Equipment and Personnel Safety

Related Hazards
This equipment contains dangerous voltages that are capable of serious injury or death.
WARNING
If any defects are observed or malfunctions occur, stop operating the equipment and perform the proper action for the patient. Inform a
qualified service person and contact a Service Representative for information.
There are no user serviceable components inside the console. Refer all servicing to qualified service personnel only.
Only approved and recommended peripherals and accessories should be used. All peripherals and accessories must be securely
WARNING
connected to the LOGIQ C5/C5 PRO.
The concerns listed below can seriously affect the safety of equipment and personnel during a diagnostic ultrasound examination.
DANGER
Explosion
Risk of explosion if used in the presence of flammable anesthetics.
Hazard
To avoid injury:
Electrical
Hazard
Do not remove protective covers. No user serviceable parts are inside. Refer service to qualified service personnel.
To assure adequate grounding, connect the attachment plug to a reliable (hospital grade) grounding outlet.
Never use any adaptor or converter of a three-prong-to-two-prong type to connect with a mains power plug. The protective earth
connection will loosen.
Do not place liquids on or above the console. Spilled liquid may contact live parts and increase the risk of shock.