Omega HHP-SG User Manual page 20

Handheld strain gage indicator
Table of Contents

Advertisement

5
C
6
Look up the gain value calculated in step 5 and find the
closest gain value in Table 5.1. (If the value of gain required is
not contained in Table 5.1, skip to step 14)
7
Read across the table to find the gain desired and the gain
switch settings necessary to give you that value.
Example: S1-1 S1-2 S1-3 S1-4 S1-5 S1-6 S1-7
C
8
Open up the Indicator case by removing the four screws which
hold the rear cover in place. (see Figure 3.2)
9
Set the gain resistor switches to the positions indicated in step 7.
(open is up, and closed is down)
10 Select the decimal point location (if desired) by using the
push-on jumpers "Q" through "T" on the display circuit board.
Also select the dummy zero (if desired) by installing the
push-on jumper "U" on the display circuit board.
11 Close the Indicator case by installing the four screws removed in
step 8.
12 Connect your device to the Indicator terminals.
13 Adjust the ZERO and SPAN pots on the Indicator as
necessary to obtain the exact reading you desire.
16
USTOMIZED
Table 5.1 (partial)
O
O
S
P
CALING
O
O
ROCEDURE
O
O
S8
GAIN
O
9