I WARNING:
To avoid e lectric shock, donot o perate your meter until the
battery door i sinplace a nd fastened
securely.
REPLACING
THE FUSES
WARNING:
To avoid e lectric shock, disconnect
the test l eads from any
|
source of voltage before removing the fuse door.
I
1. Disconnect the test leads from the meter and any item under test.
2.
Open the fuse door by loosening the screw on the door using a Phillips
head screwdriver.
3.
Remove the old fuse from its holder by gently pulling it out.
4.
Install the new fuse into the holder.
5. Always use a fuse of the proper size and value (0.5A/250V fast blow
for the 400mA range, 10A/250V fast blow for the 10A range).
6.
Put the fuse door back in place. Insert the screw and tighten it securely.
l WARNlNG:
To avoid electric shock, do not operate your meter unti. the]
fuse door is in place and fastened securely.
UL LISTED
The UL mark does not indicate that this product has been evaluated for the
accuracy of its readings.