The safety and effectiveness of this device has been verified against previously distributed devices. Although all The MAC 5000 System bears CE mark CE-0459 indicating standards applicable to presently marketed devices may its conformity with the provisions of the Council Directive not be appropriate for prior devices (i.e.
Page 4
CE Marking Information For your notes CE-2 MAC 5000 System — 2000657-057, Revision B...
Purpose Conventions This manual contains the instructions necessary to operate Note á the MAC 5000 system in accordance with its function and Provides additional user information. intended use. Styles Intended Audience Bold text Indicates keys on the keyboard, text to be entered, or This manual is intended for the person who uses, hardware items such as buttons or switches on the equipment.
Press to acquire ECG data and to print or stop printing writer reports. Insert or remove a diskette as indicated by the direction of the arrow. Press to perform various exercise test functions and adjust test variables. Select the displayed screen item. 002B MAC 5000 System — 2000657-057, Revision B...
Indicates a potentially hazardous situation which, if not avoided, COULD result in death or serious injury. &$87,21 Indicates a potentially hazardous situation which, if not avoided may result in minor or moderate injury. MAC 5000 System — 2000657-057, Revision B...
Page 22
EXPLOSION HAZARD - Do not use this equipment in the presence of flammable anesthetics, vapors, or liquids. MAC 5000 System — 2000657-057, Revision B...
If you elect to collect this type of data, it is your The equipment is used in accordance with the responsibility to ensure that you comply with all applicable instructions for use. legal requirements. MAC 5000 System — 2000657-057, Revision B...
ECG tracing interpretation. To ensure accuracy, only use computer-generated tracings and not the display for physician interpretation. This equipment will not cause abnormal operation of a patient’s pacemaker or other electronic stimulator. MAC 5000 System — 2000657-057, Revision B...
Page 25
If polarizing electrodes are used, we recommend disconnecting the leadwires from the patient before delivering the shock. MAC 5000 System — 2000657-057, Revision B...
Page 26
ACCESSORY has been performed in accordance to the information before connecting any devices to this appropriate IEC 60601-1 and/or IEC 60601-1-1 equipment that are not recommended in this manual. harmonized national standard. 1-10 MAC 5000 System — 2000657-057, Revision B...
AC power to re-charge the battery. Do NOT throw the battery into the garbage. Recycle the battery. Consult accompanying documents. This position of the switch removes battery power from the equipment. MAC 5000 System — 2000657-057, Revision B 1-11...
XXXXXXXXX á XXXXXXXX XXXXXXX XXXXXXX XXX The product code for the MAC 5000 system is MP. XXXXXXXXX XX XXXX XX XXXXX J6XX0415FXX 022C 1-12 MAC 5000 System — 2000657-057, Revision B...
View the waveform and text data. disk drive slot Insert a diskette to store ECG orders or recorded ECG data. keyboard Press the keyboard keys to control the system or to enter data. 115A MAC 5000 System — 2000657-057, Revision B...
Connect the acquisition module cable here. module connector paper tray Place paper here. STD or A4 Indicates the size of paper (standard or A4) the tray holds. battery Recharge when the screen battery symbol flashes. 116A MAC 5000 System — 2000657-057, Revision B...
Indicates the battery is re-charging. light green AC power Indicates the system is connected to AC light power. back panel Connect peripheral devices here. connectors 117A MAC 5000 System — 2000657-057, Revision B...
Connect a device requiring analog data or TTL trigger. 118A EXT.VID. Connect an external video display. Point at a MAC 5000 or MUSE CV system’s :$51,1*6 IR transceiver to transmit or receive ECG Keep leakage current within acceptable limits data.
Page 35
118A á Ergoline bicycle ergometers require connections to both ports. :$51,1*6 Keep leakage current within acceptable limits when connecting auxiliary equipment to this device. Total system leakage current must not exceed 100 microamperes. MAC 5000 System — 2000657-057, Revision B...
Creates a capital letter. Press shift+k to type a capital K. function keys Selects screen menu functions. Power Powers the system on or off. delete Erases typed characters. Copy Prints another ECG report. MAC 5000 System — 2000657-057, Revision B...
*Or advance to next stage within the Pretest Press to advance to the pretest phase*. selected phase. Exercise Press to advance to the exercise phase*. Recovery Press to advance to the recovery phase*. MAC 5000 System — 2000657-057, Revision B...
(C) to discard the data. 161B :$51,1* To ensure defibrillator protection and protection against high-frequency burns, use only the CAM-14 acquisition module with this equipment. Otherwise, serious injury could result. 2-10 MAC 5000 System — 2000657-057, Revision B...
Page 39
V3R V4R V4 V5 V6 LA LL 10 Lead IEC 14 Lead IEC AUX C1 C2 C3 C1 C2 C4 C5 C6 L R C1 C2 C3 C4 C5 C6 L 205A 204A MAC 5000 System — 2000657-057, Revision B 2-11...
Page 40
Equipment Overview: Equipment Description Leadwire Adapters The MULTI-LINK leadwires require an adapter to connect to an electrode. 4 mm pin Leadwire end Grabber 119B MACTRODE clip 2-12 MAC 5000 System — 2000657-057, Revision B...
Equipment Overview: Getting Started Getting Started Prepare the Equipment for Use Mount the System to the Trolley (Option) 171A, 172A 211A MAC 5000 System — 2000657-057, Revision B 2-13...
Page 42
Equipment Overview: Getting Started Mount the Modem to the Trolley (Modem Option) 206A 208A 207A 2-14 MAC 5000 System — 2000657-057, Revision B...
Page 43
Total system leakage current must not exceed 100 Contact GE Medical Systems Information Technologies for microamperes. information on the recommended modem. :$51,1* This is Class I equipment. The mains plug must be connected to an appropriate power supply. MAC 5000 System — 2000657-057, Revision B 2-15...
