3. Basic Operations
3.1 Installation
1
Positioning the measurement pan
First, mount the pan base on the
main unit of the balance and place
the measurement pan on top of it.
Be sure to mount the pan base in the
correct direction, as illustrated
below.
2
Securing the exact level of the balance
Position of air bubble
in the level
Turn the adjusters until the bubble
rests in the center of the red circle on
the level.
Round-pan type: total of two
adjusters, one on
either side at front
Square-pan type: total of four
adjusters, one on
either side at front
and rear
<Caution>
Use caution when operating the
adjusters on the square-pan type to
prevent them from lifting up.
3
Connecting the AC adapter
Connect the AC adapter to the
balance, as illustrated at left.
S&Take 5 minutes before operation.
<Caution>
If the balance has the battery
installed, refer to "11. Operating the
Balance with the Battery," on page
39.