Biological Hazard; Pacemaker Hazard - GE Vivid i User Manual

Hide thumbs Also See for Vivid i:
Table of Contents

Advertisement

Safety
the unit must be checked by GE Ultrasound service personnel.
No attempt should be made to replace the fuses with others of
a higher rating.

Biological hazard

For patient and personnel safety, beware of biological hazards
while performing transesophageal procedures. To avoid the
risk of disease transmission:
Use protective barriers (gloves and probe sheaths)
whenever necessary. Follow sterile procedures as
required.
Thoroughly clean probes and reusable accessories after
each patient examination and disinfect or sterilize as
needed. Refer to Chapter 10, "Probes" on page 263, for
probe use and care instructions.
Follow all in-house infection control policies as they apply
to personnel and equipment.

Pacemaker hazard

The possibility of the system interfering with pacemakers is
minimal. However, as this system generates high frequency
electrical signals, the operator should be aware of the potential
hazard this could cause.
360
Vivid i User's Manual
2378958-100 Rev. 02

Advertisement

Table of Contents
loading

Table of Contents