General Safety Instructions - GE CRxVision Operating Manual

Computed radiography
Table of Contents

Advertisement

CAUTION:
Magnetic particles can be transported in the CRxVision and stick on the drum.
As a consequence, image artifacts are possible.

General safety instructions

The specified image quality is reached only, if the CRxVision is operated in a dark
room.
The USB flash drive connected to the CRxVision may not be removed during
operation. The CRxVision is constantly using the USB flash drive to write data on it.
If an image plate has not been used for more than 48 hours or if an image plate has
been exposed to an exceptionally high X-ray dose you must re-erase the image plate
before re-using it in order to prevent ghost images from interfering with the image of
interest.
Make sure that the CRxVision is constantly monitored in order to avoid inappropriate
handling, especially by children.
Only trained service personnel must make repairs. Only authorized service personnel
must make changes to the CRxVision.
If there is any visible damage to the CRxVision casing, do not start nor use the
CRxVision.
Do not override or disconnect the integrated safety features.
Do not apply excessive shock or vibration to the CRxVision during operation. This
may decrease the image quality. Neither should the CRxVision be moved during
operation.
Do not allow the CRxVision to be subject to excessive vibration during operation, due
to unstable ground (e.g. vibration of nearby equipment or footsteps).
Switch off the CRxVision before performing any maintenance work or repairs.
Disconnect the CRxVision from the mains before making repairs or performing any
maintenance activities during which live electrical components may be exposed.
As is the case for all technical devices, the CRxVision must be operated, cared for
and serviced correctly. A regular quality control is recommended.
If you do not operate the CRxVision correctly or if you do not have it serviced
correctly, the manufacturer is not liable for resulting disturbances, damages or
injuries.
If you notice conspicuous noise or smoke, disconnect the CRxVision immediately.
Do not pour water or any other liquid over the CRxVision.
Perform no other operations on the CRxVision than those described in this document.
Switch the system off before moving it. When reaching the new position, switch the
system on again.
Do not transport the CRxVision without packaging
CRxVision Operating Manual V1.2
Take care to keep the image plate clean.
Take special care to keep it free from magnetic particles like swarfs.
25