Supply information from the Setup Supplies dialog box. Select the Supplies icon from the Alert Status Icons. Select Setup to set up supplies. With the Setup Supplies dialog box open, scan the bar-codes on the Beckman Coulter reagent container and hit Enter.
Page 18
Configure Diluent Containers Step Action You can configure the system to run with one diluent container, or two (dual) diluent containers. The system will automatically switch between two containers if so configured. Select the Supplies Icon > Configure Containers. You may select either the Single Container or Dual Container option from the Diluent drop-down list.
Page 19
UniCel® DxH 800 Coulter® Cellular Analysis System Running Patient Samples Cassette Presentation (LIS Bidirectional Download to Pending) Step Action The SPM must be online to run samples. You can view the status of the SPM in the Status Mode area at the upper left hand corner of any screen. Ensure your specimens have been collected and stored properly.
Page 20
Single Tube Presentation Step Action Select the Single-tube Presentation icon The single-tube Presentation Dialog Box will come up. Place the specimen on the bar-code reader platform. If you type or scan the specimen identifier you need to press enter. Verify the Specimen Identifier and Test request. Follow the prompts on the screen. Mix the specimen according to your laboratory standards.
Page 21
Running a Sample in the Predilute Function Step Action The predilute function is a fixed x5 dilution. Obtain diluent from the instrument, using the Diluent Dispense, to use in making a dilution of a previously run sample. You will need an empty tube to obtain the diluent.
Page 22
Running Sample in Predilute Function (Cont.) Step Action If SPM timed out, select the Single-tube Presentation icon Scenario One Scenario Two Specimen ID in the Pending List Specimen ID is Not in the Pending List 1. Identify the specimen and press Enter. 1.
Page 23
Place the specimen into the correct Single-tube position. Once the sample is aspirated and the tube is returned. Select the button to finish. PN: B02684AA (August 2011) Professional Development...
Page 24
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. PN: B02684AA (August 2011) Professional Development...
Page 25
UniCel® DxH 800 Coulter® Cellular Analysis System Calibration Step Action NOTE Before you can start or restart the calibration process, the SPM must be offline. Select Menu>QA>CBC Calibration The screen will have the results of the last calibration run that was performed. Select Calibration Setup on the CBC Calibration Screen.
Page 26
• After replacement of any component that involves dilution characteristics (such as the BSV) or the primary measurements (such as the apertures). • When advised to do so by your Beckman Coulter Representative. • If you fail verify calibration procedure.
Page 27
Fitting #2 Fitting #1 BECKMAN COULTER, INC. URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER.
Page 28
UniCel® DxH 800 Coulter® Cellular Analysis System Front Area Air Mix Temperature Control (AMTC) Module Retic Clearing Diff Retic Stain NRBC Mixing Mixing Mixing Mixing Chamber Chamber Chamber Chamber Single Tube Presentation Barcode Camera Right Left Cradle Cradle PN: B02684AA (August 2011) Professional Development...
Page 29
UniCel® DxH 800 Coulter® Cellular Analysis System Right Side Area Pumps Sheath Regulator VCSn Waste Chamber Sample Regulator Sample Tank Sheath Tank Solenoids 25 psi Chamber PN: B02684AA (August 2011) Professional Development...
Page 30
UniCel® DxH 800 Coulter® Cellular Analysis System Left Side Area Pumps Bath Count Chamber Hemoglobin Lamp Hemoglobin Count Cuvette Chamber Bath Foam Trap PN: B02684AA (August 2011) Professional Development...
Page 31
UniCel® DxH 800 Coulter® Cellular Analysis System Multi-Transducer Module (MTM) Flow Cell PN: B02684AA (August 2011) Professional Development...
Page 32
UniCel® DxH 800 Coulter® Cellular Analysis System Cell Population NEUTS WBC’S PLT/DEBRI MATURE RBC’S RETICS NRBC PLT/DEBRIS PN: B02684AA (August 2011) Professional Development...
Need help?
Do you have a question about the UniCel DxH 800 and is the answer not in the manual?
Questions and answers
test principle
How to setup a Protocol not in the list
To set up a custom protocol on the Beckman Coulter UniCel DxH 800, follow these steps:
1. Navigate to Menu > Worklist > Custom Tab > Advanced Search.
2. Select an option from the Report Transmitted drop-down list:
- Equal To
- Not Equal To
3. Choose one of the following statuses from the second drop-down list:
- Preliminary Not Sent
- Amended Not Sent
- Final Not Sent
- Preliminary Sending
- Amended Sending
- Final Sending
- Preliminary Queued
- Amended Queued
- Final Queued
- Preliminary Sent
- Amended Sent
- Final Sent
4. Click OK to save the configuration.
This process allows customization of worklist filters based on report transmission status.
This answer is automatically generated