Rear Caster - GE LOGIQ 5 Service Manual

Hide thumbs Also See for LOGIQ 5:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300000, R
IRECTION
EVISION
8-6-5

Rear Caster

Purpose: This is a description on how to remove and replace the Rear Caster.
8-6-5-1
Tools
Common pillips screwdrivers
Allen/Unbraco wrench
Lifter (180Kg)
8-6-5-2
Needed Manpower
1 person, 15 minutes + travel
8-6-5-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 3.
Lift the system before proceed.
8-6-5-4
Removal Procedure
1.) Unscrew eight hexagonal bolts (1-8).
2.) Remove the Rear Caster. Refer to Figure 8-85.
8-84
2
6
5
2
1
4
3
Figure 8-85 Removing the Rear Caster
Section 8-6 - Body Block