Cellular analysis system, hematology specimen processing module with system manager (666 pages)
Summary of Contents for Beckman Coulter AU2700
Page 1
Users Guide Vol. 1 AU2700 ® Chemistry Analyzer For In Vitro Diagnostic Use BM0561V1 NOV 2009 Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821...
Main Contents AU2700 User’s Guide Volume I: Precautions, System Outline, Basic Operations, and Software Document Change Notification Sheet ........Behind Cover Page Table of Contents Introduction ........................i Conventions for this Manual ......................ii Terms Used in This Manual ......................ii Notices on the Supplied Programs ..................... iii Trademarks ..........................
Page 4
Chapter C Basic Operations Introduction ......................C-1 1. General Operating Procedures ................C-3 2. Preparing for Analysis ..................C-7 2.1 Starting the AU2700.........................C-8 2.2 Starting the ISE ........................C-12 2.3 Performing Calibrations and a Reagent Blank...............C-15 2.3.1 Performing a Calibration and Reagent Blank on the Analyzer........C-16 2.3.2...
Page 6
6. User Defined Menus ..................D-197 Chapter E Software Menu Outline Routine Menu Outline ......................E-2 Parameter Menu Outline......................E-3 Auxiliary Menu Outline ......................E-4 Maintenance Menu Outline ......................E-5 List of Screen Operations ......................E-6 Automated Chemistry Analyzer Contents - AU2700 User's Guide May, 2001...
Beckman Coulter, or operators who have been trained by those who have taken the Beckman Coulter training course. Please read this manual carefully to understand instructions before operating the system.
Provides helpful tips on important information essential for understanding system operation and functions. See Also Gives reference locations for further information on a topic. Automated Chemistry Analyzer Introduction AU2700 User's Guide March, 2000...
Notices on the Supplied Programs • The CD-ROM and floppy disks that contain AU2700 software should be stored carefully in • appropriate place. • The programs may not be copied except for backup purposes. • The programs may not be subjected to any reverse engineering including decompiling, disassembling, etc.
See "What to do when service is needed." Beckman Coulter, at its sole discretion, will repair, replace, or adjust the defective Product, provided that Beckman Coulter investigation and inspection disclose that (i) such defect developed under normal and proper use and (ii) the Product is covered under this warranty.
Page 11
Representations and warranties made by any person, including dealers and representatives of Beckman Coulter, which are inconsistent or in conflict with the terms of this warranty, shall not be binding upon Beckman Coulter unless reduced to writing and approved by an express authorized officer of Beckman Coulter.
Introduction This chapter describes the precautions on system use, installation environments for proper operation, standard accessories, warning and caution labels. To use the AU2700 safely and to obtain the highest possible system performance, be sure to observe the precautions listed on the following pages.
• Please contact Beckman Coulter Technical Services, or the Sales Department when the system status is to be changed. Preventing Electric Shock To avoid an electric shock, observe the following precautions.
Page 15
Power outlet connections should be properly installed so as not to seriously affect other important equipment (e.g., surgical equipment power supply) or facilities due to system failure. Automated Chemistry Analyzer Chapter A - Precautions and Installation Environments AU2700 User's Guide March, 2000...
Page 16
If reagents or samples accidently come in contact with the mucous membranes or are accidently ingested, rinse immediately or induce vomiting. Follow the manufacturers or distributors instructions also located on the package insert, and contact a physician immediately. A-4a Automated Chemistry Analyzer AU2700 User's Guide November 1, 2009...
Page 17
This page intentionally left blank. A-4b Automated Chemistry Analyzer AU2700 User's Guide November 1, 2009...
Store samples according to your laboratory protocol. Improper storage of samples may compromise the integrity of the sample. Example: Refrigeration of blood increases potassium in the blood. Handling samples (Pretreating samples) • Fibrin suspended in the serum may clog the probe. Automated Chemistry Analyzer AU2700 User's Guide March, 2000...
Page 19
• To analyze serum, urine, or other samples using this system, the appropriate reagent, reference material, and control serum are required. For information about which products to use, consult the reagent manufacturer, the distributor, or Beckman Coulter Technical Services. •...
Page 20
To learn about total system operations, take the training course offered by Beckman Coulter. This manual is intended for personnel who have completed the training course offered by Beckman Coulter, or personnel who have been trained by someone who has taken the course.
Page 21
Never use the computer for a purpose other than that intended. Frequently scan diskettes or other media used in the AU2700 to avoid transmitting viruses. The danger of exposing the AU2700 computer to a software virus exists. A computer virus is malicious software, which is designed to intentionally destroy programs and data.
Page 22
After using the analyzer, observe the following precautions. • Follow proper shutdown procedures. Use caution when performing troubleshooting procedures. For assistance with troubleshooting, contact Beckman Coulter Technical Services. Do not modify the equipment. Inspecting and maintaining the system. • Be sure to perform scheduled analyzer maintenance.
The back of the system is warmer than the environment since the heat inside the system is ventilated from the back. Install the system where air circulates well. • The site should be well ventilated. Use ventilation equipment if necessary. • The system should not be exposed to direct air flow from air conditioners. A-10 Automated Chemistry Analyzer AU2700 User's Guide March, 2000...
Voltage: AC208 V ±10% (U.S.A) 60 Hz AC230 V ±10% (Europe) 50 Hz AC240 V ±10% (Australia) 50/60 Hz Power: 6 kVA [Maximum] Distribution panel: Within 30 feet of the system A-11 Automated Chemistry Analyzer AU2700 User's Guide March, 2000...
The grounding terminal should conform to class 3 grounding construction (less than 100 ohms of grounding resistance). Warning To be performed by Beckman Coulter personnel only: Install a grounding wire to prevent electric shock and to prevent system malfunction. OLU4056 Crimp terminal Hole diameter: Ø5.4 mm...
65 L/hour, and the maximum instantaneous water consumption, 3.8 L/minute. For information about requirements for a deionizer, refer to the quantity consumed and maximum consumption of the AU2700, described above in "Water Supply and Drainage." The purity of deionized water should meet the requirements of the College of American Pathologists (CAP) Type II, and its electric µ...
Page 27
Water supply hose (30’) Washing waste liquid hose (30’) Exhaust air hose (30’) Concentrated waste liquid hose (30’) See Also For more information on system specifications, refer to the "Specifications Chapter." A-14 Automated Chemistry Analyzer AU2700 User's Guide March, 2000...
60 mL Bottle Cap A Note Additional accessories are normally included at delivery of the instrument. Those items vary depending on the particular application intended for the instrument. Consult your Beckman Coulter contact about additional items. A-15 Automated Chemistry Analyzer...
Never remove the covers secured by screws, such as the rear cover and side covers, unless being guided by an Beckman Coulter Representative. • If liquid spills or leaks occur inside the system, contact Beckman Coulter Technical Services. Careless operation with liquid present may result in an electric shock. OLU4128...
Page 30
Do not place your hands or face close to this area. Instruction Labels: Provides information on the correct handling of the instrument. If instructions are not followed, system damage may occur because of incorrect handling. A-17 Automated Chemistry Analyzer AU2700 User's Guide March, 2000...
Page 31
DO NOT STARE INTO BEAM OR VIEW DO NOT STARE INTO BEAM OR VIEW DIRECTLY WITH OPTICAL INSTRUMENTS DIRECTLY WITH OPTICAL INSTRUMENTS STOP RESET CLASS 1 LASER PRODUCT complies with IEC60825-1 A-18 Automated Chemistry Analyzer AU2700 User's Guide March, 2000...
3.1 Keyboard Functions ....................B-10 3.2 Using the Mouse .....................B-13 3.3 Selecting Menus and Items ..................B-14 3.4 Help Functions ......................B-21 4. Login/Logout Function ................B-22 5. Description of Measure Modes ............B-23 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide January 20, 2006...
Outline of Functions Purpose of the System The AU2700 is a fully automated chemistry analyzer designed for operating ease, speed, and flexibility. It uses liquid, barcoded reagents, and user-friendly Microsoft Windows TM software. The operator is assured a fast turnaround time for response panels, maximum 2133 tests per hour with the ISE, 1600 tests per hour photometric.
Page 34
Batch Online The host computer and AU2700 must be online. Multiple sample requisitions are batch processed by the host computer through the operation of the AU2700 prior to sample analysis. Keyboard Entry Analysis is requisitioned by using the keyboard to enter the sample number or the barcode ID located on the sample cup.
Detergent tanks EM STOP button Reset button Printer Sampler Monitor Data processor DIAG button Photometer lamp Analyzer Power supply unit Breaker Ethernet cable Water supply/drain Sampler unit Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Page 36
Photometer Lamp The AU2700 has a polychrometer photometry system. A halogen lamp transmits white light through the cuvette. A diffraction grating splits the white light into 13 wavelengths. A silicone photodiode array detects the amount of light transmitted through the cuvettes. Each cuvette is read 28 times (P0 to P27) with 18 seconds between each read point.
Page 37
The personal computer is equipped with a large hard drive to store analysis data, RAM, a disk drive to use floppy disks, and a CD-ROM drive. The AU2700 uses software with user-friendly displays that make the analyzer easy to learn and operate.
