MAINTENANCE

Replacement of glow starter

The glow starter is located at the switch box on the lower right side.
1. Turn off the power switch and disconnect the power supply plug from the outlet.
2. Turn the glow starter to counter clockwise to remove. (see figure below)
3. Set a new glow starter. (type: FG-1P)
• The glow starter is consumables. Please replace the glow starter at the same time with the lamp.
Replacement of glow starters
(Optional Light add-on kit (MIR-L15))
The glow starters of optional light add-on kit (MIR-L15) are located inside the unit cover on the lower left side.
1. Turn off the power switch and disconnect the power supply plug from the outlet.
2. Remove 4 screws fixing the unit cover by using a screw driver.
3. Turn the glow starter to counter clockwise to remove. (see figure below)
4. Set new glow starters. (type: FG-1P)
• The glow starter is consumables. Please replace three glow starters at the same time with the lamps.
Note: Consult Sanyo sales representative or agent when replacing the glow starters.
Glow starter (FG-1P)
50
4 screws (fixing the unit cover)
Glow starter
(FG-1P)