Description Of Functions (Cont.); Interface Section - Hitachi AJ-CE series Operation Manual

High-precision tuning fork electronic balance
Table of Contents

Advertisement

4.2 Description of Functions (cont.)
Setup of units of
measurement to be
displayed*"
1
Display of the
auxiliary scale
interval"
2
GLP-compliant
print
1
"
3
Output format
while the auxiliary
scale interval is
displayed"
2
8 l.5.u.
essu
c
l. R <
0.0 LP
Ft.Fr F
* 3 30
TV4
IF
*5 00
n
u
a i
*o
i
i
i
i
3
*3
[Pcs]
[%!
Unit not set
Disable
Enable (0 is the auxiliary scale interval.)
Note: The auxiliary-scale-interval place does not
represent a verification scale interval. Use it
only as a reference value.
Disable
Enable
Prints a GLP-compliant format at calibration.
(Refer to "9.3 GLP-compliant print" (Page 31).)
No output is made while the auxiliary scale interval is
displayed.
Output is made even while the auxiliary scale interval is
displayed.
Output is made with 7" added to the left of the auxiliary-
scale-interval place.
•>K2
•>K 3
-jir5: default factory settings [H I. h. u.]~[H h. 5.u.]
Can be set only for a model on which the lock switch is off.
For a model on which the lock switch is on, only the units set when the switch was off are
effective.
The AJ[H]-820CE does not provide this function.
This setting is only available with the AJH-E Series.
4.3 Interface Section
Displayed when ['i!
E"3] is set to [ / ] o r [ c ' ]
Item
Output Control
Baud Rate
Parity
Set Value
' " ? 1. O.C.
13. b.L
r
l3. PR.
n
u
i
i
a
3
S'
s
s
' "a'
3
'-/
1
' 3
Description
Stop output
Output continuous at all times
Output continuous if stable (stop output if unstable)
Outputs
of wheth
once by pressing I Memory | key (irrespective
er stable).
Outputs once if stable. Outputs if the balance is stable
when a sample is loaded after the preceding sample has
been removed and the balance indicated zero, or less.
Outputs once if stable, and stops output when unstable.
Even if the sample is not replaced, the balance is output
once when it stabilizes next time (including the zero
indication).
Outputs
unstable
the balar
once.
Pressing
once wh
once if stable, and outputs continuously when
. Even if the sample is not replaced, output of
ice stops when it stabilizes after being output
I Memory | key causes the balance to output
an stable.
1200 bps
2400 bps
4800" bps
9600 bps
None
Odd
Even
Displayed only when [1.
l.h.
a
(7-digit numeric format)] is specified.
denotes a factory-setting
The data interval in continuous output mode is 0.1 to 1 second.