Managing device setups
Configuring or reconfiguring your instrument could take a non trivial amount of time. With that in mind,
once you have properly configured your instrument, you can save your setup as a baseline configuration
that you can revert to if necessary. The following pages explain how to proceed.
Saving a default user setup
To save your current configuration as your default user setup:
1 Select Device Setup > Save Setup & reset > Save as User Default.
2 Press OK. Your current setup is saved in the instrument non-volatile memory.
Resetting to default user setup
Provided that you have already saved a setup as your default setup, you can return to that default user
setup at any time.
To do so:
1 Select Device Setup > Save Setup & reset> Reset to User Default.
2 Press OK. Your current setup is erased and replaced with the default user setup.
Resetting to factory default
Should you be in a situation where the default user setup does not work anymore (e.g., you have moved
the LWT instrument to a new vessel), you can simply revert to the factory default setup and start
configuring your instrument anew.
To do so:
1 Select Device Setup > Save Setup & reset. > Global Factory Reset.
2 Press OK. Your current setup is erased and replaced with the default factory setup.
57
Configuring the instrument