Displays the system date and time controlled by the system clock. Menu bar Displays the six main menus: [Routine], [Parameters], [Auxiliary], [Maintenance], [User], and [Logout]. If these menu items are selected, a pull-down menu is displayed. Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Page 39
Dialog box Print Start Printing? Cancel Image Format Print Report No. Help Exit ALARMCLEAR The term "dialog box" is also referred to as "prompt." Example of a Window Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Page 40
Format Print Report No. ALARMCLEAR ALARMCLEAR Message display window Displays information about notification messages, etc. Help window Displays the description of alarms or operations if the Help icons are pressed. Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Performs an End Process and turns off system power, but not the reagent refrigerator or incubator. Cursor keys Arrow keys: Moves the cursor in the direction indicated by the arrow. B-10 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Page 42
For detailed information about the [System Status] key, refer to [System Status] in the Software Chapter. Function key Eight functions keys are displayed at the bottom of the screen. B-11 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Page 43
Each function displayed on the screen is executed by pressing the corresponding function key. For example, to close the currently displayed screen, press the F2 key, which corresponds to "Exit” on the screen. B-12 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
(P)Photocal Monitor (Q)QC Monitor (D)Data Management (L)Data Report (R)Report (D)Data List (N)Online Sub-menu Function ALARMCLEAR Alarm Clear : Selectable with the mouse or the optional touch panel. B-13 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Pressing the F2 key If function key F2 key is pressed, the current screen will close. The F2 key can only be used to close the current screen. B-14 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Page 46
Selected Selected item state state state G-GTP G-GTP G-GTP Cursor Move the cursor. Press the Space key. Press the ↓ key or Cursor the Enter key. B-15 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Page 47
This item is loaded. Cursor Type: SER. Type: SER. Type: URI. SER. URI. XXXXXXXXXX XXXXXXXXXX XXXXXXXXXX OTHER (Year/Month) (Year/Month) Press the Space key. Press the Enter key. B-16 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Page 48
ALARMCLEAR Deselected state Selected state Deselected state Multi check level Multi check level Multi check level Cursor Cursor Press the Space key. Press the Space key. B-17 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Page 49
.nnnnn .nnnnn ID(S) Serum Routine: Emergency: STAT: Urine Button Routine: Emergency: STAT: Others Routine: Emergency: STAT: Press the Enter key or click the left mouse button. B-18 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Page 50
Move the mouse to the button you want to select and click the left mouse button. STANDBY 1997/09/03 10:10 Report Format Common Sample Results(Fix) Results(Seq.) Comment Line Dialog box Print Start Printing? Cancel Image Format Print Report No. Help Exit ALARMCLEAR B-19 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Page 51
If a multi-digit number contains zeros in the first digits, they do not need to be entered. To enter “0001,” for example, type “1” and press the Enter key and “0001” will automatically appear. B-20 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Help Functions OLP1047 The AU2700 is equipped with an on-line help function. The operator can easily obtain information on alarms, operations, and software menus. Alarm Help icon Operation Help icon STANDBY 03/15/200010:10 (R) Routine (P) Parameters (A) Auxiliary (M) Maintenance...
Procedure for Executing Logout Select [Logout] when the menu bar is displayed. After logging out, the name and time are recorded in the alarm log file. B-22 Automated Chemistry Analyzer Chapter B - System Outline AU2700 User's Guide March 2000...
Description of Measure Modes AU2700 Measure Modes Measure 1 Sample analysis has started. Racks are moving from the rack supply belt to the sample transport belts. Measure 2 No more racks are detected on the sample transport belt. All R1 dispensing is being completed.
Chapter C Basic Operations Introduction ....................C-1 1. General Operating Procedures .............C-3 2. Preparing for Analysis ................C-7 2.1 Starting the AU2700 ....................C-8 2.2 Starting the ISE ......................C-12 2.3 Performing Calibrations and a Reagent Blank ............C-15 2.3.1 Performing a Calibration and Reagent Blank on the Analyzer........C-16 2.3.2...
Page 56
6.10 Programming the Auto Power On .................C-113 6.11 Creating a User Menu ...................C-114 6.12 Standby Mode .......................C-115 6.13 Programming a New Test ..................C-116 7. Operation Guide.................C-121 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide March 31 , 2005...
During operation of the AU2700, at least two levels of control material should be tested a minimum of once a day. In addition, these controls should be performed after calibration, with each new lot of reagents, and after specific maintenance or troubleshooting steps described in the Maintenance Chapter or Troubleshooting Chapter in this guide.
Page 58
During operation of the AU2700, at least two levels of control material should be tested a minimum of once a day. In addition, these controls should be performed after calibration, with each new lot of reagents, and after specific maintenance or troubleshooting steps described in the Maintenance Chapter or Troubleshooting Chapter in this guide.
Page 59
It is possible that the effects of preservatives on results may change over time. • Control sera recommended by Beckman Coulter are listed below. Changes in added components (especially preservatives) may occur in the future. Recommended Control Sera...
Page 60
If an error occurs while using the system, an alarm message is displayed on the monitor screen. Use the on-line help to assist in correcting the problem. Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Preparing for Analysis Before starting analysis, review the sections listed below: Preparations for Analysis Contents 2.1 Starting the AU2700 ....................C-8 2.2 Starting the ISE ......................C-12 2.3 Performing Calibrations and a Reagent Blank ............C-15 2.3.1 Performing a Calibration and Reagent Blank on the Analyzer........C-16 2.3.2...
The sub-power is turned on with the On button. The sub-power is turned off by pressing the End key on the keyboard. To start the AU2700, both the main and sub-power must be turned on. The main power should be on at all times except in an emergency! The main power supplies power for refrigerator, the incubator, and the automatic priming of the ISE.
Page 63
If the probes or mix bars appear dirty, clean them with an alcohol prep. • Verify that the printer “Ready” light is on and that there is sufficient printer paper. • Change DI water in the pre-dilution bottle. Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 64
Guide. OLU1013E Position Display Type: Pos. Test Shot Onboard Cal. Expired Lot Bottle Seq. Comment (mL) Stability Stability aaaaa 9999 999 R.ID Read Error Close C-10 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 65
12. Press function key F2 (exit) key. The system status screen appears. Press function key F2 (exit) again. See Also For detailed information on starting analysis, refer to the section in this chapter called “Starting Analysis.” C-11 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
3. Verify the ISE buffer syringe is not leaking and that condensation is not visible. OLU2065 I S E ISE power switch Fixing screws Case head Syringe case : Possible leakage locations C-12 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 67
(small) Sample cup STAT table STAT ROTATION/DIAG switch Top view of STAT table D. Replace the small STAT table cover. E. Select [System Status], [ISE Status]. C-13 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 68
H. After checking results, press function key F2 (exit). The main screen returns. See Also For calibration errors, refer to the ISE section of the Troubleshooting Chapter. For information about replacing electrodes, refer to the Maintenance Chapter. C-14 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Then use the ▼ at “Type” to select urine. Add a ü to “Use serum cal,” then click on the box with the mouse. This feature is required for urine tests using the pre-dilution feature: • urine urea • urine uric acid • urine phosphorus C-15 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide February 4, 2004...
8. Exit the [Calibration] screen. 9. First place the reagent blank rack, then the calibration rack on the analyzer. 10. Close the cover, then press Start. C-16 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide February 4, 2004...
Reagent Blank on the Analyzer” listed on the previous page. Place the calibrator with the appropriate barcode ID in any position on the yellow rack and start analysis. Note To change calibrators, select function key F5 (Cal) C-17 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
If the STAT LED is off, press the STAT button twice to start. If errors are present, an error list will appear. Review the error list and correct any problems before starting analysis. C-18 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
4. Press function key F2 (exit). Verify the information entered in this screen. 5. Press function key F2 (exit) again to return to the main menu. C-19 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
C. Use the pull-down menu from the “Position No.” prompt to set all positions. D. Select the “Close” button. 7. Select the “Test” tab. 8. Select the test to program for auto calibration. 9. Press function key F5 (edit). C-20 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 75
17. Press function key F2 (exit) again to return to the main menu. 18. Verify programming by selecting [System Status], [STAT Table Status]. The calibrator numbers display in assigned yellow positions. Function key F6 (cal req) is gray and not operational. C-21 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
QC names and assigned positions. 7. Load the QC tubes in the correct green rack using the Detail Inf. Button. 8. Press the Start Key or icon. C-22 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide March 31 , 2005...
7. Select function key F2 (exit) twice to return to the main menu. 8. Set sample containers on a green rack in any order. 9. Place racks on the analyzer. 10. Close the cover. 11. Start analysis. C-23 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
If the STAT LED is off, press the STAT button twice to start. If errors are present, an error list will appear. Review the error list and correct any problems before starting analysis. C-24 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
QC in [System Maintenance] and cups are not detected. 4. Reagent blank, calibration and QC are performed automatically based on programming from the [Parameters] menu. C-25 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
19. Verify the programming by selecting [System Status], [STAT Table Status]. The QC numbers display in the assigned positions. Function key F7 (QC Req.) is gray and not operational. C-26 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
If normal analysis is being performed using the ISE unit, select the ISE tests (Na, K, Cl) or requisitions can be downloaded from a host computer to the AU2700. Profiles can be created to requisition a common group of tests at once.
3. Verify the sample number is correct for the next sample number to be processed on the AU2700. If you are processing emergency samples: Type an “E” to test emergency samples in the red rack or type a “P” to test emergency samples on the STAT table. Use function key F8 (S.