Page 44
Equipment Overview: Getting Started Mount the Client Bridge to the Trolley (Wireless Option) á The wireless option is available only in the United States. 259A, 260A, 261A 2-16 MAC 5000 System — 2000657-057, Revision B...
Page 45
Equipment Overview: Getting Started Connect the System Cables (Wireless Option) Connect External Devices (Exercise Option) Your MAC 5000 system can connect at port 1 with the following devices: GE Medical Systems Information Technologies Series T2000 treadmills, SunTech Tango blood pressure device, Colin STBP-780 blood pressure device, or Ergoline 900/900L integrated blood pressure device.
Page 46
Does the system start up without displaying error messages? The system is power operational. power Are the errors resolved? Contact GE Medical The system is Systems Information operational. Technologies Service. 156A 006A, 150A 2-18 MAC 5000 System — 2000657-057, Revision B...
Main Menu Use the Main Menu to select the different functions available on this system. The functions displayed in your Main Menu may vary due to the installation of purchased software options. 157A MAC 5000 System — 2000657-057, Revision B 2-19...
Page 48
(220 - age). The line below that is the icon available. target heart rate (a percentage of 220 - age). At the start of EXERCISE phase, a third line representing the resting heart rate will appear. 2-20 MAC 5000 System — 2000657-057, Revision B...
Page 49
More Resting ECG PediatricECG Vector Loops 15 Lead ECG EditProtocol More Exercise12 Exercise15 Hi-Res PHi-Res Master’s Step MUSE More File Manager SystemSetup Receive RemoteQuery Ord Mgr Int. Return More 182A More 180A MAC 5000 System — 2000657-057, Revision B 2-21...
Page 50
Defines the operating parameters of the system. EditProtocol or erased. 082A Exercise12 conducts the 12-lead exercise test and SystemSetup allows you to print reports. 251A Receives ECG data from other devices. Exercise12 081A Receive 2-22 MAC 5000 System — 2000657-057, Revision B...
To select More, press the function key directly below More. Acquires, prints, and stores ECG orders received from a MUSE CV system with a Hospital Information 086A Ord Mgr Int. System (HIS) interface. More 162A MAC 5000 System — 2000657-057, Revision B 2-23...
Page 52
Equipment Overview: Getting Started Using the Arrow Pad Press the middle of the pad to select More. To select More: Press the right arrow until More is highlighted. More More 177A 176A 2-24 MAC 5000 System — 2000657-057, Revision B...
Page 53
Equipment Overview: Getting Started To select ECG: Press the middle of the pad to select ECG. The system displays the next screen. Press the right arrow to highlight ECG. 178A 179A MAC 5000 System — 2000657-057, Revision B 2-25...
Type Data into a Highlighted Field The cursor goes to the next data field. Press the right arrow to highlight the First name field. Last name Last name First name Lisa First name 163A 164A 2-26 MAC 5000 System — 2000657-057, Revision B...
Page 55
Press the down arrow to highlight Female. Press the right arrow to highlight Gender. Press the middle of the pad to lock the list in place. Male Female Male Female 165A 167A MAC 5000 System — 2000657-057, Revision B 2-27...
Page 56
Equipment Overview: Getting Started Press the middle of the arrow pad to select Female. The cursor goes to the next data field. Female 166A 2-28 MAC 5000 System — 2000657-057, Revision B...
Alcohol Use a skin preparation analyzer to test each electrode site for good conductivity. :$51,1* 064B Keep the conductive parts of lead electrodes and associated parts away from other conductive parts, including earth. MAC 5000 System — 2000657-057, Revision B...
D E F LA black L yellow Right and left deltoids. LL red F green Right and left thighs. RL green N black Right and left thighs. RA white R red Right and left deltoids. MAC 5000 System — 2000657-057, Revision B...
Page 61
RA white R red Right and left deltoids. V4R gray C4R gray Right anterior chest opposite of D. V8 gray C8 gray Under midscapular line. V9 gray C9 gray Left paraspinal border. 089A MAC 5000 System — 2000657-057, Revision B...
Page 62
Just below the clavicle of the right and left arms. H orange H light blue Back of neck, avoid the carotid artery and jugular vein. M orange M light blue Center of spine on the same horizontal level as C and D. MAC 5000 System — 2000657-057, Revision B...
Page 63
Attachment point of the 2nd rib to the right sternal edge. A2 orange white 5th intercostal space on the left posterior axillary line. (Same position as V8 or C8.) V4 blue white Mid-clavicular line in the fifth intercostal space. (Same position as C4.) MAC 5000 System — 2000657-057, Revision B...
Page 64
C4R gray Mid-clavicular line in the fifth right intercostal space. V3R gray C3R gray Halfway between A and K. V7 gray C7 gray Same horizontal level of D in the posterior axillary line. MAC 5000 System — 2000657-057, Revision B...
In addition to the standard electrodes, apply one electrode on the sternum (A1) and one in location V5R/C5R (A2). CM5, CC5, ML Lead Placement Electrode Electrode Placement Mid-sternum at the second intercostal space. In the fifth intercostal space in the anterior axillary line (V5R/C5R). 254A MAC 5000 System — 2000657-057, Revision B...
Page 66
Electrode Electrode Placement Mid-sternum at the second intercostal space. In the fifth intercostal space in the anterior axillary line (V5R/C5R). On either side of the neck or anywhere above the shoulders. 255A 3-10 MAC 5000 System — 2000657-057, Revision B...
Entering Patient Information: Enter Patient Information Enter Patient Information Preparation Patient Data power 061A MAC 5000 System — 2000657-057, Revision B...
Insert the 6-pin Insert the 9-pin cable reader when you are prompted. cable connector connector into the “HOST” into port “A” on the port on the card reader. MAC 5000. HOST 219A 256A MAC 5000 System — 2000657-057, Revision B...
30.4-91.4cm necessary to re-initiate the process of receiving from the MUSE system. You may also wish to consult your hospital IT department or your local GE Medical Systems Information MUSE CV infrared device Technologies networking professional regarding modification of your wireless LAN to improve system performance.
