5.5.3 Incubator Heater Replacement

Refer to "Figure 5-6 Heat Sink and Fan" and "Figure 6-10 Lower Chassis".
Incubator heater
Motor shaft

5.5.3.1 To Remove the Heater Closest to the Foot of the Bed

1. Remove the chassis cover. (Refer to section 5.5.1.)
2. Remove the screw (4 mm hex key) in the heatsink directly in front of the heater cartridge.
3. Disconnect the heater's electrical connector.
4. Remove the retaining clip and slide the heater out of the heat sink.
NOTE: On early units, the incubator heater screwed into a threaded hole in the heat sink. For these units you
must order kit 6600-0228-850 to replace both the heaters and heat sink.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
fan hub
Fan seal
Chassis

Figure 5-6 Heat Sink and Fan

Fan
Heat sink
Incubator heater
(not shown)
Heat sink gasket
Side chassis cover
6600-0356-000 103