Page 83
Month Age: Comment(C) Help Exit Start Entry Test Total S.No. Search S.ID ALARMCLEAR 10. Press function key F2 (exit) twice to return to the main menu. C-29 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
3. Verify the sample number is correct for the next sample number to be processed on the AU2700. If you are processing emergency samples: Type an “E” to test emergency samples in the red rack or type a “P” to test emergency samples on the STAT table. Use function key F8 (S. No.) to select a specific sample number.
3. Verify the sample number is correct for the next sample number to be processed on the AU2700. If you are processing emergency samples: Type an “E” to test emergency samples in the red rack or type a “P” to test emergency samples on the STAT table. Use function key F8 (S. No.) to select a specific sample number.
B. From the “Type” pull-down menu select: serum, urine or other. C. Verify the sample number is correct for the next sample number to be processed on the AU2700. Type a “P” to test emergency samples on the STAT table. Use function key F8 (S. No.) to select a specific sample number.
Page 87
Condensation may be generated under high-temperature and high- humidity environmental. Remove all samples from the STAT table when the END LED is lit before placing new samples on the STAT table. C-33 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
ID field, then select the desired tests following test requisition. The system will automatically process the barcoded sample with the revised requisition. C-34 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
ID field, then select the desired tests following test requisition. The system will automatically process the barcoded sample with the revised requisition. C-35 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide June 16, 2006...
31-40 (Etc. up to Rack ID 400) For all methods, all of the sample containers must have a requisition, or the default profile (#0) will be performed. C-36 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide June 16, 2006...
Page 91
IDs 1, 3 and 5 and analyzed, the assigned sample numbers (requisitioned with an E prefix) are E001, E002 and E003 sequentially. If the sample containers are barcoded, they can be placed in any position on the rack. C-37 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 92
Orient the label so the numbers are located to the left if viewed from the front . Rack side Protruding part of the rack C-38 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 93
The label should not be inserted into the rack. Up to 36mm Rack Note Nested cups used in analysis on the AU2700 should meet the specifications for tubes and micro-cups, See Chapter I - Analyzer Specifications for Sample Cups and Tubes. C-39 Automated Chemistry Analyzer Chapter C - Basic Operations...
Page 94
• The following should not be present: Scratch with a diameter greater than 0.1 mm, omitted pattern, blur, or void (black point) on white bars. C-40 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide July 21, 2006...
Page 96
Serum, urine, or other samples must be requisitioned separately and placed in racks with the correct rack ID. The rack ID settings can be selected in [Parameters], [Online]. C-42 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 97
Instruction If a sample is centrifuged in a primary tube to be analyzed on the AU2700 system, ensure that the serum is adequately separated from the red blood cells and does not contain clots or fibrin, and that the amount is sufficient for sample detection and analysis.
Page 98
Moves to Analysis 2 mode Sample Nos. determined automatically 1 2 3 4 5 6 7 8 9 11 12 13 White rack No change in mode C-44 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 99
The inclination angle must be 5° or less. Sample cup The label should not rise above the container. Label Barcode The label should not be inserted into the rack. Up to 36mm Rack C-45 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 100
CODE128, 2 of 5 standard, 2 of 5 interleaved, mixed, ISBT-128 (alone). If “mixed” is selected, analysis can be performed with multiple types of barcodes being mixed (except ISBT-128). C-46 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 101
G (gap) can be identified easily. If less than 0.2 mm of NB (narrow bar) is used, it is recommended to use the barcode check character to retain the reliability of the barcode read. C-47 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 102
6. If the SET LED and END LED turn on, remove the STAT table cover (small). 7. Take out the emergency sample cup where the sample aspiration is complete. 8. Remount the STAT table cover (small). C-48 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 103
The sample cup should be placed in an outer hole on the STAT table with the barcode facing outward. STAT table C-49 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 104
13.6 mm. When using containers with a diameter of 14.0 to 15.0 mm, the operator will need to determine if an adapter is necessary since plastic sample containers may be tapered and the diameter may be difficult to judge. C-50 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 105
2. When the upper edge of the adapter comes out from the rack, grasp the adapter and pull it out. Adapter lock Rack window Push up lightly with a finger C-51 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Top view of rack feeder Surface where the rack ID label is applied A maximum of 30 racks can be set at one time. Window for reading rack ID C-52 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 107
If emergency samples are to be processed, they can be run anytime during routine analysis by inserting red racks ahead of white racks. Tips A maximum of 30 racks can be set at one time. C-53 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Clean the Cuvette Wash Nozzle. • Clean the Deionized-Water tank. • Clean the sample and reagent probe wash wells. • Clean the Deionized-water and the sample probe filters. C-54 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide ????, 2006...
Page 109
Replace the Na, K, and Cl electrodes. • Replace the ISE reagent (buffer) syringe. • ISE Cleaning (Manual) procedure. See Also For detailed information on performing Maintenance, refer to the Maintenance Chapter. C-55 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide ????, 2006...
The standards listed in the table above are values calculated under the assumption that normal, routine work is being performed on the analyzer. C-56 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide ????, 2006...
30 mL reagent bottle 15 mL reagent bottle Beckman Coulter reagents have a reagent ID. Bottles can also be used without a reagent ID. Load Reagent Bottles The first and second reagent refrigerator can each hold 48 reagent bottles maximum.
Page 112
Replace old bottles with new ones if the reagent level is low, or if the reagent stability or expiration date has expired. Warning It is imperative that the AU2700 reagent bottles are placed correctly in the refrigerator. If the end of a bottle is up, reagent probe problems could occur. •...
Page 113
If the bottle is not detected correctly, apply a label to the reagent bottle as shown below. If stains or condensation occur on the label, a barcode read error may occur. Reagent bottle Label C-59 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 114
When storing the reagent bottle, remove the insertion pipe. If the bottle is stored with the pipe inserted, the reagent will be deposited between the pipe and opening causing instability in reagent. Pouring opening Dispensing opening Scale Insertion pipe C-60 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Display, Detail button, and review the comments. For more information on each comment refer to “Reviewing Reagent Position Errors,” located in [System Status], [Reagent Status] of the Software Chapter. C-61 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide March 31, 2005...
Also, if any wash solution gets in your eyes or mouth, rinse it out immediately and consult a physician. See Also For detailed procedures on adding wash solution, refer to the Maintenance Chapter. C-62 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide March 31 , 2005...
Setting portion of sample prior-diluent bottle Perform the following steps if the analyzer has not been set up to run automatic pre-dilutions: 4. Select [Parameters], [Specific Test Parameters]. C-63 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 118
For reference, the main sample pre-dil. rate and the required original sample quantity for each are given in the following table. ample ample ample dilutinG otal rate quantity µ water quantity µ µ C-64 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Preparing for Analysis section of this chapter. 4. Press function key F2 (exit). The main screen is displayed. 5. Press the start key. The window for starting analysis is displayed. C-65 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 120
8. After performing the above check, select “Yes” and press enter. The analysis operation starts. After “STANDBY → MEAS. 1” is displayed on the upper left of the screen, “MEAS. 1” is displayed. After starting analysis, the first result is calculated. Results print as they are completed for the sample. C-66 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
4.6 Printing Edited or Corrected Data ................C-88 4.7 How to Reprint Reagent Blank, Calibrator, QC and Patient Data ......C-89 4.8 Saving Parameters and Data to a Floppy Disk ............C-90 C-67 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide March, 2003...
1. Look at the printout and check it for error flags. If error flags appear on the printout, perform the corrective action corresponding to that flag For a detailed description of the printout, please refer to your AU2700 Training Manual, in the ‘Modes of Data Interpretation’ Section.
“OK.” 3. Select the test(s) to be referenced using the mouse. 4. Press function key F6 (daily data). The chart with daily variation data is displayed. C-69 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 124
8. Optional Step: Select function key F4 (comment) if you want to enter a comment. 9. After checking the chart, press function key F2 (exit). The main screen returns. C-70 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 125
Daily Data Change Axis Control/Test Control/Test ALARMCLEAR 5. Check the chart for abnormal data. See Also For information about handling errors, refer to the Troubleshooting Chapter. C-71 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 126
See Also For information about shifts and trends, refer to the Troubleshooting Chapter. C-72 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 127
After entering a comment, select the “close” button. 7. After checking the chart, press function key F2 (exit) until the main screen returns. C-73 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Perform a repeat run manually according to the contents of that file. Important The following repeat run procedures also explain how to program a repeat run. If the analyzer has been preprogrammed, refer to the AU2700 Quick Response Guide for basic repeat run procedures. C-74 Automated Chemistry Analyzer...
7. Check the contents of the repeat-run work list. 8. Use the worklist to perform a manual repeat, or modify the worklist before performing repeat analysis. C-75 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
9. Depending on the preferred repeat-run method, use one of the procedures on the following pages. If changes are made to the repeat run requisitions, always print an updated repeat run work list. C-76 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
How to Perform an Auto-Repeat Run The following procedure explains how to program and perform an auto-repeat run. If the system has been pre-programmed to perform an auto repeat run, follow the procedures in the AU2700 Quick Response Guide. The auto-repeat method involves pre-programming flags which will be used to generate repeat run requisitions.
Page 132
STAT serum, urine or other repeat. (If no barcode is used, place in numerical order and remove when finished or concordance errors can occur.) C-78 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide June 16, 2006...