Enter the location(s) from which the device should retrieve the orders. Ord Mgr Int 005A, 006A Select the Orcers to Receive Select one or more orders. power 090A Return The system stores the orders. 019A, 006A, 005A MAC 5000 System — 2000657-057, Revision B...
Run the test to complete this order? Edit the data. Cancel Continue Select a different order The system goes to the to complete. ECG test. Return 005A, 006A, 020A 005A, 006A, 022A, 017A, 066A Record an ECG. MAC 5000 System — 2000657-057, Revision B...
Entering Patient Information: Enter Orders Manually (Option) Enter Orders Manually (Option) Preparation Create an Order Create Order Enter the patient data. Ord Mgr Int Return power The system stores the order. 005A, 006A, 017A 072A MAC 5000 System — 2000657-057, Revision B...
Run the test to complete this order? Edit the data. Cancel Continue Select a different order to The system goes to the complete. ECG test. 005A, 006A, 020A Return Record an ECG. 005A, 006A, 022A, 017A, 066A 4-10 MAC 5000 System — 2000657-057, Revision B...
“Enable or Disable the ACI-TIPI Option” on page 14-8 To Transmit the ECG enable or disable ACI-TIPI. Did the Establishing network connection message appear? 006A See Chapter 9, “Transmitting an The ECG is transmitted. ECG” to transmit. MAC 5000 System — 2000657-057, Revision B...
To Average the ECG Data Average 015A á 005A, 006A To record a PHi-Res ECG, GE Medical Systems Information Technologies recommends a target noise level of 0.3MV or To Store the ECG less. Store 005A, 006A To Transmit the ECG See Chapter 9, “Transmitting an ECG”...
Number of Steps, Test Type, Post J (ms), Step Counter Display, Continuous Recording, and Post Exercise ECG Time. Press Return. Record the ECG Press the ECG button to record a pre-exercise ECG. 013A power 216A MAC 5000 System — 2000657-057, Revision B...
Check the waveform quality on the screen to confirm that all leadwires are correctly reattached. The MAC 5000 automatically records any additional ECGs you program. Once you record all ECGs, a final report prints.
Enter the patient data. Patient Data Return 006A, 017A Press the ECG button. 013A Enter the ACI-TIPI data. power Return 006A, 036A See the beginning of this chapter to print, store, or transmit the ECG. 061A MAC 5000 System — 2000657-057, Revision B...
Page 84
Not Present: Select this option if the patient has no chest or left arm pain or equivalent discomfort. MAC 5000 System — 2000657-057, Revision B...
You may store data during an exercise test to diskette (see Chapter 14, Final Report, Storage Setup). The data is stored Power to the protocol diskette. Make sure there is sufficient free space on the diskette before starting the test. 217B MAC 5000 System — 2000657-057, Revision B...
Press to enter comments that will be stored with the record and printed on some of the final reports. Enter BP Press to enter BP readings or to trigger a reading from an external device. MAC 5000 System — 2000657-057, Revision B...
Press Pretest to advance to the next PRE-TEST stage. If you are using a treadmill to conduct the exercise test, tell the patient to place his/her feet on the treadmill frame, not on the belt. MAC 5000 System — 2000657-057, Revision B...
Page 90
(on the treadmill controller keyboard) to manually control the test. However, once you press these keys, you control the speed and grade during the remainder of the EXERCISE and RECOVERY phases, not the protocol. MAC 5000 System — 2000657-057, Revision B...
During the test you can manually perform operations from the function keyboard. Press the Start tmill button if treadmill or ergometer has not been started yet. If you are using an ergometer, the ergometer workload is automatically controlled. MAC 5000 System — 2000657-057, Revision B...
Page 92
Select a lead to act as the median lead. This can be a fixed lead or scanned for lead with most ST depression. Writer Change the writer’s arrhythmia documentation (doc.), cubic spline (baseline control), paper speed, gain, filter, and writer on/off settings. MAC 5000 System — 2000657-057, Revision B...
Select a lead to act as the median lead. This can be a fixed lead scanned for lead with most ST depression. Writer Change the writer’s arrhythmia documentation (doc.), cubic spline (baseline control), paper speed, gain, filter, and writer on/off settings. MAC 5000 System — 2000657-057, Revision B...
Press and hold the Test end button to end the test and your system. (Select System Setup, Exercise Report, then start the test end phase. Final Report. Select Reason for termination or Comments to enter information about this exercise test. 6-10 MAC 5000 System — 2000657-057, Revision B...
Page 95
More Select to see More menu options. Main Menu Return to the system Main Menu. New Patient Remain in the exercise application and start a test for a new patient. MAC 5000 System — 2000657-057, Revision B 6-11...
Page 96
Exercise Stress Test (Option): Test Phases For your notes 6-12 MAC 5000 System — 2000657-057, Revision B...
You can edit an existing ergometer or treadmill protocol or, you can create a new protocol. These protocols are used to run an exercise test. Follow these steps to edit an existing protocol or create a new protocol: MAC 5000 System — 2000657-057, Revision B...
Select a protocol to edit, or select ‹‹ spare ›› if you want to create a new protocol. EditProtocol Power 220A á If using a blank diskette or one without stored protocols, default protocols will be written to the diskette. MAC 5000 System — 2000657-057, Revision B...
Peak report style Choose one of the following reports to print a time of peak exercise (during transition to RECOVERY phase from EXERCISE phase): No report, 12/15 Ld, Medians, and 5 second Rhythm. MAC 5000 System — 2000657-057, Revision B...
Repeat indicates the frequency of blood pressure The last stage always has an infinite duration. This prompts after the first prompt occurs. means that your MAC 5000 remains in the last stage for an infinite duration, or until you stop the exercise test. Median Set how often Median complexes are saved during a stage for the final report.
