2.1 Removal of Transport Fixed Plate (ChromasterUltra R
Appended Fig. 1-1 Removal of Transport Fixed Plate
(h) Power supply to motor-driven equipment (such as
stirrer and vibrator) without a noise suppressor shall
not be turned ON/OFF frequently on the same power
supply line as that of this instrument.
(1)
Open the front door of ChromasterUltra R
Autosampler.
(2)
There is a Transport Fixed Plate fixed by three screws in
the autosampler. Remove them using a Phillips screwdriver.
NOTE:
Please be careful not to drop the screws in the
instrument while at work.
It is recommended to use a magnetic driver since the
iron screws are used.