Specifications

Type and Range and Display Resolution:
Differential Non-Isolated Types:
28 inH2O (1 psid) – XX.YYY
200 inH2O (7.21 psid) – XXX.YY
2000 inH2O (72.1 psid) – XXXX.Y
Gauge Isolated Types:
15 psig – XX.YYY
30 psig – XX.YYY
200 psig – XXX.YY
300 psig – XXX.YY
2000 psig – XXXX.Y
Absolute Isolated Types:
17 psia (900 mmHg) – XXX.YY
38 psia (2000 mmHg) – XXX.YY
Wet/Wet Types:
1 and 5 psid – X.YYYY
15, 30 and 50 psid – XX.YYY
100, 300 and 500 psid – XXX.YY
Accuracy:
±0.025% of Full Scale or ±0.05% of Full Scale
(±0.1% of Full Scale for 28 inH2O Differential range)
Includes the combined effects of temperature, linearity,
repeatability, hysteresis and resolution.
Warm up time = 5 minutes.
Zero unit at working ambient temperature before use.
Temperature:
Storage = -40°C to +60°C (-40°F to +140°F)
Operating = -5°C to +50°C (23°F to +122°F)
Media Compatibility:
PORTS:
DN: Differential pressure non-isolated sensors for use with
clean, dry, non-corrosive gases only.
DI, GI , AI: Differential, Gauge, or Absolute pressure sensors
for use with gases and liquids compatible with 316L SS
(and O-ring material on DI wet/wet differential sensors)
O-RING Material (for DI sensors only):
Standard: Viton
Available options: Buna-N, Neoprene or Ethylene-Propylene
Pressure Limits:
DN units: 2 × range when pressurized on high side only. 150
PSI (10.5 Kg/cm²) static when applied to both sides of
sensor simultaneously
GI and AI units: 2 × range
DI units: 5 PSI: 4 × range on high side only. >5 PSI: 3x
Page 22 of 28