Type in your own time value or press the delete Press enter. and BP First/ key to indicate no Report, Median, or BP for this Repeat values for stage. a stage. Press enter. MAC 5000 System — 2000657-057, Revision B...
Change the EXERCISE phase information. See “Pre- Test Protocol Information” on page 7-7 as a sample guide for editing this protocol information. MAC 5000 System — 2000657-057, Revision B...
Advance to the RECOVERY phase using one of the following methods: Press Phase. Next press Recovery. Change the RECOVERY phase information. See “Pre- Test Protocol Information” on page 7-7 as a sample guide for editing this protocol information. MAC 5000 System — 2000657-057, Revision B...
Press Menu when you finish editing the phases of the protocol. An Edit menu similar to the following appears. Return Edit protocol/phase names Save current protocol Print current protocol Edit different protocol Erase current protocol New diskette Main Menu 7-10 MAC 5000 System — 2000657-057, Revision B...
Erase current protocol New diskette Main Menu To add or change another protocol, select Edit different protocol. Select Main Menu to display the Main Menu. Select Exercise to run an exercise test. MAC 5000 System — 2000657-057, Revision B 7-11...
Page 108
Edit Protocols Options: Save Current Protocol For your notes 7-12 MAC 5000 System — 2000657-057, Revision B...
Select the ECGs Select Select one or more ECGs. File Manager 005A, 006A, 019A Print the ECGs power Print Print ECGs from a different disk? Main Menu 062A New Disk 005A, 006A, 071A MAC 5000 System — 2000657-057, Revision B...
Page 112
Printing an ECG Report: Print Stored ECG Reports For your notes MAC 5000 System — 2000657-057, Revision B...
Transmitting an ECG: Transmit Stored ECGs by Modem (Option) Transmit Stored ECGs by Modem (Option) Preparation Transmit via Modem File Manager Line MAC 5000 Modem power 075A 070A MAC 5000 System — 2000657-057, Revision B...
Transmit the ECGs Type the telephone number Transmit of the receiving device. Transmit ECGs from a different disk? Main Menu Select a modem type. New Disk 005A, 006A, 071A Return 005A, 006A, 035A, 063A, 036A MAC 5000 System — 2000657-057, Revision B...
MAC 5000 occupees par des patients. Diskette MUSE CV PN 700060-003 File Manager Transmit via Infrared 1-3ft 30.4-91.4cm MUSE CV infrared device Transmit via Infrared power 1-3ft 30.4-91.4cm 070A MAC 5000 MAC 5000 107A MAC 5000 System — 2000657-057, Revision B...
Transmit Transmit ECGs from a different disk? Serial Line MUSE Network 005A, 006A, 068A Main Menu New Disk 005A, 006A, 071A Select the ECGs Select Select one or more ECGs. 005A, 006A, 019A MAC 5000 System — 2000657-057, Revision B...
Transmitting an ECG: Transmit Stored ECGs by Wireless (Option) Transmit Stored ECGs by Wireless (Option) Preparation File Manager 263A power 070A MAC 5000 System — 2000657-057, Revision B...
Select one or more ECGs. MUSE system. You may also wish to consult your hospital IT department or your local GE Medical Systems Information Technologies networking professional regarding modification of your wireless LAN to improve system performance.
Receiving an ECG: Receive ECGs by Modem (Option) Receive ECGs by Modem (Option) Preparation Receive via Modem Receive Line MAC 5000 Modem power 075A 190A MAC 5000 System — 2000657-057, Revision B 10-3...
Receiving an ECG: Receive ECGs by Modem (Option) Receive the ECGs Receive Phone Line The system is ready to receive ECGs. Are all the ECGs received? The system is ready to Cancel receive the ECGs. Main Menu 005A, 006A 10-4 MAC 5000 System — 2000657-057, Revision B...
Receiving an ECG: Receive ECGs Locally Receive ECGs Locally Preparation Receive via Infrared 1-3ft 30.4-91.4cm Receive MAC 5000 MAC 5000 106A power 190A MAC 5000 System — 2000657-057, Revision B 10-5...
Receiving an ECG: Receive ECGs Locally Receive the ECGs Receive LocalLine The system is ready to receive ECGs. Are all the ECGs received? The system is ready to Cancel receive the ECGs. Main Menu 005A, 006A 10-6 MAC 5000 System — 2000657-057, Revision B...
Receiving an ECG: Retrieve Confirmed ECGs from a MUSE CV System via Modem (Option) Retrieve Confirmed ECGs from a MUSE CV System via Modem (Option) Preparation Receive via Modem MUSE Remote Query Line MAC 5000 Modem power 075A 191A MAC 5000 System — 2000657-057, Revision B 10-7...
Type Location Retrieve ECGs from a pre-defined MUSE CV system? Return Select the location. Connect 063A, 006A, 111A, 036A, 063A, 005A Manual Dial Connect Type the telephone number. 005A, 006A, 035A, 063A 10-8 MAC 5000 System — 2000657-057, Revision B...
Disconnect If you do not know the patient’s ID number, type the patient’s last name. Then select your patient from the displayed list of patients. Select an ECG. Main Menu 005A, 006A MAC 5000 System — 2000657-057, Revision B 10-9...
MUSE system. You may also wish to consult your hospital IT department or your local GE Medical Systems Information Technologies networking professional regarding modification of your wireless LAN to improve system performance.
006A, 036A, 005A, 019A á If you do not know the patient’s ID number, type the patient’s last name. Then select your patient from the displayed list of patients. Connect 005A, 006A, 035A, 063A MAC 5000 System — 2000657-057, Revision B 10-11...
The system retrieves then displays the test. To Print the ECG Print The system retrieves then prints the test. Display or print another ECG? Return Disconnect Select an ECG. Main Menu 005A, 006A 10-12 MAC 5000 System — 2000657-057, Revision B...
