Chapter 4: Troubleshooting

4.2 Service Screen

To access the service screen, hold in the override button (>37) during power up until the software revision
screen appears. Release the button and the first service screen will appear.
Default options that may be selected from this screen appear along the left side of the screen.
Item
Description
Digital/Analog Converter voltage. This is the over temperature voltage that is used by the system
to verify the computer independent circuitry is working. To manually test this circuit enter
DAC Volt
voltages from 0 to .5 V. The DV value at the bottom right corner of the screen should match this
value within 10mV.
English is the default language that appears on the EL display, but you can select French,
Language
Spanish, etc., depending on what software is installed.
Changes temperature displays to show "Fahrenheit", "Celsius" or "Celsius Only" so the Fahrenheit
Temp U
option is not present on the user Set-up screen (Celsius is factory set default).
Volume
Select one of four volume settings; 1 is minimum, 4 is maximum.
Set the default Hot Baby/Cold Baby alarm to activate when either 0.5ºC or 1.0ºC difference is
Pat Alarm
read between a set temperature and the baby probe temperature.
Elevate
Disable or enable the bed up/down pedal switches.
Pat Ctrl
Allows you to disable patient control.
Pat Algo
To be used for future software options.
Select from "Grams", "Pounds" or "Grams Only" so that the pounds option is not present on the
Scale U
user Scale screen (grams is the factory default).
Scale R
Select from 2 scale resolution settings; 10 grams or 5 grams (10 grams is the factory default).
28

Figure 4-1 First Service Screen

6600-0356-000 103