Alarms
3. Adjust the alarm settings by moving the sliders to the desired values, by pressing
the up or down arrows with a field selected, or by entering values in the available
fields.
The profile used to set default values is displayed at the top of the Edit Alarm
Settings form. Settings modified from a default value are highlighted in blue.
Select Reset to Default to set all alarm values to the default values. Note that
this means reverting to the settings currently configured for the selected care
profile, not the settings that the patient was admitted with.
The high limit for SpO
range. The limit indicates OFF.
The critical low limit for SpO
specific percentage below the defined low limit for SpO
cannot be modified. As the low limit is adjusted, the critical low limit changes
and is displayed in the Edit Alarm Settings form.
4. Select Save to update the alarm settings to the modified values. When you select
Save, modified alarm settings take effect immediately.
Select Cancel to return to the Single Patient View without modifying alarm settings.

Adjusting alarm volume on the Central Viewer

You can adjust the alarm volume of the computer running the Central Viewer
application. This setting does not affect the alarm volume of the computer or the
Hub, but it does change the volume of the alarming unit. You can access the setting
from any view.
1. Select
The alarm volume adjustment bar appears.
2. Move the volume adjustment handle to the desired volume level:
The selected volume is shown as a percentage number above the adjustment
bar. You cannot select a volume below the configured minimum volume setting
(shown as MIN on the control bar).

Adjusting alarm volume on the Hub

You can adjust the alarm volume from the Hub menu. This is a patient-specific setting
so it is only available when there is an admitted patient selected on the Hub. It is
independent from the Central Viewer. You can access the menu from any view. The
minimum allowed volume can be configured for the care unit through Config UI with
service access rights. By default it is 70%.
86
% is disabled by moving the slider to the right of the
2
% is defined by the care area configuration as a
2
from the upper right corner of the screen.
Where: MIN (MIN) = configured minimum volume
Portrait
Mobile Monitoring Solution
% . This percentage
2