The following procedure explains how to program and perform a manual repeat run. If the system has been preprogrammed to perform a manual repeat run, follow the procedures in the AU2700 Quick Response Guide. Any of the flags on the [Repeat Run Common] menu can be selected for repeat run analysis. J and K flags allow specific test ranges to be entered.
To quickly perform repeat runs on a small number of samples, use the STAT table. The following procedure explains how to program and perform a manual repeat run. If the system has been preprogrammed to perform a manual repeat run, follow the procedures in the AU2700 Quick Response Guide.
Page 135
STAT table after being requisitioned for STAT table analysis. Press the STAT button. See Also For detailed information on using the STAT table, refer to the section called “Using the STAT Table: Normal Analysis Mode.” C-81 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
The “S” flag will display on reprinted data from the AU2700, or data that is batch transferred to a host computer. There are two main fields in [Parameters] that affect final results: [Parameters], [Repeat Parameters], [Repeat Common], “Repeat Data Overwrite”:...
Page 137
(rewrite). The data should now have an “S” flag to indicate the test was repeated, and the original result overwritten with the repeat result. See Also For more information about error flags, refer to the Error Flags Chapter. C-83 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Enter the test number and click the “OK” button. 5. Press function key F5 (data display). The analysis result data displays for tests or samples. C-84 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 139
Press F4 Reagent INF button to display the R1 and R2 lot nmber and bottle number used on the test where the cursor is located. 8. Select function key F2 (exit). 9. Press function key F2 (exit) again. The main screen returns. C-85 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide February 4, 2004...
6. Enter correction factors A and B. Click the “OK” button. The correction is executed. 7. After the correction has finished, press function key F2 (exit). The main screen returns. C-86 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide February 4, 2004...
4. A window appears to select “Resume Printing” or “Cancel Print.” 5. All data will print if “Resume Printing” is selected. C-88 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
The screen will default to the “FD Data Save” tab. 2. From the “Index” field, use the pull-down menu to select the index that will be saved. C-89 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 144
5. Insert the floppy disk in the computer. Select function key F5: Start Save. From the “Insert FD Start” prompt, select yes. The message “Operating: Wait for a while” displays. C-90 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide November 1, 2009...
The reagent refrigerators and incubator require power to be maintained. The ISE unit also requires power to perform an automatic, periodic priming cycle. Contents Ending ISE Analysis ...................C-92 Ending Analyzer Analysis ................C-92 C-91 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 146
Auto ON: with auto preparation: Month Date: Hour Time: Set(S) End(E) NEXT ALARMCLEAR See Also For detailed information about performing a W2, refer to the Maintenance Chapter. C-92 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
6.9 Turning On/Off the Sample ID Barcode Reader: ..........C-112 6.10 Programming the Auto Power On .................C-113 6.11 Creating a User Menu ...................C-114 6.12 Standby Mode .......................C-115 6.13 Programming a New Test ..................C-116 C-93 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Stopping the Rack Feed Unit During analysis, to insert a rack for an emergency sample, use the Feeder Stop key on the keyboard or icon on screen. C-94 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
(6)DPR Status common alarms, then use corrective actions found in the on-line help function. (7)STAT Table Status Measure S.No.: STAT Cal./QC Status Online S.No. C-95 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide Input May, 2002 Output Calibration:...
Page 150
QC and calibration status performed on the STAT table and the type of operation to perform. The STAT LED light status is also displayed. C-96 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
If many emergency samples require analysis, use the red rack. Instruction If a sample is centrifuged in a primary tube to be analyzed on the AU2700 system, check that there is sufficient serum for analysis, in addition to the dead volume required for sample detection. If the serum height in the sample tube is less than the limit, transfer the serum into a smaller sample container, then start analysis.
5. To use the one touch option, remain in this menu. To end the one touch mode, select function key F2 (exit). 6. When the prompt “End One Touch Mode” appears, select “Yes.” C-98 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 153
3. Each time a STAT sample is performed, remove the current sample from position 1, and replace it with the next sample. Press the STAT Button. C-99 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
“P” category. If condensation appears on the STAT table or sample containers, be sure to wipe it off with a clean cloth. C-100 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 155
7. Press the STAT Button. Analysis starts. 8. The END LED comes on when STAT sample aspiration is complete. Remove containers from the STAT table. C-101 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 156
Do not place the barcode label below the holes on the STAT table. This may cause barcode read errors. The barcode label should be applied to the sample container as shown in the figure. C-102 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
5. Enter sample ID, sex, age and profile if required. 6. Select the tests to be performed on the sample. 7. Press function key F4 (entry). The next sample number is available for requisition. C-103 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 158
11. Press start to begin or resume analysis. See Also For information on how to start analysis, refer to “Starting Analysis” located in this Chapter. C-104 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
“no cup” alarms are generated and requisitions will not be performed. This could lead to concordance errors. 5. Press start. C-105 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide June 16, 2006...
Page 160
Also the operating system and application files could become corrupted. The following steps are the preferred method of shutting down the AU2700 during an emergency: 1. Press the following keys on the keyboard: CTRL + ALT + Delete. The Task Manager will appear.
A reagent check is the only procedure requiring the 14 minute wait. After adding Wash Solution or removing the racks from the collection area, press START to resume analysis. C-107 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide June 16, 2006...
2. Start analysis. Analysis begins with the first rack in the rack feeder. See Also For information on how to start analysis, refer to Chapter C, “Starting Analysis.” C-108 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
8. Optional: Select function key F3 (print) for a list of the profile no., name, and tests selected in that profile. 9. Select function key F2 (exit) again to return to the main menu. C-109 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
5. Optional: Select function key F3 (print) to print a list of the new mean, SD, and range. Example: The Beckman Coulter Control package insert lists the mean as 120, and the range as 115 to 125. You would type in 120 as the mean, 10 as the range and 2.5 as the SD.
3. Select function key F4 (set). Enter the new calibrator concentration at the “CONC” field from the package insert. 4. Select function key F2 (exit) to set the new concentration. C-111 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
STAT field. 3. Select “Sequential” to turn off the barcode reader or select “Barcode” to turn on the barcode reader. C-112 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
4. When the End key is selected, the next “On” time will be displayed. If you want to change the next “On” time, select the “set” button. C-113 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
4. Repeat steps 2 and 3 until all the screens you want to add are part of the User Menu. Press function key F2 (exit). 5. When [User Menu] is selected from the Main Menu, a pull-down menu of the choices selected will appear. C-114 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
2. Select [Auxiliary], [Standby Set]. If the analyzer is currently in Stop, select the Stop key again. The analyzer can be initialized and returned to the previous mode (Warm-Up or Standby). C-115 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
The order in which the tests are added to the round is the order they will print from the data log. D. Press function key F2 (exit) twice. C-116 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide July 7, 2009...
Page 171
Click on the “close” button. Example: The Beckman Coulter Control package insert lists the mean as 120, and the range as 115 to 125. You would type in 120 as the mean, 10 as the range and 2.5 as the SD.
Page 172
F. Select function key F4 (set). Refer to the methodology sheet and set the Calibration type, Formula, and MB type factor. G. Press function key F2 (exit) twice. The main menu returns. C-118 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Page 173
G. Select the test arrangement by using the pull-down menu next to the “Test Arrangement” field. H. Select the test(s) to be added to the list. I. Press function key F2 (exit) twice. The main menu returns. C-119 Automated Chemistry Analyzer Chapter C - Basic Operations AU2700 User's Guide May, 2002...
Using a worksheet created by an external device, and using the Requisition operation for normal analysis [Normal] test requisition menu: See "C-22, D-57". Create a worksheet using AU2700. Set on the rack feeder. Set the samples. Set on the STAT table.
Page 176
3.10.2 System ........................D-148 3.10.3 Auto Power On ......................D-150 3.10.4 Password .........................D-151 3.10.5 Login Name & Menu Level ..................D-152 3.10.6 User Menu .......................D-154 3.11 Comment Masters ....................D-156 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide March 31, 2005...
Page 177
5.5 Maker Maintenance ..................... D-186 5.5.1 Program Version ......................D-186 5.5.2 Analyzer Diagnostics ....................D-187 5.5.3 ISE Diagnostics .......................D-192 5.5.4 Alarm Log ........................D-195 5.5.5 File Management .....................D-196 6. User Defined Menus ................D-197 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide March 31, 2005...
Page 178
Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide March 31, 2005...
1.5 Analyzer (ANL) Status ....................D-19 1.6 Cuvette Status ......................D-22 1.7 Data Display ......................D-24 1.8 Alarm List ........................D-26 1.9 Data Processor (DPR) Status .................D-28 1.10 STAT Table Status ....................D-29 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Error occurred and analysis is not possible Note The ISE Status turns red and displays “Busy” during ISE Calibration and cleaning. This is normal and not cause for alarm. Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
For procedures, refer to the Maintenance Chapter, “Perform a Photocal.” Shift to the One Touch Mode: For procedures, refer to the Basic Operations Chapter “Using the STAT Table: One Touch Mode.” Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
2. Select either “R1” or “R2” from the pull-down menu. 3. Enter the position number for the reagent on which you want more information. 4. Select “OK.” Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Also, it does not detect a bottle placed in the wrong position. Illustration of Reagent Refrigerators The reagent refrigerators are illustrated on the left side of the screen. This illustration shows the reagent type and status. Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
(tests), and remaining volume: 1. Select [System Status], [Reagent Status]. See Also For information about troubleshooting reagents and samples, refer to the Troubleshooting Chapter “Troubleshooting Reagents and Samples.” D-10 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Page 185
-F5 (Check Start) Perform a reagent volume check: For information about performing a reagent volume check, refer to “Starting the AU2700” located in the Basic Operations Chapter. F6 (Edit) Set (Fix) a reagent position: For procedures, refer to the Basic Operations Chapter “How to Set (fix) a Reagent Position.”...