Editing an ECG: Edit Demographic and Interpretive Data Edit Demographic and Interpretive Data Preparation Select the ECGs Select Select one or more ECGs. File Manager Edit power 005A, 006A, 019A 062A MAC 5000 System — 2000657-057, Revision B 11-3...
Test Information Questions Enter the Overreader Password Type the overreader password. Edit the data. Type the reviewer information. Return 189A, 006A, 044A Return 006A, 063A, 054A 11-4 MAC 5000 System — 2000657-057, Revision B...
Page 137
Hi-Res Measurements Edit the data. Edit the QRS or P-wave Onset? Offset Onset Return 055A, 006A, 056A Press the left or right arrows to increment or decrement the measurement. 057A, 005A, 006A, 063A MAC 5000 System — 2000657-057, Revision B 11-5...
Page 138
To Insert or Append an Acronym Append Insert Type a statement. Select a statement. Select a statement. Insert Append Type an acronym. Return 005A, 006A, 063A, 050A Return 058A, 005A, 006A, 063A, 050A 11-6 MAC 5000 System — 2000657-057, Revision B...
Return Select a statement. Save the edited file? Delete 005A, 006A To Join Two Statements Store 005A, 006A, 052A, 066A Select a statement to join with the preceding statement. Join 005A, 006A MAC 5000 System — 2000657-057, Revision B 11-7...
Page 140
Editing an ECG: Edit Demographic and Interpretive Data For your notes 11-8 MAC 5000 System — 2000657-057, Revision B...
Deleting an ECG: Delete Stored ECGs Delete Stored ECGs Preparation Select the ECGs Select Select one or more ECGs. File Manager 005A, 006A, 019A power 062A MAC 5000 System — 2000657-057, Revision B 12-3...
Type the System or Overread password. Delete these ECGs? The system deletes the Select a different ECG to ECGs. delete. Delete ECGs from a different disk? Main Menu New Disk 005A, 006A, 063A, 071A 12-4 MAC 5000 System — 2000657-057, Revision B...
Delete Stored ECG Orders (Option) Preparation Select the Orders Load Orders Delete Ord Mgr Int Select an order. á Only orders that have not been completed may be deleted. power Delete 072A 005A, 006A, 019A MAC 5000 System — 2000657-057, Revision B 12-5...
Deleting an ECG: Delete Stored ECG Orders (Option) Delete the Orders Delete The system deletes the orders. Receive ECG orders now? See chapter 4 to Receive Cancel orders. Cancel Main Menu 005A, 006A 12-6 MAC 5000 System — 2000657-057, Revision B...
The system will automatically ask if you want to format an unformatted disk. Lock the disk. Unlock the disk to format. You can now store or delete data. File Manager 006A, 110A, 109A power 062A MAC 5000 System — 2000657-057, Revision B 13-3...
Completing Other Tasks: Display Stored ECGs Display Stored ECGs Preparation Select the ECGs Select Select one or more ECGs. File Manager 005A, 006A, 019A power 062A MAC 5000 System — 2000657-057, Revision B 13-5...
005A, 006A To Display Medians or Rhythm Data To Display ECGs From a Different Disk Medians Rhythm New Disk 005A, 006A, 071A 005A, 006A To Display Measurement and Analysis Statements Text 005A, 006A 13-6 MAC 5000 System — 2000657-057, Revision B...
Select the System Setup Function Enter the System Setup Password Type the password. Preparation Is the password correct? Select a menu function. More System Setup System Setup 005A, 006A Type the password. power 151A MAC 5000 System — 2000657-057, Revision B 14-3...
Define the System Parameters Use the information contained in this chapter to define your system’s operating parameters. Save Your Changes System Setup Save Setup To System Main Menu 006A, 026A, 027A, 028A 14-4 MAC 5000 System — 2000657-057, Revision B...
Basic System Power Up Options Select the function the system always uses when you power on Preview before analysis your system. Return 168A, 006A, 057A, 022A Return 168A, 006A, 057A, 025A, 185A MAC 5000 System — 2000657-057, Revision B 14-5...
Select the type of ECGs transmitted. Select the type of ECGs stored. Auto ECG transmission All ECGs All ECGs No ECGs Only ABNORNAL ECGs Return 168A, 006A, 057A, 025A, 185A Return 168A, 006A, 057A, 25A, 185A MAC 5000 System — 2000657-057, Revision B 14-7...
Defining the System Setup: Program the System to Automatically do a Task Enable or Disable the ACI-TIPI Option System Setup ECG Acquisition/Analysis Enable ACI-TIPI 168A, 057A, 25A Select Yes to enable the ACI-TIPI Option; select No to disable the ACI-TIPI Option. 14-8 MAC 5000 System — 2000657-057, Revision B...
Select Low to set the system’s speaker to low volume. transmit receiving device. Select High to set the system’s speaker to high Text on bottom Select Yes to print the ECG test information on the volume. bottom of the ECG reports. MAC 5000 System — 2000657-057, Revision B 14-9...
Select Age in years to enter age in years, months, weeks, days, or hours. Gender Select Yes to display a prompt asking whether the patient is male or female. Height Select Yes to display a prompt asking the patient’s height. 14-10 MAC 5000 System — 2000657-057, Revision B...
Page 163
If you elect to number for this ECG. collect this type of data, it is your responsibility to ensure that you comply with all applicable legal requirements. MAC 5000 System — 2000657-057, Revision B 14-11...
Define the screen colors you want to system to display. Select Monochrome to view white screen elements. Select Option 1 to view white, green, yellow, and red screen elements. Select Option 2 to view white, yellow, and red screen elements. 14-12 MAC 5000 System — 2000657-057, Revision B...
Select 9600 baud rate to transmit or receive data Select Dialing only to hear the modem tones while rate between another MAC 5000 or a MUSE CV system. your system dials a telephone number. Default Location Select the default receiving device your system Dialtone required Select Yes when the system is connected to telephone transmits ECGs to.
