Item
Description
Comfort
Allows you to remove the Comfort Screen feature.
Set real time clock for time, day and date. Choose how date is displayed (North American or
Set Time
European). Choose a 12 hour (AM/PM) or 24 hour time display.
Display the current software revision of the options installed on this specific unit (Humidifier,
View Mods
Scale, SPO2, etc.)
Down
Go to second service screen.
Up
Return to previous screen.
Check status of all self test the
software runs continuously.
These include: incubator heater
on (INCHTRON), incubator
heater off (I/HTROFF), humidifier
heater on (RHHTR), remote
monitoring data stream
(RS232LOOP), incubator fan on
(FANON), and incubator fan off
Status
(FANOFF).
If the RS232 option is not
installed, RS232LOOP will display
fail. To test the circuit if the
option is installed, short pins 2
and 3 on the 9 pin connector
on the back of the electrical
enclosure.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 4: Troubleshooting
Figure 4-2 Second Service Screen
Figure 4-3 Status Menu
6600-0356-000 103