If you do not want a bottle in this position, click on that position, select function key F6 (Edit) and use the pull-down menu at the test prompt to select “None.” D-12 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide March 31, 2005...
Fix the reagent position and edit the lot number. Perform a reagent volume check. If the test is R1 only, and the lot number is not entered, results print in D-13 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide March 31, 2005...
(not empty). For more information on the multi-reagent switch feature, refer to step 1 in the procedure called “Programming a New Test” located in the As-Needed Tasks section of the Basic Operations Chapter. D-14 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide March 31, 2005...
Electrode indicates whether the latest slope value falls within the range of normal values. essage eaning hange in baCkground Color Electrode Slope values in range Blue Electrode Slope out of range Yellow D-15 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Displays whether the slope is in range or out of range. ISE Reagent Status Indicates the status of the buffer, mid-standard, and reference solution. Disp. Range Shows the display range with start and end dates. Max Slope D-16 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Page 191
Clean the Mix Bar, Liquid Level Sensors and Sample Pot: For procedures, refer to the Maintenance Chapter “Clean the Mix Bar, Liquid Level Sensors and Sample Pot.” D-17 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
B/P Tube Exchange Prime • C/Total (All) Prime • D/Buffer Prime • E/Mid/Ref Prime • F/Prime Bypass 4. Press the STAT ROTATION/DIAG Button. The prime operation starts. D-18 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
AUTO (Ready) Analysis is possible. Blue STOP Analysis is not possible. Red or Yellow STANDBY Analyzer is standing by. Blue MEASURE Analysis in progress. W1 in progress. D-19 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Sample No. at reagent 2 dispense position. Rack: Indicates the rack feeder status. eaning hange in baCkground Color Rack tray is full or jammed Not Full Blue STAT Table Status: Cover Status D-20 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
READY Analysis is possible. Blue COVER ISE cover is open BUSY Analysis is in progress. STOP ISE is in stop mode Yellow No Connection ISE disconnected Yellow D-21 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Note: The mean check range 0.03 has to be set for both the inner and outer wheel individually. Absolute Check Range: Not performed in Beckman Coulter maintenance protocol. Select 6.5 to maximize the range. Note: The absolute check range has to be set for both the inner and outer wheel individually. D-22 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
F8: Save Reference: Used to save the most recent data, after performing a cuvette check, as reference data. 1. Press function key F8 (save reference) and select “Yes. D-23 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
“BB” in the figure will be substituted by the following symbols that indicate each sample type. Sample data prefixes: H: Repeat-run sample U: Urine sample E: Emergency sample X: Other sample P: Priority Stat sample No prefix: Serum sample (routine) D-24 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
F5 (measured data). This list displays a maximum of 1,000 lines of data starting with the most current. If no relevant data exists, a message indicating no data is displayed. D-25 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
1. In the [Alarm list] screen press function key F3 (print). A window to select the print range is displayed. 2. Select either “Displayed Date” or “All Date” and select “OK.” Printing begins. To stop printing, select “CANCEL.” D-26 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
F5: Re-Display: To display the most current alarms, select function key F5 (re-display). F6: Display Help: To display the alarm help screen, select function key F6 (display help). D-27 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
While in real-time printing, “Real-time printing” displays. While in report printing, “Report printing” displays. In other printing states, “Batch printing” displays. Sample number The sample number printed displays while in real-time printing or report printing. D-28 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Indicates the status of the STAT table sample aspiration. essage eaning hange in baCkground Color Sample aspiration complete Blue Sample aspiration incomplete Grey D-29 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Page 204
For procedures, refer to the “Performing Calibrations and a Reagent Blank” section of the Basic Operations Chapter. F7: QC Req. (Requisition QC STAT analysis): For procedures, refer to the “Quality Control Analysis” section of the Basic Operations Chapter. D-30 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Data Edit ........................D-75 2.7.2 Data Correction ......................D-77 2.7.3 Repeat Data Verification ....................D-80 2.8 Data Report ......................D-81 2.8.1 Report ........................D-81 2.8.2 Data List ........................D-83 2.8.3 Online ........................D-84 D-31 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide March 31, 2005...
Urine Start No. Urine start numbers can be entered for each of the samples classified as Routine, Emergency, and STAT. They can be selected for normal and repeat-run analysis. D-32 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Use this software screen to: Start Analysis: Refer to the Basic Operations Chapter “Starting the AU2700” for detailed procedures. Change start conditions: A data index can hold up to 8,000 samples. Three rounds of tests can be programmed, with up to 51 tests.
After pressing function key F8 (inventory check), the inventory check dialog box displays the projected amount of reagent that will be used. After requisitioning a test, this screen will show how much reagent will be used in mLs. D-34 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide March 31, 2005...
Routine samples must have a requisition for the tests that will be run on each sample. Specifiying the profile Deleting the sample Requisitions can be performed on the AU2700 in this screen, or downloaded from a host Profile information computer to the AU2700. Requisitions are processed sequentially by sample number or by the previously registered sample ID number.
F3 Print (Print a pending list or work list): A pending list prints samples not yet run. A work list prints all tests requisitioned. 1. Select [Routine], [Test Requisition], [Normal]. D-36 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Click or tab in the next column and enter the ending sample number to be deleted. If an asterisk (*) is entered, all sample numbers or IDs will be deleted. 5. Click on the delete button using your mouse. D-37 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Page 212
C. Select “OK” to save changes. 4. After the entry is completed, press function key F4 (entry). The setting contents are saved and the next sample number will be displayed. D-38 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
< and > buttons: Switches the sample number being displayed. The keyboard can also be used. Shift + Page Down = < button, Shift + Page Up = > button. D-39 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
003405 H0003 0005 003405 H0002 0008 003408 H0004 0008 003408 H0003 List of Rep: Use this button to display a list of samples requiring repeat run analysis. D-40 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
If this analyzer is connected to a host computer, the F7: Online function key can be used to receive requisitions from the host computer. F8: S. No.: Use this function key to select the sample number. D-41 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Chapter C, Basic Operations. F7: Change Dil: Use this function key to change the repeat run type from the following options: normal, dilution, or condense. D-42 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Requisition screen has priority over the main screen. If a test is selected from the main screen, the system requisitions only bottle sequence 1. All for one test button: On the Individual Requisition screen, requisitions all sequenced bottles of the selected test. D-43 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Use this function key to view QC settings and rack positions. Select the Detail button to display the QC rack ID number, rack position, QC number, and QC name. F8: Deselect All Tests: Deselect all QC requisitions for all tests. D-44 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Analyzer Requisition: Use this button to perform an automatic calibration requisition for tests with expired cal or new bottles/lots using Advanced calibration. D-45 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Use this function key to view calibration settings and rack positions. Select the Detail button to display the calibrator rack ID number, rack position, calibrator number, and calibrator name. D-46 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Sample No.: Enter the sample number range in this field. Enter an asterisk (*) to designate all samples. Sample ID: Enter the sample ID range according to the range set in [Parameters], [Requisition Format]. Enter an asterisk (*) to designate all sample IDs D-47 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
2. Press function key F6 (file initialize). A window to start initialization is displayed. 3. Select “OK.” Initialization starts. See Also For procedures, refer to “Performing a Repeat Run” located in the Basic Operations Chapter. D-48 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Sample ID: Enter the sample ID range according to the range set in [Parameters], [Requisition Format]. Enter an asterisk (*) to designate all sample IDs and enter a space for no specification. D-49 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
STAT. 4. Move the cursor to the “Sample No.” column and enter a range of sample numbers for analysis. A range of sample IDs can be entered instead. Enter an asterisk (*) for an item which will not be specified. D-50 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
QC, Cal./Count - right box: the cases of calibration samples or reagent blank samples, the cuvette count will be displayed in this box. D-51 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
If the corresponding Photocal data does not exist, this box remains void. Index: The corresponding index date and time will be displayed. Data: mmmm nnnn mmmm: Data being displayed D-52 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Search for an analysis test: A. In the [Reaction Monitor] screen press function key F5 (search data), press function key F8 (test number). The cursor moves to the test number column. D-53 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
“Auto” is selected, the largest possible graph will be displayed within an area in which all data can be contained. Graph Display: The X axis displays the measuring point. The Y axis displays the measured OD value. D-54 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
1. Select the test for which calibration data is to be viewed and perform the following: F5 (Recal) Recalculate the calibration factor. F3 (Print) Prints a list of calibration information. F4 (Comment) Add a comment to the list print. D-55 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
5. Select “OK.” The result of recalculation is saved and the calibration is updated. If you select “Cancel” here, the recalculation will not be saved D-56 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Calibration data of only the currently displayed analysis tests will be printed. All tests: All the calibration data of all analysis tests will be printed. 3. Select “OK.” Printing begins. D-57 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
