TROUBLESHOOTING
ERROR CODES
General Information
The Intelect Legend XT Therapy Systems incorporate error messages, and warnings to inform the user of problems or potential problems
with the system, modality, or accessories. These are numbered so the user can possibly correct the problem without the aid of service
personnel. Use the following Troubleshooting Charts to define the error codes, and locate the probable cause and possible remedies
before contacting the dealer or factory for technical service.
Code
Type
Number
Message
100
Warning
101
Warning
102
Warning
103
Warning
104
Warning
106
Warning
107
Warning
Probable Cause
Overcurrent
Shorted Lead Wires
Bad Contact Quality
Blank Patient ID
1. Blank Protocol Name
2. Blank Sequence Name
1. Attempting to delete factory set Sequence.
2. Attempting to delete Clinical Protocol.
A. Check Electrodes and Lead Wires. Make certain Lead Wires are not damaged and are
properly connected to the system. Make certain Lead Wires are properly connected to
the Electrodes and that electrodes are not damaged and are making proper contact with
treatment area.
B. Replace Lead Wires and Electrodes.
A. Check Electrodes and Lead Wires. Make certain Lead Wires are not damaged and are properly
connected to the system. Make certain Lead Wires are properly connected to the Electrodes
and that electrodes are not damaged and are making proper contact with treatment area.
B. Replace Lead Wires and Electrodes.
A. Make certain Electrodes are making proper contact with the treatment area.
B. Make certain Lead Wires are properly connected to Electrodes.
C. Replace Electrodes and Lead Wires.
Properly enter Patient ID. Refer to Therapy System User Manual for Patient Data Card instructions.
Properly enter Protocol or Sequence Name. Refer to the appropriate section of the Therapy System
User Manual.
Cannot delete factory set Clinical Protocols or factory set Sequences.
86
Intelect Legend XT® Therapy System
Possible Remedies
Need help?
Do you have a question about the Intelect Legend XT 2760 and is the answer not in the manual?
Questions and answers
Is there a way to change the default time setting for different modalities (such as ultrasound and interferential estim) so that every time we turn on the machine, we do not have to adjust the time to what we most typically use? Rather than setting up protocols I'd like to just change the default treatment time. Thank you.