Power the system off then on to view the new 2. Select the month. language. 3. Type the year. Current time Enter the current time: 1. Type the hour. 2. Type the minutes. 14-14 MAC 5000 System — 2000657-057, Revision B...
Select how you want to sort displayed ECG orders. Application to start up in every time you power on the system. Create orders Select Yes to allow ECG orders to be entered locally manually into the system. MAC 5000 System — 2000657-057, Revision B 14-15...
168A, 057A Item Description PS/2 Port Device Select the input device connected to the PS/2 port (keyboard or card reader). Card Reader Select Automatic or Manual configuration of the Configuration optional card reader. 14-16 MAC 5000 System — 2000657-057, Revision B...
Select Yes to prevent the Normal ECG 12SL analysis ECG data is ready for analysis when Pre-acquisition is NORMAL statement from appearing on printed, stored, and turned on. statement transmitted ECG reports. MAC 5000 System — 2000657-057, Revision B 14-17...
Page 170
á The Storage format item does not display if the program has classified as abnormal. MAC 5000 has the ACI-TIPI option installed. With Enable ACI-TIPI Select No to disable the ACI-TIPI function. ACI-TIPI, you can only store 500 Hz ECGs.
Select Numbers only to answer the prompt using Filter Select the writer’s default filter setting. numbers. á The screen filter is always set to 40 Hz. Select Yes or No to answer the prompt using either yes or no. MAC 5000 System — 2000657-057, Revision B 14-19...
Rhythm reports. This allows you to print the data you see on the screen. Resting ECG Reports Select 10 sec delayed to print ECG data delayed by Pediatric ECG Reports 10 seconds. 15 Lead Reports Vector Loops 168A, 183A 14-20 MAC 5000 System — 2000657-057, Revision B...
Page 173
12 (or 15) leads and to all reports that include Swedish format rhythm display the chest electrode placement. lead 1. Autorhythm Select the group of Rhythm leads printed in the Autorhythm report. MAC 5000 System — 2000657-057, Revision B 14-21...
Page 174
Confirmed appears on the ECG report. Select Reviewed by to display the reviewer’s name on a confirmed ECG report. If the ECG report is not confirmed, then no name appears. 14-22 MAC 5000 System — 2000657-057, Revision B...
Type a value between 0 and 100 to set a delay in milliseconds for the TTL Output QRS detector signal. QRS Beep Select On to hear a beep for each QRS complex. MAC 5000 System — 2000657-057, Revision B 14-23...
Numbers only = the answer to the prompt can only be reason at the end of the test. in numbers. Yes or No = the answer to the prompt must be yes or 14-24 MAC 5000 System — 2000657-057, Revision B...
Arrhythmia Doc. Select On to automatically print a report when an arrhythmia occurs during the exercise test. Tic marks Select Yes to add tic marks on the E, J, and J-plus-a- measurement points. MAC 5000 System — 2000657-057, Revision B 14-25...
Description Lead Set Choose the three additional leads used. You can: select one of the pre-defined lead sets, or select Custom 3 to define the electrode positions of A1, A2, and A3. 14-26 MAC 5000 System — 2000657-057, Revision B...
Page 179
Select the group of Rhythm leads printed in the reports, the first three Swedish format rhythm leads Autorhythm report. are used. When printing or storing 6 lead median reports, all six of the Swedish format rhythm leads are used. MAC 5000 System — 2000657-057, Revision B 14-27...
Report, Selected Medians, Trend & Medians, Median Report, or Trend Report format to print as a preview report before Reason for Termination and Comments are entered and all final reports are printed at test end. 14-28 MAC 5000 System — 2000657-057, Revision B...
Inputs / Outputs Use the Screen function to define how your exercise ECGs Define Inputs / Outputs when you connect additional are displayed on your screen. equipment, like an ergometer, to your MAC 5000 system. System Setup Exercise Test System Setup...
Page 182
I, II, V1, V2, V3, V4, V5, or V6. Attach an QRS Beep Select On to hear a beep for each QRS complex. acquisition module to the MAC 5000 system in order to use this output. Blood Pressure Select Manual, Ergoline Ergometer, Suntech, or Nipon-Colin.
40-250Hz. yes or no. Averaging target Select the method to average the target. Target Beat Type a value from 1 to 999. GE Medical Systems Count Information Technologies recommends averaging to a minimum of 250 beats. Target Noise Type a value from 0.1 to 1.0µV.
Number of ms after J point. Used to determine the ST level. Step Counter Select Up to display steps taken so far. Select Down Display to display steps to go during exercise. 14-32 MAC 5000 System — 2000657-057, Revision B...
GE Medical Systems Save Setup Information Technologies settings. Select From diskette to install System Setup 168A parameters stored on a disk. Select Do Not Restore Setup to exit this function. MAC 5000 System — 2000657-057, Revision B 14-33...
Page 186
Defining the System Setup: Miscellaneous Setup For your notes 14-34 MAC 5000 System — 2000657-057, Revision B...
á Regular maintenance, irrespective of usage, is essential to ensure that the equipment will always be functional when required. See the documentation provided with your peripheral equipment for appropriate maintenance procedures. MAC 5000 System — 2000657-057, Revision B...
Verify that all cords and connectors are securely seated. Inspect keys and controls for proper operation. Toggle keys should not stick in one position. Knobs should rotate fully in both directions. MAC 5000 System — 2000657-057, Revision B...
The head cleaner disk must spin for a total of 30 seconds to clean the drive. Repeat steps 1-3 until the head cleaning disk spins for 30 seconds or more. Eject the head cleaning disk from the drive slot. MAC 5000 System — 2000657-057, Revision B...
Appendix A – Maintenance: Changing the Paper Tray Size Changing the Paper Tray Size To Change to A4 Paper Size To Change to Standard Paper Size 187A 188A MAC 5000 System — 2000657-057, Revision B...