For procedures, refer to “How to reprint reagent blank, calibrator, QC and patient data” located in the Basic Operations Chapter. Type: The sample type can be changed to serum, urine, or other. Select the sample type from the pull-down menu. D-58 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Display comments: Entered comments can be reviewed as necessary. Press function key F5 (ref. comment). Comments will be displayed in the window. To exit select “Close.” D-59 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Use the pull-down menu located by the "RB Measured Date" field. Display a history of reagent blank data: Press function key F6 (history) to graphically display the last 30 reagent blanks. D-60 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Prints the data of all tests. Data only: Prints data only. Comments won’t be printed. Data + Comment: Prints the data and comment at the same time. 3. Select “OK.” Printing begins. D-61 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
2. Select a cuvette number. 3. Select function key F5 (history graph) to display data in a graph or F6 (history data) to display data in a list. D-62 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
1. Press function key F3 (print). A window to specify the printing mode is displayed. 2. To print currently displayed data, select “Display Cuvette.” To print all cuvette data, select “All cuvettes” and select "OK." Printing begins. D-63 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
See Also For information about [QC Control], see [Quality Control] located in the parameters section of this chapter, or refer to the QC section of the Specifications Chapter. D-64 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Index: Select the index (date/time) from the pull-down menu. Type: The sample type can be changed to serum, urine, or other. Select the sample type from the pull-down menu. D-65 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
2. Select the tests displayed on screen using your mouse. 3. Press function key F6 (daily data). 4. Press function key F8 (change axis). A window to set the display condition appears. D-66 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Prints a graph and comment. Comment: All Data + Prints the numerical data and comment. Comment: 5. Select “OK.” Printing Begins 6. Press function key F2 (exit) twice. D-67 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Display and print day to day QC charts: 1. Select [Routine], [QC Monitor], [Day to Day Control]. 2. Set the index using the pull-down menu next to the "index" field. D-68 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
1-30 for serum, 31-45 for urine, and 46-60 for others. These settings are user defined and can be changed by authorized personnel. 3. Select "OK." D-69 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Type: The sample type can be changed to serum, urine, or other. Select the sample type from the pull-down menu. Test - List: Use the mouse to deselect and select tests. D-70 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
4. Select function key F6 (Twin Plot Display). 5. Select function key F7 (Daily Data). Statistics for the control number selected will display. D-71 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
QC Sample No.: Set the sample number range within the range specified in parenthesis. Enter an asterisk (*) to select all samples, or enter a space for no selection. D-72 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Update the cumulative value to reflect the edited contents in the cumulative value. Perform the update in [Parameters], [QC Control], [QC Specific]. D-73 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
QC sample is replaced, a comment specifying this activity can be added by pressing function key F7 (test comment). Enter the comment, then select OK. D-74 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Routine, Emergency, STAT (Serum, Urine, Others): Select the sample type to be searched. Each time the space bar is pressed, the check box switches from “Checked” to “Unchecked.” D-75 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
AFTER the analyzer has shifted to the STOP, PAUSE, STANDBY, or WARM UP mode. Do NOT edit any data during analyzer MEASURE modes (MEASURE 1, MEASURE 2, MEASURE3 OR MEASURE END) as indicated on the analyzer screen mode display area. D-76 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Each time the space bar is pressed, the check box switches from “Checked” to “Unchecked.” Sample No.: Enter the sample number range in this field. Enter an asterisk (*) to designate all samples. D-77 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
2. Set “Sex”, “Age’s upper limit”, “Age’s lower limit” , “Cuvette” and 6 types of “Patient information”. 3. After the entry has been completed, select the “YES” button. The setting contents will be registered as the search keys. D-78 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
2. Press function key F6 (recal. data). A window to execute recalculation is displayed. 3. Select “OK.” Data recalculation starts. Data saved as an OD value can be modified by recalculating. Data saved as a concentration value cannot be modified. D-79 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Print the repeat run data list: 1. In the [Repeat Data Verification] screen, press function key F3 (print). A window to start printing is displayed. 2. Select “OK.” Printing begins. D-80 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Enter a space if no selection is required. Report No.: Data lists entered as reports will display in numerical order. Make a selection from the pull-down menu. D-81 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
A maximum of 6 kinds of patient information can be set. 3. After the setting is completed, select “YES” at the prompt titled "Are the above criteria sufficient?" D-82 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
If new tests are added, they must be added to the format or the test won't print. Follow the procedure called "How to format a list" in the section called "List Format" located in this chapter. D-83 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
[Parameters], [Requisition Format]. Enter an asterisk (*) to designate all sample IDs. QC, Reagent Blank and Calibration: Set the sample type to be searched. Select a check box for either QC, reagent blank, or calibration. D-84 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
4. Select function key F7 (list pending) to display a list of samples that have not been transferred to the host. 5. Select function key F6 (search add.) to restrict the data transferred to a specific sex or age. D-85 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
3.10.2 System ........................D-148 3.10.3 Auto Power On ......................D-150 3.10.4 Password .........................D-151 3.10.5 Login Name & Menu Level ..................D-152 3.10.6 User Menu .......................D-154 3.11 Comment Masters ....................D-156 D-87 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide March 31, 2005...
The use cuvette tab displays the names of tests and the cuvette side used by that test. Print settings: 1. In the [Test Name] screen press function key F3 (print). A window to start printing is displayed. 2. Select “OK.” Printing begins. D-89 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Test Name - List: Lists the test names entered in a profile. Use this software screen to: Create a profile: For procedures, refer to "Creating a New Profile" located in the Basic Operations Chapter. D-90 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
ISE can be added to one round. Tests Selected: Displays the total number of tests added to a single round. Test Name - List: Lists the test names added to a round. D-91 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
LIH screening. When set to "all," LIH analysis will be performed on all samples. When set to "select," LIH will only be performed on those samples with a requisition for the LIH linked reagent. D-92 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Description: Use this screen to set specific parameters for each of the individual tests. Enter specific parameters based on the reagent manufacturer’s methodology sheet. Beckman Coulter methodology sheets are located in the User’s Guide Vol. II. Use this software to: View and/or Program Specific Test Parameters 1.
Press the spacebar, then enter, if no setting is desired. The “a” flag is generated when the onboard stability period is exceeded. D-94 D-94 Automated Chemistry Analyzer Automated Chemistry Analyzer Chapter D - Software Chapter D - Software AU2700 User's Guide AU2700 User's Guide April 18, 2008...
Methodology sheet). • Saline OLP1003 The following is a procedure to program parameters using saline the LIH reagent. The following screen provides an example only. Please contact Beckman Coulter Technical Services for additional information STANDBY 11/30/2000 10:10 Specific Test Parameters...
When all parameters are entered, press function key F2 (exit). Parameters are set. Press function key F2 (Exit) again to return to the main menu. D-97 D-97 Automated Chemistry Analyzer Automated Chemistry Analyzer Chapter D - Software Chapter D - Software AU2700 User's Guide AU2700 User's Guide April 18, 2008...
2. Enter a calculated test name. Use the pull-down menu located by the "Test Name" field. The </> buttons can also be used to select a test name. 3. Select a sample type. Use the pull-down menu located by the "Type" field. 4. Press function key F4 (set). D-99 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
5. Set a test name, constant, and formula. 6. After the settings are complete, press function key F2 (exit). Parameters are saved. D-100 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
For example, specify a formula such as (GOT / AST) / (GPT / ALT) using GOT / AST and GPT / ALT. A calculation formula can use a maximum of 4 tests. D-101 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
In the [Calculated Tests] screen, select function key F8 (Inter-test Check Number). A window appears. Enter an inter-test check number between 1 and 10, then select "OK." To change the sample type, use the pull-down menu next to the "Type" field or use the </> buttons. D-102 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
4. Press function key F2 (exit) twice. The settings will be saved. See Also For more information on sample blanks, refer to the "Calculations" section of the Specifications Chapter. D-103 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
“OR mode” and “AND mode.” • OR mode: Samples that are appended with any of the selected abnormal data symbols will be extracted as repeat-run samples. D-104 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Page 279
Tests that do not have a test name, or have been assigned to the inter-item calculation test, cannot be selected as a deciding test or related test. D-105 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
(1, 5 to 100 times) as that of time analysis. the first time analysis sample. Decrease the sample dilution rate less than that of the first time analysis sample. D-106 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Page 281
The sample probe then aspirates from the dilution cuvette into the test cuvette. Reagent blank, calibrators, and OC samples are also pre-diluted for tests programmed to pre-dilute. D-107 D-107 Automated Chemistry Analyzer Automated Chemistry Analyzer Chapter D - Software Chapter D - Software AU2700 User's Guide AU2700 User's Guide January 20, 2006...
Single check level: Set the check limits of quality control data about a reference value (mean). This can be set to multiples of the set standard deviation of control rules. D-108 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Expiration: Set the expiration date of a control sample. To set “No designation” or leave blank, enter a space. Enter the expiration date as MM/DD/YYYY (Example: 11/30/2000). Lot No.: Enter 15 characters maximum. D-109 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
6. Press function key F4 (set) to enter the control name, ID, expiration, and lot number. 7. Press the function key F2 (exit). The settings are saved. D-110 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
The display of the Mean and SD settings will contain more decimal places than what was entered because trailing zeros are automatically added to the software for QC data management purposes. D-111 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide June 16, 2006...
