Err no.:
Type:
Usertext:
6115
E
Not possible to start visualization, as no test is running
6116
E
Automatic program selection not possible as no program or report field input available
6117
W
-Cancel 'Error' and 'Ready' states; -Cancel 'Drying, internal/external adjustment not performed'; -Cancel 'Test repeat, PLC
programs, last result'; -Abort program in progress Do you want this to be carried out?
6118
W
Once you start the following procedure, it must be performed completely. Do you want to start this procedure?
6119
E
Password was entered incorrectly too often. Please inform your system administrator.
6120
E
New user needs to enter new password.
6121
E
Temperature informations: <ARG1> <ARG2>
6122
E
<ARG1> ----------------------------------------------------- Temperature informations: <ARG2> <ARG3>
6123
E
No test result.
6124
E
Cleaning and drying record not available.
6125
E
Your password does not have the required minimum length of <ARG1> character(s); please enter new password.
6126
R
The drying process was completed. - Close shutoff valve (f) - Disconnect "Venting 1" tubing from pressure tank (g) - Dis-
connect remaining tubing from pressure tank (g). Caution: Tank (g) is being vented - Please disconnect all tubing from the
connectors b-e.
6127
E
The user file of the management unit is corrupt.
6901 ...
E
Printer server program cannot be addressed.
6902
6903
E
Unknown print job.
6904
E
Job cannot be sent to printer server.
6905
E
Response cannot be recieved from printer server.
6906
E
Invalid print data.
6907
E
Print job list is full.
6908
E
Invalid test number for printing
6909
E
Cannot write print file.
6910
E
Cannot read print file.
6911
E
Insufficient main memory for sending print job
6912
E
Unknown error occurred during printing.
6999
E
Internal error! <ARG1>
8000 ...
E
Communication error with test unit
8300
9001
E
It is not possible to start the selected procedure at the moment, because the former procedure has not yet been completed
- Please try again later
9004
E
Only the internal reference tank is active
9005
E
Both the internal and external reference tanks are active
9006
E
The basic configuration data were re-initialized: (Address of the test unit, serial number, software version of the test unit) -
when the unit was initially operated at the factory - when the format of the data mentioned above was changed
9010 ...
E
The raw data of the graphs were not sent to the processing unit because the required component currently does not exist
9011
9012
E
The operating pressure ist too low. Please increase the operating pressure.
9013
E
The operating pressure ist too high. Please reduce the operating pressure.
9014
E
The operating pressure is not sufficient. Please increase the operating pressure.
9015
E
The operating pressure is not stable. Please wait.
9101
E
Error while venting Warning: The system is probably still pressurized!
9102
E
The required test pressure exceeds the calibration range of the pressure sensor - Recalibrate using a wider calibration range
(0-8 bar/0-116.03 psi) - Work with a lower test pressure
9103
E
The +12 V power supply is out of tolerance.
9104
E
The -12V power supply is out of tolerance
9105
E
The +24 V power supply is out of tolerance
9106
E
Valve V1 does not close correctly.
9107
E
No compressed air supply is available or valve V1 does not open correctly
Troubleshooting
75