3 - 5 years. If your retention requirements exceed separators containing polyvinyl chloride or other these guidelines, we recommend you consider alternate vinyl chlorides. image storage techniques. Avoid contact with: cleaning fluids and solvents such as alcohols, ketones, esters, ether, etc. MAC 5000 System — 2000657-057, Revision B...
The following applies to Archivist thermal paper only. GE Medical Systems Information Technologies warrants that the image produced on Archivist papers by GE Medical Systems Information Technologies equipment will not fade for seven (7) years when handled according to the instructions outlined below: Archivist papers must be continuously stored below 104°F...
To ensure a fully charged battery, Between Acquisitions power off the system and connect it to an AC wall outlet 1/2 Full until you use the system again. This prolongs battery runtime. Full 160B A-10 MAC 5000 System — 2000657-057, Revision B...
Page 197
159B Your When the Battery is Completely Discharged system powers off when the battery is completely 158B discharged. To operate your system, you must connect the system to an AC wall outlet. MAC 5000 System — 2000657-057, Revision B A-11...
á After removing and replacing the battery, the battery gauge cycles. See the MAC 5000 Service Manual for more battery symbol resets to empty. Connect your system to an AC wall maintenance and diagnostic information.
Page 199
Appendix A – Maintenance: Maintaining the Battery 113A 112A, 114A MAC 5000 System — 2000657-057, Revision B A-13...
Appendix A – Maintenance: Mounting or Dismounting the Trolley Mounting or Dismounting the Trolley Mount the System Onto the Trolley 171A, 172A, 211A A-14 MAC 5000 System — 2000657-057, Revision B...
Appendix A – Maintenance: Mounting or Dismounting the Trolley Dismount the System From the Trolley 212A, 173A, 174A, 175A MAC 5000 System — 2000657-057, Revision B A-15...
Appendix A – Maintenance: Handling a Disk Handling a Disk 35 C (95 F) 10 C (50 F) 001Adisk, 002Adisk, 003Adisk, 004Adisk, 005Adisk, 006Adisk, 007Adisk, 008Adisk, 009Adisk MAC 5000 System — 2000657-057, Revision B A-17...
Page 204
Appendix A – Maintenance: Handling a Disk For your notes A-18 MAC 5000 System — 2000657-057, Revision B...
Is the battery installed? When connected to an AC wall outlet, does the green AC power light glow? Is the writer door closed? Is the software card installed? MAC 5000 System — 2000657-057, Revision B...
Check the patient’s position. The patient should remain Make sure you enter an age range less than 16. motionless during the acquisition of a resting ECG. The original ECG was acquired in an electrocardiograph without the ACI-TIPI option. MAC 5000 System — 2000657-057, Revision B...
See The following errors may occur while you are operating this “How to Reach Us” to find out how to contact GE Medical system. You may be required to perform some action. Systems Information Technologies.
750 µ V amplitude, 50 µ s duration, Orthogonal LA, LL and V6 Sensitivity Special Functions Disconnected and/or shorted lead detection, excessive AC noise, baseline wander, muscle tremor messages, preacquisition, and electrode impedance MAC 5000 System — 2000657-057, Revision B...
Page 222
Remote retrieval of confirmed ECG records from MUSE Network Series using the Async/CSI modem. Records can be Query (option) displayed and printed. RS-232 MAC and MUSE system compatible. Auxiliary Peripheral I/O port (magnetic card reader, bar code, etc.) Infrared Optional I/O MAC 5000 System — 2000657-057, Revision B...
0.3 A (typ.) Analog Outputs 2 channels 0-10 V 1 channel ±10 V (ECG-OUT) Analog Input 2 channels, 0-10 V, sampled @ approx. 100 sps 1 output channel with programmable polarity delay and width MAC 5000 System — 2000657-057, Revision B...
6.8 kg (15 lb) without paper Table Top Unit w/Cart Dimensions Handle Height 88.9 cm (35 in) Width 41.9 cm (16.5 in) Depth 66 cm (26 in) Table Top Unit w/Cart Weight 20.4 kg (45 lb) MAC 5000 System — 2000657-057, Revision B...
Degree of protection against Ordinary ingress of liquids Handling of disposable Use only parts and accessories manufactured or recommended by GE Medical Systems Information Technologies. Follow supplies and other manufacturer’s instructions for use for disposable/consumable products. Follow local environmental guidelines concerning consumables the disposal of hazardous materials.
Page 226
GE Medical Systems Information Technologies will make available upon request circuit diagrams and component parts lists for printed circuit boards deemed repairable by qualified technical personnel.
Disposal Batteries Disposing of battery by fire or burning will cause the battery to explode. The battery is recycleable. Follow local environmental guidelines concerning disposal and recycling. Batteries may be returned to GE Medical Systems Information Technologies for recycling. Device Recycleable.
User replaceable, 18 V rechargeable NiMH pack Capacity 3.5 AH ±15% Charge 5.0 hours (4.5 hours typical) (with display off) á Cannot charge battery at or above 45°C (best if below 40°C) D-10 MAC 5000 System — 2000657-057, Revision B...
Thermal, perforated, 214.63 mm x 280 mm (8.45 in x 11 in), z-fold, 300 sheets per pack (European 210 mm x 297.5 mm (8.27 in x 11.7 in) A4 style paper available) MAC 5000 System — 2000657-057, Revision B D-11...
Appendix D – Technical Specifications: Pediatric Analysis Pediatric Analysis Item Description Acquisition Simultaneous 15-lead acquisition. (Standard 12 lead plus pediatric leads V3R, V4R and V7.) MAC 5000 System — 2000657-057, Revision B D-13...
1000 samples per second per channel Digital Sampling Rate 4000 samples per second per channel High/Low Pass Filters Spectral filter using Fast Fourier Transform (FFT) ADC Resolution 1.22 µ V/LSB Analysis Resolution 0.1525 µ V/LSB D-14 MAC 5000 System — 2000657-057, Revision B...