Page 286
When QC is edited using the [QC Monitor], [Data Edit] screen, the edits are not reflected in the cumulative value. When the edited data are within the same index, the edited contents must be reflected in the cumulative value. D-112 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
"Auto." To enable the manual repeat run, set to "Manual." To disable repeat runs on the STAT table, set to "None." Position No., Type, Control No., Control Name: Displays the position number, type, control number, and control name for each control. D-113 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Execution after Calibration: Allows the user to set the quality control analysis with the STAT table to automatically execute after the reagent blank analysis or calibration analysis (ACAL) has been performed. 8. Enter the control name and select the "Close" button. D-114 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Perform QC on the STAT Table: For procedures, refer to "Performing QC on the STAT Table" located in the "Quality Control Analysis" section of the Basic Operations Chapter. D-115 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Calibrator No.: The calibrator number will determine the position the calibrator will be placed on the yellow rack. Calibrator name: For each calibration sample set a name. Use 26 characters maximum. D-116 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
8. Repeat steps 4 -7 for each additional calibrator material. See Also For more information refer to "Performing calibrations" and "How to reprint reagent blank, calibrator, QC and patient data" located in the Basic Operations Chapter. D-117 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
If the calibration type is MB to 7MB, it is necessary to set the 'MB Type Factor' for both inner and outer cuvette wheels. D-119 D-119 Automated Chemistry Analyzer Automated Chemistry Analyzer Chapter D - Software Chapter D - Software AU2700 User's Guide AU2700 User's Guide March 9,2007...
Cal. No., point 3: This can be set only if the calibration type is set to 3MB - 7MB or 3AB - 7AB. D-120 D-120 Automated Chemistry Analyzer Automated Chemistry Analyzer Chapter D - Software Chapter D - Software AU2700 User's Guide AU2700 User's Guide March 9, 2007...
If an error occurs the user will hear an audible alarm and a error message will be generated. Processing the Slope Check When the AU2700 performs a Multi-point calibration, it compares the OD values for the setting (“+” of “-”) selected. If the result of this comparison does not agree with the values selected, an audible alarm is generated and the error message displayed below is displayed where “nn”...
2. Enter a test No. and select the “OK” button. The selected analysis test is displayed. The analysis test can be also specified from the pull-down list in the “Test Name” field, or with the “<<” or “>>” button. D-122 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
MCAL. Counts: 1 to 4 for ACAL only. Factor A: slope or multiplier, B: offset or constant, cal. no., conc. factor range: for ACAL. 6. Press function key F2 (exit). Parameters settings will be saved. D-123 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
OD values and concentration values, set a calibration chart. OD 5 OD 4 ● This will be used for immunoassay and turbidimetric assays. OD 3 OD 2 OD 1 OD 0 D-124 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
OD 6 sections has a constant slope. OD 5 Typically used for turbidimetric and immunoassays. OD 4 OD 3 OD 2 OD 1 CONC D-125 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
: A potential difference between LOW STANDARD liquid and MID liquid Concentration Set a calibration using a potential difference value between the two points of known concentration. (logarithm) mmol/L D-126 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
A-CAL, the instrument calculates the difference between C and C’. This A-CAL correction is applied to values in addition to any M-CAL correction coefficients B. D-127 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
STAT table, set to "None." Position No., Type, Control No., Control Name: Displays the position number, type, control number, and control name for each calibrator. D-128 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Use "Change Bottle No." when the reagent bottle # of a test you have selected has changed. Use "Change Lot No. when the reagent lot # of a test you have selected has changed. D-129 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Page 304
For procedures, refer to "Performing Calibrations and a Reagent Blank" located in the Basic Operations Chapter. Caution Tests set as a sample blank cannot be tested with auto-calibration on the STAT table. D-130 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
This can result in the wrong S. ID being associated with the test results. D-131 D-131 Automated Chemistry Analyzer Automated Chemistry Analyzer Chapter D - Software Chapter D - Software AU2700 User's Guide AU2700 User's Guide March 20, 2009...
End Code - right pull-down list: Sets the second byte of the End code to “No” or within the range of “01h” to “1Fh”. To change the setting, select one of them from the pull-down list. D-132 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
4. According to the messages displayed, set the online parameters to accommodate the host. Note The function key F7 (special) is used to make different Beckman Coulter analyzers compatible with each other when communicating with the host computer. 5. After the settings are complete, press function key F2 (exit). Parameters are saved.
5. Press function key F5 (edit). Enter the test number, then select the "Close" button. 6. Repeat steps 3-5 for each test to be set. 7. After the settings are complete, press function key F2 (exit). Parameters are saved. D-134 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Real-time Report: Prints reports on a real-time basis. This function can be used only if real- time report was selected from the "Real-time Print" field. Use the report name set-up in the [Report Format] screen. D-135 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Set up a second printer: 1. First set the "second printer" field to "Yes" located in [Parameters], [System], [Option]. 2. Select [Parameters], [Format], [Printer]. Set the second printer fields for use as required in your laboratory by selecting function key F5 (second printer). D-136 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
ID field length that communicates with the host. Sex (Optional): Select the check box for the sex field to be available for entry in the [Test Requisition] screen. D-137 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
[Test Requisition] screen and on the formal report. A maximum of 20 digits can be entered. 3. After completion, press function key F2 (exit) twice. The settings are saved. D-138 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Data Format: Set data field at 6 or 9 digits Test _ _ _ _ X X Test _ _ _ _ _ _ _ X X D-139 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Example: Set the line coordinates [AST] to 26, [GLU] to 27. Continue for all tests. Test Result: [4] [16] Normal Range: [25] [26] Unit: [50] [26] Data Flag: [65] [16] Spaces Name and Test Result: D-140 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Print a report according to the set parameters: 1. Select [Parameters], [Format], [Report Format]. 2. Press function key F7 (format print). A window to start printing is displayed. 3. Select “OK.” Printing begins. D-141 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Use this software screen to: Format a list: For instructions, refer to step 6 of the "Programming a new test" procedure located in the Basic Operations Chapter. D-142 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Use this software screen to: Set a Contamination Table A maximum of 100 test combinations can be defined in contamination parameters. 1. Select [Parameters], [Special], [Contamination Parameters]. D-143 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide June 16, 2006...
If cancel does not have a ü, the probe will wash with the appropriate cleaner this occurs under all conditions. Beckman Coulter strongly recommends against using the ü (except where indicated) because special wash procedures will NOT occur if the instrument goes to measure 2, stop, pause, or standby and restarts testing sensitive tests..
1 of 4 different abnormal reaction types for prozone effects in an increasing turbidimetric assay. For detailed procedures, contact a qualified Beckman Coulter representative. See Also For parameters, refer to C-Reactive Protein, Ferritin, IgA, IgG, and IgM located in the Methods Chapter.
3. Set the optional conditions using your mouse. First printer: Select 80, or 136. Second printer: Select none, 80, or 136. 4. Press function key F2 (exit) twice. D-147 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
The following restrictions apply if either ISBT-128 or Mixed is selected from the barcode types: If ISBT-128 is selected: • Check that the mode is set to "Yes" D-148 D-148 Automated Chemistry Analyzer Automated Chemistry Analyzer Chapter D - Software Chapter D - Software AU2700 User's Guide AU2700 User's Guide March 20, 2009...
3. Set the following: test requisition, barcode, rack number limit, and auto/standard repeat. 4. After settings are complete, press function key F2 (exit). The settings are saved. D-149 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Enter each auto start time in the corresponding field. Use the “HH:MM” format in military time. 4. After settings are complete, press function key F2 (exit) twice. The settings are saved. D-150 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
4. Enter the same password in the "confirm" field and press enter. 5. Select function key F2 (exit). The new password is set, if the old and new passwords were typed consistently. D-151 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
8. Select "Close" and press function key F2 (exit). The change will not be enabled until the next power on. 9. Check the log-in function box. D-152 D-152 Automated Chemistry Analyzer Automated Chemistry Analyzer Chapter D - Software Chapter D - Software AU2700 User's Guide AU2700 User's Guide March 20, 2009...
5. Select F2: Exit. The message "This function will not be enabled until next Power On" displays. 6. Perform an End Process. 7. Press the ON button. 8. The Login Window appears after F2: Exit is selected from [Start Condition]. D-153 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
7. Repeat steps 2 through 5 until all the screens you want to add are part of the User Menu. 8. Press function key F2 (exit). 9. When [User Menu] is selected from the Main Menu, a pull-down window of the added screens will appear. D-154 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Modify a shortcut key or screen name: Warning Do not modify the software names unless instructed to do so by authorized Beckman Coulter personnel! 1. Move the cursor over a screen for which the shortcut key or name is to be modified, then press function key F6 (change).
