7 PROGRAMMING AND ELECTRICAL CALIBRATION

7.1 Access to Secure Parameters

A 5-digit security code is used to prevent tampering with the secure parameters.
0 0 0 0 0
SECURITY CODE
– – – – –
Espanol

7.2 Language Page

_ _ _ _ _
Espanol
˜
Francais
– – – – –
SET UP PARAMETER

7.3 Set Up Parameters Page

– – – – –
SET UP PARAMETER
– – – – –
Disp. Units ug/kg
ug/l
ppb
760
Pressure
mmHg
0
Salinity
ppt
– – – – –
Temp. Units
(C)
– – – – –
Autocal
4 Weeks
1 Week
1 Day
Off
– – – – –
Reset Cal.
NO
YES
– – – – –
SET UP ALARM
18
Security Code
Enter the required code number between 00000 and 19999 to gain access to the secure
parameters. If an incorrect value is entered, access to subsequent programming pages
is prevented and the display reverts to the
Advance to
, Section 7.2.
Language Page
Language Page
Select the language to be displayed on all subsequent pages: Español, Francais,
Deutsch or English.
Advance to
Set Up Parameters Page
Press
to advance to next parameter
or
Press
to advance to
Set Up Alarm Page
These two switches are used to advance to all subsequent parameters and pages. If a
parameter is changed it is automatically stored on operation of either switch.
Display Units
Select the required display units:
µg/kg, µg/l, or ppb.
Barometric Pressure Correction
Set the local barometric pressure in mm Hg (between 500 and 800).
If the local barometric pressure is unknown the default value, which is the standard sea-
level value of 760 mm Hg, should not be changed.
Salinity Correction
Required when monitoring sea water or other waters containing high concentrations of
dissolved salts.
Enter the appropriate value between 0 and 80 parts per thousand (ppt).
Leave at the default value of 0 ppt if correction is not required.
Temperature Units
Select either °C or °F.
Auto Calibration
Select the frequency of automatic calibrations: 1 Day, 1 Week or 4 Weeks. Select Off to
disable automatic calibrations. Only manual calibrations may be carried out.
Reset Auto Calibration
Select
to reset the timing for automatic calibrations. To fix the calibration to a specific
YES
time of day select
at the correct time of day. (This will be automatically reset following
YES
a power failure.)
Advance to
Set Up Alarm Page
.
Operation Page
, Section 7.3
, Section 7.4.
, Section 7.4.