1.1.3 Equipment safety symbols

6
.
Document no
M1144951-004
-
Attention, consult accompanying documents.
-
When displayed next to the O
is set below 21%.
-
When displayed next to the HR value, indicates that the pacer is set on R.
-
On the modules or frames indicates that modules with identical
measurements should not be used in the same monitor. If such modules
have been inserted, remove the module that has been most recently
connected. You can also remove both modules and re-connect the new
module after five seconds.
-
On the Interface Module E-INT, M-INT, indicates that it is for connecting
external devices. Do not connect patient cables to the module.
-
On the E-TONO, M-TONO module indicates that the module should only be
used with Tonometrics catheters.
-
On the E-miniC module indicates that airway gases should be calibrated
every six months in normal use and every two months in continuous use.
-
BIS: On the Aspect DSC indicates that the converter must not be opened for
any reason or autoclaved.
-
On the E-PRESTN, E-PRETN, E-RESTN, E-PSM, E-PSMP, E-P, E-PP, E-PT, E-COP
and E-COPSv module indicates that protection against cardiac defibrillator
discharge is due in part to the accessories for pulse oximetry (SpO2),
temperature (T) and invasive pressure (P) measurement.
-
On the E-NMT, M-NMT module indicates the following warnings:
-
Do not place the NMT stimulating electrodes on the patient's chest.
-
Always stop the NMT measurement before handling the stimulating
electrodes.
-
Never subject a patient with an implanted electronic device to electrical
stimulation without consulting a medical specialist first.
-
On the rear or bottom panel this symbol indicates the following warnings
and cautions:
-
Electric shock hazard. Do not open the cover or the back. Refer servicing
to qualified service personnel.
-
For continued protection against fire hazard, replace the fuse only with
one of the same type and rating.
-
Do not touch a battery-operated monitor during defibrillation procedure.
-
Disconnect from the power supply before servicing.
-
Do not use the monitor without manufacturer approved mounting
attached.
-
Lithium battery on the CPU board: follow the regional regulations for
disposal.
value, indicates that the FiO
2
low alarm limit
2