TROUBLESHOOTING
ERROR CODES
The Intelect Transport Combo displays error messages to inform the user of problems or potential problems with the unit, modality, or
accessories. These are numbered so the user can possibly correct the problem without the aid of service personnel. Use the following
Troubleshooting Charts to define the error codes, and locate the probable cause and possible remedies before contacting the dealer or
factory for technical service.
Code
Type Message
Number
104
Message
105
Message
106
Warning
107
Warning
108
Warning
109
Message
200-399
Message
Probable Cause
User attempted to perform an electrotherapy session, but both
channels are already in use.
User selected a two channel electrotherapy treatment, but at least
one channel is already in use.
Overcurrent
Bad Contact Quality
Shorted Lead Wires
While performing ultrasound therapy using the 10 cm
2
an intensity greater than 15 W (1.7 W/cm
), user attempted to
begin an electrotherapy session on a second channel.
Refer to Warning on next page.
A. Wait until the previous therapy session finishes.
B. Press the STOP button to end the therapy session on either channel.
A. Wait until the previous therapy session finishes.
B. Press the STOP button to end the therapy session on either channel.
A. Check Electrodes and Lead Wires. Make certain Lead Wires are not damaged and are properly
connected to the system. Make certain Lead Wires are properly connected to the Electrodes
and that electrodes are not damaged and are making proper contact with treatment area.
B. Replace Lead Wires and Electrodes.
A. Make certain Electrodes are making proper contact with the treatment area.
B. Make certain Lead Wires are properly connected to Electrodes.
C. Replace Electrodes and Lead Wires.
A. Check Electrodes and Lead Wires. Make certain Lead Wires are not damaged and are properly
connected to the system. Make certain Lead Wires are properly connected to the Electrodes
and that electrodes are not damaged and are making proper contact with treatment area.
B. Replace Lead Wires and Electrodes.
2
sound at
A. Use a smaller sound head.
B. Set the ultrasound at an intensity less than 15 W (1.7 W/cm
C. Wait until ultrasound session is complete.
Refer to Warning on next page.
78
Intelect® Transport Combo
Possible Remedies
2
).
Need help?