4 by 2.5s + 3 rhythm leads in each lead. 4 by 10s One 10 second rhythm lead. 5 by 2s 5 by 2s + 1 rhythm lead 5 by 2s + 3 rhythm leads 5 by 10s 147A MAC 5000 System — 2000657-057, Revision B...
Stag Hold is ON. Stag Hold is OFF. Cubic Spline is ON. Cubic Spline is OFF. 50Hz AC filter is ON. 60Hz AC filter is OFF. Binary encoded format for heart rate leads. MAC 5000 System — 2000657-057, Revision B...
Averaged Standard Hi-Res or PHi-Res Signal 400mm/s of expanded X,Y,Z medians and a RMS voltage function/VM plot. Averaged Expanded ACI-TIPI The analysis of the acquired ECG data appears at the top of the report. MAC 5000 System — 2000657-057, Revision B...
Page 238
5 seconds for each of the 12 leads at writer speed 50mm/s. Text is on the top of the page. Vector Loops Sagittal, horizontal, and frontal plane vectorgrams. Marks on sample X,Y,Z complexes identify P onset and offset, Q onset and offset, and T onset. MAC 5000 System — 2000657-057, Revision B...
Based on exercise setups, a Linked Median, Medians & Rhythms, 6 or 12 Lead Comparative Medians & Rhythm report will print. See the ’Additional Report Names’ section for a description of these formats. MAC 5000 System — 2000657-057, Revision B...
ST/HR Loops Report A two-dimensional representation of ST Level vs. Heart Rate. ST/HR Slope Report Records linear regression of heart rate-adjusted slope for all leads, plus median morphology of lead with highest slope. MAC 5000 System — 2000657-057, Revision B...
10-14 15-19 20-24 25-29 30-34 35-39 40-44 45-49 50-54 55-59 60-64 65-69 70-74 75-79 Weight (kg) Male 18-22 Female Male 23-26 Female Male 27-31 Female Male 32-35 Female Male 36-40 Female Male 41-44 Female Male 45-49 Female Male 50-53 Female Male 54-58 Female MAC 5000 System — 2000657-057, Revision B...
Page 244
10-14 15-19 20-24 25-29 30-34 35-39 40-44 45-49 50-54 55-59 60-64 65-69 70-74 75-79 Weight (kg) Male 59-63 Female Male 64-67 Female Male 68-72 Female Male 73-76 Female Male 77-81 Female Male 82-85 Female Male 86-90 Female Male 91-94 Female Male 94-99 Female Male 100-104 Female MAC 5000 System — 2000657-057, Revision B...
ST depression ≥ 0.05mV (any leads). ST elevation ≥ 0.1mV (any leads). T wave change ≥ 0.5mV (any leads). No. 1, 2, or 3 is satisfied. Negative Positive and borderline criteria are NOT satisfied. MAC 5000 System — 2000657-057, Revision B...
Page 246
Appendix F – Master’s Step Data: ST-T Change For your notes MAC 5000 System — 2000657-057, Revision B...
Page 247
Index MAC 5000 System — 2000657-057, Revision B Index-1...
Page 248
2-19, 2-20 ANA/TTL 2-6 beat count averaging target 14-31, 14-32 analog outputs setup 14-23 Blood pressure 2-20 analysis filter 14-31, 14-32 blood pressure 14-11 archivist paper storage A-9 Borderline F-5 arrow pad 2-8 Index-2 MAC 5000 System — 2000657-057, Revision B...
Page 249
15 Lead ECG reports 14-20, 14-26 correlation threshold 14-31, 14-32 analog outputs 14-23 create orders locally setup 14-15, 14-16 ECG acquisition/analysis 14-17, 14-24 cubic spline 14-24 patient questions 14-19 pediatric ECG reports 14-20, 14-26 MAC 5000 System — 2000657-057, Revision B Index-3...
Page 250
Event names 14-24 height 14-10 exercise 2-9 height/weight in 14-11 Exercise function 2-22 Hi-Res setup EXT.VID. 2-6 analysis filter 14-31, 14-32 external video port 14-9 averaging targe 14-31 Extra questions 14-24 averaging target 14-32 Index-4 MAC 5000 System — 2000657-057, Revision B...
Page 251
ECG reports 8-3 lead labels 2-11 receive ECGs by modem 10-3 lead set 14-23, 14-26 receive ECGs locally 10-5 lead Z display 14-23 record a signal averaged ECG 5-4, 5-5 leads MAC 5000 System — 2000657-057, Revision B Index-5...
Page 252
Master’s Step F-3 patient medians 2-9 history 14-11 medications 14-11 ID number length 14-10 modem D-4 ID number required 14-10 modem speaker 14-13 questions function 14-19 muscle tremor warning 14-17 skin preparation 3-3 Index-6 MAC 5000 System — 2000657-057, Revision B...
Page 253
14-11 recall 2-9 receive safety confirmed ECGs 10-7, 10-10 definitions 1-5 via infrared communication 10-5 equipment symbols 1-8 via modem 10-3 general information 1-8 recovery 2-9 statements 1-5 Referred by 14-11 MAC 5000 System — 2000657-057, Revision B Index-7...
Page 254
14-13 sort test 2-9 ECG orders 14-15, 14-16 test Indication 14-11 file manager 14-9 text entry 14-9 space bar 2-8 thermal paper storage A-8 speaker volume 14-9 Tic marks 14-25 speed 2-9 Index-8 MAC 5000 System — 2000657-057, Revision B...
Page 255
TTL Output 14-23, 14-30 two second pause 14-13 type 14-13 BF equipment 1-11 visual inspection A-4, B-3 weight 14-11 Width 14-30 width 14-23 writer filter 14-19, 14-25 gain 14-19, 14-25 speed 14-19, 14-25 MAC 5000 System — 2000657-057, Revision B Index-9...
Page 256
Index Index-10 MAC 5000 System — 2000657-057, Revision B...