3. Select function key F5 (edit). A window appears. 4. Enter a comment up to 50 characters maximum. 5. Select "Close." 6. Press function key F2 (exit) twice. The settings are saved. D-156 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Sample ID: Set the sample ID range within the range specified in [Parameters], [Requisition Format]. To specify all samples IDs, enter an asterisk (*). To leave the field blank, enter a space. D-158 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
(Low and high limit)," and 6 types of demographics.Select "Yes." The settings will be saved as search keys. Note An asterisk (*) can only be used if specific demographic fields are used in sample processing. D-159 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
The operator is allowed to specify a search key for searching the data on the correlation diagram after statistical calculation. However, a correlation diagram cannot be produced for serum samples versus urine samples. D-160 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
3. After specifying the sample number of data points to be removed, press function key F2 (delete). A window to start removal is displayed. 4. Select “YES” and press the Enter key. The specified data will be removed from the calculation and a recalculated correlation diagram will be displayed. D-161 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Sample ID: Set the sample ID range within the range specified in [Parameters], [Requisition Format]. To specify all samples IDs, enter an asterisk (*). To leave the field blank, enter a space. D-162 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
• Number of samples exceeding the largest class interval and its % to the whole. 2. Press function key F2 (exit). The window will be closed. D-163 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
2. Select the index, start date, end date, and sample type by using the pull-down menu located by each field. 3. To check the reagent volume, press function key F5 (reagent volume). The quantity of reagent assigned to R1 displays. D-164 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Select "Save files to FD." Highlight "Reagent Consumption." Press F5 Execute. Note The file can be opened as an Excel file, and emailed as an attachment. D-165 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Type: Select the sample type from the pull-down menu: serum, urine, or other. Test (Sunday-Saturday): This chart displays the mean reagent consumption that has been calculated for each day of the week. D-166 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
5. Enter a value for the expected reagent consumption. 6. Press the "Close" button. 7. Repeat steps 2 through 6 until all settings are complete. 8. Press function key F2 (exit). D-167 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Load files from FD: Loading will overwrite the parameters set by each submenu of the [Parameters] menu with those which are loaded. Save the old parameters on another floppy disk if they need to be retained. D-168 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
4. Insert the floppy disk in the computer. Select function key F5: Execute. From the “Insert FD Start” prompt, select yes. The message “Operating: Wait for a while” displays. D-169 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Sample ID: Set the sample ID range within the range specified in [Parameters], [Requisition Format]. To specify all samples IDs, enter an asterisk (*). To leave the field blank, enter a space. D-170 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
1. Function key F5 (analyst initial): Allows the operator to enter their name or initials on the graph. 2. Function key F6 (note edit): Allows the operator to enter a "note" or comment on the graph. D-171 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Use this software screen to: Shift to the Standby Mode: For procedures, refer to "Standby Mode" located in the Basic Operations Chapter. D-172 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
The [Maker Maintenance] submenu should be performed according to the instructions provided by Beckman Coulter Technical Services or authorized service personnel. The [Maker Maintenance] submenu will be used for service, inspection, and repair work performed on this system by service personnel.
Use the "FD Data Initialize" tab to initialize (Format) a floppy disk. Initializing a floppy erases data from an old floppy or prepares a new floppy for use with this system. Floppy disks can be initialized for use as an AU2700 parameter disk, or AU2700 data disk. Tips Before initializing a floppy diskette, always make sure that the data saved on it can be erased.
5. Insert the floppy disk in the computer. Select function key F5: Start Save. From the “Insert FD Start” prompt, select yes. The message “Operating: Wait for a while” displays. D-175 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
2. Press function key F4 (set). A cursor appears in the "year" field. 3. Make the necessary changes. Tab to the next field. 4. After the settings are complete, press function key F2 (exit) twice. The settings are saved. D-176 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
3. Press function key F5 (execute). A window to start rebuilding displays. 4. Select “Yes” at the prompt. The rebuilding of the database begins. D-177 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide February 4, 2004...
Use this screen to save serum, urine, or other sample data to external floppy disk media in text format. Data saved in text format cannot be loaded back onto an AU2700. To use this function, the online selection in parameters must be set to batch.
ID numbers displayed. Choose a storage media and save the data: To save data to an Beckman Coulter 128 MB SmartMedia card, do the following: 1. Press F7 (Mode Select) to toggle the display to Data Output. 2. Press F8 (Set Format). The Set Format window appears. Select the desired demographic information, text format, and output sequence.
Stop ALARMCLEAR Description: Use this screen to connect to remote technical support provided by Beckman Coulter Technical Services. Press F5 Connect to start or stop Remote Maintenance: If the message in this box says "Running Remote Maintenance," then this gives OAI personnel access to connect to the PC via the modem.
2. Select a maintenance function listed on the screen. 3. Press the STAT ROTATION/DIAG button. The maintenance process starts. For detailed maintenance procedures, refer to the Maintenance Chapter. D-181 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide March 31, 2005...
Unit: Used to set the method for calculating the consumption and average live of the consumable. Select days, weeks, months, or years from the pull-down menu. Use the field that follows the unit field to add the life expectancy. D-182 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide March 31, 2005...
Recent: The most recent date of supply or replacement will be printed. History: All the records, including all past dates of supply or replacement and the use values are printed. A message indicating printing is displayed. D-183 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
2500, Months - 1 to 600, Years - 1 to 50. Period Unit: Set the maintenance period in terms of days, weeks, months or years, using the pull-down menu. D-184 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
2. Select function key F5 (execute). The "executed" date and "next" scheduled date will appear. Note The Photocal and W2 are automatically updated for the "executed" and "next" scheduled dates. D-185 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
1. Select [Maintenance], [Maker Maintenance], [Program Version]. 2. The screen appears with program, version and revision information. 3. After viewing the screen, press function key F2 (exit) to return to the main menu. D-186 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
C/Rotation Step: Moves one step between the wash station and items selected. Select the following: From (rack, cuvette, STAT table, or wash/STD table) and To (cuvette, or ISE). D-187 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
S/Pulse: Pulses the reagent compartment clockwise or counterclockwise the number of steps specified from 1 to 999. It is recommended to use one (1). All Pump T/On: Activates water and air pumps. U/Off: Deactivates water and air pumps. D-188 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
L/Reset: Rack holder moves to the right. M/Step: The rack push bar steps forward as far as the sampling belt and then back as the Rev./Diag button is pressed again. D-189 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
H/S. ID Movement: Performs a complete rack sequence and reads S. ID data. I/Rack ID: The rack located in the "rack ID read position" is read. J/Rack ID Movement: The barcode reader is activated. D-190 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Q/Probe Wash: The wash wells are activated and the probes move into the wells. R/8 bit LED: Performs a diagnostic sequence of the 8 bit LED. S/SUB CPU Check: Performs a diagnostics check of the photometer and optical sub- CPUs. D-191 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
"Performing Calibrations and a Reagent Blank" section located in the Basic Operations Chapter. 2. Select [Maintenance], [Maker Maintenance], [ISE Diag]. 3. The calibration begins as soon as one of the following is selected: I/Calibration for serum, J/Calibration for urine, K/Calibration for serum +urine. D-192 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Maintenance Chapter. 4. Verify the data printout results are within the acceptable range. D-193 D-193 Automated Chemistry Analyzer Automated Chemistry Analyzer Chapter D - Software Chapter D - Software AU2700 User's Guide AU2700 User's Guide March 20, 2009...
Acceptable % CV < 0.6 < 2.0 < 0.7 See Also For procedures refer to the AU2700 User's Guide Volume II Troubleshooting Chapter "Sequential Sample Measure." Explanation of functions: Prime A/Total Prime: Primes all solutions. B/Buffer Prime: A one cycle buffer solution prime is performed.
2. Choose "Displayed Date" and "All Date." Displayed date only prints the date selected. "All date" prints alarm logs for all dates. 3. Select "OK." Printing begins. D-195 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
If a floppy is removed from the drive while data is being saved, the data being saved or stored in the floppy may be lost. Floppy disks for parameter storage have an “AU2700 PARAM” volume label. Initialize a 3.5 floppy disk as a parameters disk one time before parameters are saved or loaded. The floppy disk can be 2HD or 2DD.
(7)Repeat (8)Data Edit (9)Data Correction (0)Report See Also For information on customizing the [User] menu, refer to the section in this chapter under [Parameters], [System], [User Menu]. D-197 Automated Chemistry Analyzer Chapter D - Software AU2700 User's Guide May, 2002...
Introduction ....................E-1 Chapter E Software Menu Outline Routine Menu Outline ....................E-2 Parameter Menu Outline ...................E-3 Auxiliary Menu Outline ....................E-4 Maintenance Menu Outline ..................E-5 List of Screen Operations ..................E-6 Automated Chemistry Analyzer Chapter E - Software Menu Outline AU2700 User's Guide March,2000...
Used to display/print analysis results as reports. (D)Data List Used to display/print the master data file. (N)Online Used to send data to the online host computer. Automated Chemistry Analyzer Chapter E - Software Menu Outline AU2700 User's Guide March, 2000...
Used to program a protocol if reagent cross-contamination is suspected. (D)Data Check Parameters Used to set the conditions of data check. For detailed information, contact an Beckman Coulter representative or the reagent manufacturer. (S)System (O)Option Used to set up optional analyzer equipment.
Verifies the calibration performance by putting a calibration specimen into the same analysis used for the general samples. (B)Standby Set Moves from the warm-up mode to the standby mode. Automated Chemistry Analyzer Chapter E - Software Menu Outline AU2700 User's Guide March, 2000...
Chronologically lists the alarms that have occurred. (F)File Management Used to save the parameters file in the floppy disk and to load the file from the disk. Automated Chemistry Analyzer Chapter E - Software Menu Outline AU2700 User's Guide March,2000...
Need help?
Do you have a question about the AU2700 and is the answer not in the manual?
Questions and answers