5.2.71 Error 2201 (Calibration sensor error)

Is the connector of the touch sensor type
calibration jig (40300-0970) connected to the jack
for maintenance on the USB board?
Yes
Are there any breaks in the cable of the touch
sensor type calibration jig (40300-0970)?
No
Are there any breaks in the cable ASSY (40274-
CA11)?
No
Replace the USB board (44501-BA13)
[p139]).
Is the problem resolved?
No
Replace the main board (44501-BA01)
[p135]).
Connect the connector of the touch sensor type
No
calibration jig (40300-0970) to the jack for
maintenance on the USB board
Replace the touch sensor type calibration jig
Yes
(40300-0970)
Replace the cable ASSY (40274-CA11)
Yes
[p298] and 9.4
(see 7.3.8
Yes
(see 7.3.1
73
LLE7D*RDA002B
(see 9.1
(see 8.12.1
[p189]).
(see 9.3
[p299]).
End
[p293]).