Manual Operation - Hitachi L-2100 Instruction Manual

Table of Contents

Advertisement

4.1.3

Manual Operation

This is used for supplying liquid under constant conditions
without using a program.
(1)Constant Flow Rate Mode
This section explains the setting of liquid delivery system at
constant flow rate. An example is given where flow rate at 1.0
mL/min, upper pressure limit at 15.0 MPa and lower pressure
limit at 0 MPa.
Monitor screen
MANUAL
MANUAL:
FLOW=1, PRESS=2
1
Select the constant flow rate mode.
CONSTANT FLOW:
Input the flow rate, and the upper and lower limit pressure
values.
CONSTANT FLOW:
The monitor screen will apper.
Monitor screen
NOTICE: The set conditions for manual operation will be erased
when power is turned off. Conditions must be set each time power
is turned on.
4 - 5
FLOW MAX
MIN
1.000 15.0
0.0
FLOW
PRESS
0.000
0.0
[MN]
[MN]
[MN]